Tracheal intubation of morbidly obese patients:: a randomized trial comparing performance of Macintosh and Airtraq™ laryngoscopes

被引:95
作者
Ndoko, S. K. [1 ]
Amathieu, R. [1 ,3 ]
Tual, L. [1 ]
Polliand, C. [2 ]
Kamoun, W. [1 ]
El Housseini, L. [1 ]
Champault, G. [2 ,3 ]
Dhonneur, G. [1 ]
机构
[1] Jean Verdier Publ Univ Hosp Paris, APHP, Dept Anaesthesia & Intens Care, F-93143 Bondy, France
[2] Jean Verdier Publ Univ Hosp Paris, APHP, Visceral & Obes Surg Dept, F-93143 Bondy, France
[3] Paris 13 Sch Med, F-93000 Bobigny, France
关键词
airway; equipment; laryngoscope; oxygen; therapy;
D O I
10.1093/bja/aem346
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. The Airtraq(TM) laryngoscope is designed to allow visualization of the glottis without alignment of the oral, pharyngeal, and laryngeal axes. We hypothesized that this new airway device would facilitate tracheal intubation of morbidly obese patients. We compared tracheal intubation performance of standard Macintosh laryngoscope with the Airtraq(TM) laryngoscope in morbidly obese patients. Methods. One hundred and six consecutive ASA I-III morbidly obese patients undergoing surgery were randomized to intubation with the Macintosh laryngoscope or the Airtraq(TM) laryngoscope. Induction of anaesthesia was standardized. If tracheal intubation failed within 120 s with the Macintosh or Airtraq(TM) , laryngoscopes were switched. Success rate, SpO(2), duration of tracheal intubation, and quality of airway management were evaluated and compared between the groups. Results. Preoperative characteristics of the patients were similar in both groups. In the Airtraq(TM) group, tracheal intubation was successfully carried out in all patients within 120 s. In the Macintosh laryngoscope group, six patients required intubation with the Airtraq(TM) laryngoscope. The mean (SD) time taken for tracheal intubation was 24 (16) and 56 (23) s, respectively, with the Airtraq(TM) and Macintosh laryngoscopes, (P < 0.001). SpO(2) was better maintained in the Airtraq(TM) group than in the Macintosh laryngoscope group with one and nine patients, respectively, demonstrating drops of SpO(2) to 92% or less (P < 0.05). Conclusions. In this study, the Airtraq(TM) laryngoscope shortened the duration of tracheal intubation and prevented reductions in arterial oxygen saturation in morbidly obese patients.
引用
收藏
页码:263 / 268
页数:6
相关论文
共 10 条
  • [1] The intubation difficulty scale (IDS) - Proposal and evaluation of a new score characterizing the complexity of endotracheal intubation
    Adnet, F
    Borron, SW
    Racine, SX
    Clemessy, JL
    Fournier, JL
    Plaisance, P
    Lapandry, C
    [J]. ANESTHESIOLOGY, 1997, 87 (06) : 1290 - 1297
  • [2] EFFECTIVENESS OF PREOXYGENATION IN MORBIDLY OBESE PATIENTS
    BERTHOUD, MC
    PEACOCK, JE
    REILLY, CS
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1991, 67 (04) : 464 - 466
  • [3] Brodsky JB, 2003, ANESTH ANALG, V96, P1841, DOI 10.1213/01.ANE.0000063165.15467.1B
  • [4] COUSSA M, 2001, ANESTH ANALG, V95, P1491
  • [5] Tracheal intubation of morbidly obese patients:: LMA CTrach™ vs direct laryngoscopy
    Dhonneur, G.
    Ndoko, S. K.
    Yavchitz, A.
    Foucrier, A.
    Fessenmeyer, C.
    Polliand, C.
    Combes, X.
    Tual, L.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (05) : 742 - 745
  • [6] Tracheal intubation using the Airtraq® in morbid obese patients undergoing emergency cesarean delivery
    Dhonneur, Gilles
    Ndoko, Serge
    Amathieu, Roland
    el Housseini, Lodfi
    Poncelet, Christophe
    Tual, Loic
    [J]. ANESTHESIOLOGY, 2007, 106 (03) : 629 - 630
  • [7] Iked N, 2001, Rev Stomatol Chir Maxillofac, V102, P305
  • [8] Cephalometric comparison of pharyngeal changes in subjects with upper airway resistance syndrome or obstructive sleep apnoea in upright and supine positions
    Ingman, T
    Nieminen, T
    Hurmerinta, K
    [J]. EUROPEAN JOURNAL OF ORTHODONTICS, 2004, 26 (03) : 321 - 326
  • [9] A comparison of tracheal intubation using the Airtraq® or the Macintosh laryngoscope in routine airway management:: a randomised, controlled clinical trial
    Maharaj, C. H.
    O'Croinin, D.
    Curley, G.
    Harte, B. H.
    Laffey, J. G.
    [J]. ANAESTHESIA, 2006, 61 (11) : 1093 - 1099
  • [10] Endotracheal intubation in patients with cervical spine immobilization - A comparison of Macintosh and Airtraq laryngoscopes
    Maharaj, Chrisen H.
    Buckley, Elma
    Harte, Brian H.
    Laffey, John G.
    [J]. ANESTHESIOLOGY, 2007, 107 (01) : 53 - 59