Effects of Charite artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol

被引:172
作者
Goel, VK
Grauer, JN
Patel, TC
Biyani, A
Sairyo, K
Vishnubhotla, S
Matyas, A
Cowgill, I
Shaw, M
Long, R
Dick, D
Panjabi, MM
Serhan, H
机构
[1] Univ Toledo, Coll Engn, Dept Bioengn, Spine Res Ctr, Toledo, OH 43606 USA
[2] Med Univ Ohio, Toledo, OH USA
[3] Yale Univ, Sch Med, Dept Orthoped & Rehabil, New Haven, CT USA
[4] Commonwealth Orthoped, Fairfax, VA USA
[5] DePuy Spine Inc, Boston, MA USA
关键词
Charite artificial disc; biomechanics; loads; stresses; contact area and contact force; in vitro cadaver studies; finite element modeling;
D O I
10.1097/01.brs.0000195897.17277.67
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Finite element model of L3 - S1 segment and confirmatory cadaveric testing were used to investigate the biomechanical effects of a mobile core type artificial disc (Charite artificial disc; DePuy Spine, Raynham, MA) on the lumbar spine. Objective. To determine the effects of the Charite artificial disc across the implanted and adjacent segments. Summary of Background Data. Biomechanical studies of artificial discs that quantify parameters, like the load sharing and stresses, are sparse in the literature, especially for mobile-type core artificial disc designs. In addition, there is no standard protocol for studying the adjacent segmental effects of such implants. Methods. Human osteo-ligamentous spines (L1 - S1) were tested before and after L5 - S1 Charite 'artificial disc placement. The data were used to validate further an intact 3-dimensional (3-D) nonlinear L3 - S1 finite element model. The model was subjected to 400-N axial compression and 10.6 Nm of flexion/extension pure moments ( load control) or pure moments that produced the overall rotation of the L3 - S1 Charite model equal to the intact case ( hybrid approach). Resultant motion, load, and stress parameters were analyzed at the experimental and adjacent levels. Results. Finite element model validation was achieved only with the load-controlled experiments. The hybrid approach, believed to be more clinically relevant, revealed that Charite 'artificial disc leads to motion increases in flexion (19%) and extension (44%) at the L5 - S1 level. At the instrumented level, the decrease in the facet loads was less than at the adjacent levels; the corresponding decrease being 26% at L3 - L4, 25% at L4 - L5, and 13.4% at L5 - S1 when compared to the intact. Intradiscal pressure changes in the L4 - L5 and L3 - L4 segments were minimal. Shear stresses at the Charite 'artificial disc-L5 endplate interface were higher than those at S1 interface. However, in the load control mode, the increase in facet loads in extension was approximately 14%, as compared to the intact case. Conclusions. The hybrid testing protocol is advocated because it better reproduces clinical observations in terms of motion following surgery, using pure moments. Using this approach, we found that the Charite artificial disc placement slightly increases motion at the implanted level, with a resultant increase in facet loading when compared to the adjacent segments, while the motions and loads decrease at the adjacent levels. However, in the load control mode that we believe is not that clinically relevant, there was a large increase in motion and a corresponding increase in facet loads, as compared to the intact.
引用
收藏
页码:2755 / 2764
页数:10
相关论文
共 40 条
[1]   BIOMECHANICAL EVALUATION OF LUMBAR SPINAL STABILITY AFTER GRADED FACETECTOMIES [J].
ABUMI, K ;
PANJABI, MM ;
KRAMER, KM ;
DURANCEAU, J ;
OXLAND, T ;
CRISCO, JJ .
SPINE, 1990, 15 (11) :1142-1147
[2]   The spondylolytic vertebra and its adjacent segment - Mobility measured before and after posterolateral fusion [J].
Axelsson, P ;
Johnsson, R ;
Stromqvist, B .
SPINE, 1997, 22 (04) :414-417
[3]  
Chen W J, 2001, Spine (Phila Pa 1976), V26, pE519, DOI 10.1097/00007632-200111150-00024
[4]   Biomechanical evaluation of total disc replacement arthroplasty:: An in vitro human cadaveric model [J].
Cunningham, BW ;
Gordon, JD ;
Dmitriev, AE ;
Hu, NB ;
McAfee, PC .
SPINE, 2003, 28 (20) :S110-S117
[5]   COMPARISON OF IN-VIVO AND IN-VITRO ADJACENT SEGMENT MOTION AFTER LUMBAR FUSION [J].
DEKUTOSKI, MB ;
SCHENDEL, MJ ;
OGILVIE, JW ;
OLSEWSKI, JM ;
WALLACE, LJ ;
LEWIS, JL .
SPINE, 1994, 19 (15) :1745-1751
[6]   ProDisc artificial total lumbar disc replacement: Introduction and early results from the United States clinical trial [J].
Delamarter, RB ;
Fribourg, DM ;
Kanim, LFA ;
Bae, H .
SPINE, 2003, 28 (20) :S167-S175
[7]   Load-Sharing Between Anterior and Posterior Elements in a Lumbar Motion Segment Implanted With an Artificial Disc [J].
Dooris, Andrew P. ;
Goel, Vijay K. ;
Grosland, Nicole M. ;
Gilbertson, Lars G. ;
Wilder, David G. .
SPINE, 2001, 26 (06) :E122-E129
[8]   Biomechanical study on the effect of cervical spine fusion on adjacent-level intradiscal pressure and segmental motion [J].
Eck, JC ;
Humphreys, SC ;
Lim, TH ;
Jeong, ST ;
Kim, JG ;
Hodges, SD ;
An, HS .
SPINE, 2002, 27 (22) :2431-2434
[9]   COMPARISON OF RADIOGRAPHIC FINDINGS IN FUSION AND NON-FUSION PATIENTS 10 OR MORE YEARS FOLLOWING LUMBAR-DISK SURGERY [J].
FRYMOYER, JW ;
HANLEY, EN ;
HOWE, J ;
KUHLMANN, D ;
MATTERI, RE .
SPINE, 1979, 4 (05) :435-440
[10]   The fate of the adjacent motion segments after lumbar fusion [J].
Gillet, P .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2003, 16 (04) :338-345