Dosimetric and microdosimetric study of contrast-enhanced radiotherapy with kilovolt x-rays

被引:67
作者
Verhaegen, F
Reniers, B
Deblois, FO
Devic, S
Seuntjens, J
Hristov, D
机构
[1] McGill Univ, Montreal Gen Hosp, Dept Med Phys, Ctr Hlth, Montreal, PQ H3G 1A4, Canada
[2] Hop Maison Neuve Rosemont, Serv Radiotherapie, Montreal, PQ H1T 2M4, Canada
[3] Siemens Med Solut Inc, Concord, CA USA
关键词
D O I
10.1088/0031-9155/50/15/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Kilovolt x-rays are clearly suboptimal compared to MV photon beams for radiotherapy of deep-seated tumours because of the increased attenuation in tissue, causing a rapid dose fall-off. This picture could change drastically when tumours can be labelled with contrast medium, containing high atomic number elements. This causes a significant dose enhancement to the tumour by exploiting the high cross sections for the photo-electric effect for kV x-rays. In this work, we have investigated the dosimetric and microdosimetric characteristics of kV contrast-enhanced radiation therapy (CERT) for different photon energies, contrast-medium concentrations and types (I and Gd). Two idealized patient treatment plans (head and lung) for irradiation with CT-arc beams were calculated. It is shown that the dose enhancement in tumours can be highly significant (up to about sixfold for realistic 80-120 kVp x-ray spectra and an iodine concentration of 50 mg ml(-1)) but that dose homogeneity in the tumour depends on photon energy, contrast-medium concentration and type, and irradiation scheme. An attempt to optimize the irradiation scheme is discussed. The microdosimetric study of the dose mean lineal energy shows that radiation quality changes in the contrast-medium-labelled region compared to homogeneous tissue are fairly small and limited to 10%. It is concluded that kV-CERT is a promising radiotherapy technique, provided contrast medium can be delivered reliably to tumours.
引用
收藏
页码:3555 / 3569
页数:15
相关论文
共 30 条
[1]  
Al-Yahya K, 2002, MED PHYS, V29, P1294
[2]  
BERGER MJ, 1974, P 4 S MICR, P695
[3]  
Einstein A, 1905, ANN PHYS-BERLIN, V17, P132
[4]  
HALL EJ, 2000, RADIOBIOLOGY RADIOLO, P438
[5]  
International Commission of Radiation Units and Measurements (ICRU), 1983, 36 ICRU
[6]   Optimal photon energies for IUdR K edge radiosensitization with filtered x-ray and radioisotope sources [J].
Karnas, SJ ;
Yu, E ;
McGarry, RC ;
Battista, JJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (10) :2537-2549
[7]   Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498
[8]   TRION CODE FOR RADIATION ACTION CALCULATIONS AND ITS APPLICATION IN MICRODOSIMETRY AND RADIOBIOLOGY [J].
LAPPA, AV ;
BIGILDEEV, EA ;
BURMISTROV, DS ;
VASILYEV, ON .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 1993, 32 (01) :1-19
[9]   Correction factors for water-proofing sleeves in kilovoltage x-ray beams [J].
Ma, CM ;
Seuntjens, JP .
MEDICAL PHYSICS, 1997, 24 (09) :1507-1513
[10]   Study of dosimetry consistency for kilovoltage x-ray beams [J].
Ma, CM ;
Li, XA ;
Seuntjens, JP .
MEDICAL PHYSICS, 1998, 25 (12) :2376-2384