Tibial Intraosseous Insertion in Pediatric Emergency Care: A Review Based upon Postmortem Computed Tomography

被引:16
作者
Harcke, H. Theodore [1 ]
Curtin, Riley N. [1 ]
Harty, M. Patricia [1 ]
Gould, Sharon W. [1 ]
Vershvovsky, Jennie [2 ]
Collins, Gary L.
Murphy, Stephen [1 ]
机构
[1] Nemours Alfred I duPont Hosp Children, Wilmington, DE USA
[2] Delaware Div Forens Sci, Wilmington, DE USA
关键词
intraosseous infusion; pediatric emergency care; postmortem computed tomography; tibia infusion; VASCULAR ACCESS; DEVICE IMPLEMENTATION; EZ-IO; INFUSION;
D O I
10.1080/10903127.2019.1698682
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: The proximal tibia is a recommended and commonly used site for pediatric emergency intraosseous vascular access (IO). During forensic whole body postmortem computed tomography (PMCT), we evaluated accuracy of emergency placement of tibial IO access. Methods: We conducted a retrospective review of 92 state medical examiner cases to assess presence and placement of tibial IO needles. Insertions were classified as successful (needle tip in the medullary portion of the bone) or unsuccessful (all other non-medullary placements) based upon position of the needle tip. Medical records were reviewed for patient age, equipment, and where an insertion was attempted, as well as if IO placement occurred in a prehospital or hospital environment. Results: Thirty-one cases with 42 tibial devices (aged 3 weeks to 16 years, median 4 months) were identified. In 25 insertions (60%), the needle tip was in satisfactory position. In 17 placements (40%), needle tip was unsatisfactory and included tibia perforation (6), tip embedded in the cortex (6), and needle missed the bone (5). In patients older than 6 months, all six placements of a 15-mm needle were successful. In infants age 6 months or younger, 14 placements (56%) were successful and 11 (44%) unsuccessful. The 25-mm IO needle was successfully placed in five of six children older than 6 months. In infants age 6 months or younger, the 25-mm needle was unsuccessfully placed in five of five attempts. Conclusion: In infants 6 months of age or younger, tibial IO needle insertion had a 53% failure rate (non-medullary placement). Failures occur during both prehospital and emergency department care. In infants age 6 months or younger, use of a 25-mm needle should be avoided. Procedures for IO insertion in infants age 6 months or younger should be reviewed and modification considered.
引用
收藏
页码:665 / 671
页数:7
相关论文
共 19 条
[1]   Intraosseous Vascular Access in Radiology: Review of Clinical Status [J].
Baadh, Amanjit S. ;
Singh, Ayushi ;
Choi, Andrew ;
Baadh, Palvir K. ;
Katz, Douglas S. ;
Harcke, Howard T. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) :241-247
[2]  
Cooper B R, 2007, J R Army Med Corps, V153, P314
[3]  
EZ-IO Intraosseous Vascular Access Needles, 2014, EZ IO INTR VASC ACC
[4]   Use of the Pediatric EZ-IO Needle by Emergency Medical Services Providers [J].
Frascone, Ralph J. ;
Jensen, Joe ;
Rewerka, Sandi S. ;
Salzman, Joshua G. .
PEDIATRIC EMERGENCY CARE, 2009, 25 (05) :329-332
[5]   Pediatric postmortem computed tomography: initial experience at a children's hospital in the United States [J].
Gould, Sharon W. ;
Harty, M. Patricia ;
Givler, Nicole E. ;
Christensen, Theresa E. ;
Curtin, Riley N. ;
Harcke, Howard T. .
PEDIATRIC RADIOLOGY, 2019, 49 (09) :1113-1129
[6]  
Gould SW, 2017, CURR RADIOL REP, V5, P62, DOI [10.1007/s40134-017-0250-8, DOI 10.1007/S40134-017-0250-8]
[7]   Placement of Tibial Intraosseous Infusion Devices [J].
Harcke, H. Theodore ;
Crawley, Geoffrey ;
Mabry, Robert ;
Mazuchowski, Edward .
MILITARY MEDICINE, 2011, 176 (07) :824-827
[8]   EZ-IO® intraosseous device implementation in German Helicopter Emergency Medical Service [J].
Helm, Matthias ;
Haunstein, Benedikt ;
Schlechtriemen, Thomas ;
Ruppert, Matthias ;
Lampl, Lorenz ;
Gaessler, Michael .
RESUSCITATION, 2015, 88 :43-47
[9]   Powered intraosseous insertion provides safe and effective vascular access for pediatric emergency patients [J].
Horton, Mark A. ;
Beamer, Cynthia .
PEDIATRIC EMERGENCY CARE, 2008, 24 (06) :347-350
[10]   SEMIAUTOMATIC INTRAOSSEOUS DEVICES IN PEDIATRIC PREHOSPITAL CARE [J].
Myers, Lucas A. ;
Russi, Christopher S. ;
Arteaga, Grace M. .
PREHOSPITAL EMERGENCY CARE, 2011, 15 (04) :473-476