Methodological proposal for validation of the disinfecting efficacy of an automated flexible endoscope reprocessor

被引:0
|
作者
Graziano, Kazuko Uchikawa [1 ]
Auler Pereira, Marta Elisa
Koda, Elaine [2 ]
机构
[1] Univ Sao Paulo, Escola Enfermagem, BR-05403000 Sao Paulo, SP, Brazil
[2] Surg Ctr, Sao Paulo, Brazil
来源
REVISTA LATINO-AMERICANA DE ENFERMAGEM | 2016年 / 24卷
关键词
Methodology; Disinfection; Efficacy; Endoscopes;
D O I
10.1590/1518-8345.0595.2745
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Objective: to elaborate and apply a method to assess the efficacy of automated flexible endoscope reprocessors at a time when there is not an official method or trained laboratories to comply with the requirements described in specific standards for this type of health product in Brazil. Method: the present methodological study was developed based on the following theoretical references: International Organization for Standardization (ISO) standard ISO 15883-4/2008 and Brazilian Health Surveillance Agency (Agencia Nacional de Vigilancia Sanitaria - ANVISA) Collegiate Board Resolution ( Resolucao de Diretoria Colegiada - RDC) no. 35/2010 and 15/2012. The proposed method was applied to a commercially available device using a high-level 0.2% peracetic acid-based disinfectant. Results: the proposed method of assessment was found to be robust when the recommendations made in the relevant legislation were incorporated with some adjustments to ensure their feasibility. Application of the proposed method provided evidence of the efficacy of the tested equipment for the high-level disinfection of endoscopes. Conclusion: the proposed method may serve as a reference for the assessment of flexible endoscope reprocessors, thereby providing solid ground for the purchase of this category of health products.
引用
收藏
页数:9
相关论文
共 5 条
  • [1] Validation of a sample method for the whole flexible endoscope's channels are experimentally contaminated by microorganisms.
    Duc, DL
    Sing, JSC
    Soule, H
    Fauconnier, J
    Marchetti, B
    Mallaret, MR
    Lerroux, N
    Ducel, G
    Calop, J
    PATHOLOGIE BIOLOGIE, 1998, 46 (01): : 34 - 38
  • [2] Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels
    Alfa, Michelle J.
    Fatima, Iram
    Olson, Nancy
    AMERICAN JOURNAL OF INFECTION CONTROL, 2013, 41 (03) : 245 - 248
  • [3] Prion disease and recommended procedures for flexible endoscope reprocessing - a review of policies worldwide and proposal for a simplified approach
    Kampf, G.
    Jung, M.
    Suchomel, M.
    Saliou, P.
    Griffiths, H.
    Vos, C.
    JOURNAL OF HOSPITAL INFECTION, 2020, 104 (01) : 92 - 110
  • [4] A prospective, quasi-experimental study on the efficacy of a novel double-headed endoscope cleaning brush for cleaning flexible endoscopes
    Shen, Rui
    Wu, Yaping
    Lv, Jiajun
    Liu, Qukai
    Yao, Weiyan
    Chen, Shu
    Liu, Dandan
    Xi, Huijun
    Zhang, Yibo
    BMC GASTROENTEROLOGY, 2025, 25 (01)
  • [5] Comparison of the efficacy of disinfectants in automated endoscope reprocessors for colonoscopes: tertiary amine compound (Sencron2((R))) versus ortho-phthalaldehyde (Cidex((R))OPA)
    Seo, Hyun Il
    Lee, Dae Sung
    Yoon, Eun Mi
    Kwon, Min-Jung
    Park, Hyosoon
    Jung, Yoon Suk
    Park, Jung Ho
    Sohn, Chong Il
    Park, Dong Il
    INTESTINAL RESEARCH, 2016, 14 (02) : 178 - 182