Can a single-shot black-blood T2-weighted spin-echo echo-planar imaging sequence with sensitivity encoding replace the respiratory-triggered turbo spin-echo sequence for the liver? An optimization and feasibility study

被引:57
作者
Hussain, SM
De Becker, J
Hop, WCJ
Dwarkasing, S
Wielopolski, PA
机构
[1] Erasmus MC, Dept Radiol, Sect Abdominal Imaging, NL-3015 GD Rotterdam, Netherlands
[2] Philips Med Syst, Best, Netherlands
[3] Erasmus MC, Dept Epidemiol & Biostat, NL-3015 GD Rotterdam, Netherlands
关键词
liver; MR; echo-planar imaging; turbo spinecho; image optimization; image quality; T2-weighted images; black blood; diffusion-weighted imaging;
D O I
10.1002/jmri.20269
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To optimize and assess the feasibility of a single-shot black-blood T2-weighted spin-echo echo-planar imaging (SSBB-EPI) sequence for MRI of the liver using sensitivity encoding (SENSE), and compare the results with those obtained with a T2-weighted turbo spin-echo USE) sequence. Materials and Methods: Six volunteers and 16 patients were scanned at 1.5T (Philips Intera). In the volunteer study, we optimized the. SSBB-EPI sequence by interactively changing the parameters (i.e., the resolution, echo time (TE), diffusion weighting with low b-values, and polarity of the phase-encoding gradient) with regard to distortion, suppression of the blood signal, and sensitivity to motion. The influence of each change was assessed. The optimized SSBB-EPI sequence was applied in patients (N = 16). A number of items, including the overall image quality (on a scale of 1-5), were used for graded evaluation. In addition, the signal-to-noise ratio (SNR) of the liver was calculated. Statistical analysis was carried out with the use of Wilcoxons signed rank test for comparison of the SSBB-EPI and TSE sequences, with P 0.05 considered the limit for significance. Results: The SSBB-EPI sequence was improved by the following steps: 1) less frequency points than phase-encoding steps, 2) a b-factor of 20, and 3) a reversed polarity of the phase-encoding gradient. In patients, the mean overall image quality score for the optimized SSBB-EPI (3.5 (range: 1- 4)) and TSE (3.6 (range: 3- 4)), and the SNR of the liver on SSBB-EPI (mean +/- SD = 7.6 +/- 4.0) and TSE (8.9 +/- 4.6) were not significantly different (P >.05). Conclusion: Optimized SSBB-EPI with SENSE proved to be feasible in patients, and the overall image quality and SNR of the liver were comparable to those achieved with the standard respiratory-triggered T2-weighted TSE sequence.
引用
收藏
页码:219 / 229
页数:11
相关论文
共 29 条
[1]   ANALYSIS OF FLOW EFFECTS IN ECHO-PLANAR IMAGING [J].
BUTTS, K ;
RIEDERER, SJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03) :285-293
[2]   ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
FELMLEE, JP ;
GRIMM, RC .
RADIOLOGY, 1993, 189 (01) :259-264
[3]   FAST SELECTIVE BLACK BLOOD MR IMAGING [J].
EDELMAN, RR ;
CHIEN, D ;
KIM, D .
RADIOLOGY, 1991, 181 (03) :655-660
[4]   VALUE OF T1-RELAXATION AND T2-RELAXATION TIMES FROM ECHOPLANAR MR IMAGING IN THE CHARACTERIZATION OF FOCAL HEPATIC-LESIONS [J].
GOLDBERG, MA ;
HAHN, PF ;
SAINI, S ;
COHEN, MS ;
REIMER, P ;
BRADY, TJ ;
MUELLER, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (05) :1011-1017
[5]   DIFFERENTIATION BETWEEN HEMANGIOMAS AND METASTASES OF THE LIVER WITH ULTRAFAST MR IMAGING - PRELIMINARY-RESULTS WITH T2 CALCULATIONS [J].
GOLDBERG, MA ;
SAINI, S ;
HAHN, PF ;
EGGLIN, TK ;
MUELLER, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (04) :727-730
[6]   Single breath-hold T2-weighted MR imaging of the liver: Value of single-shot fast spin-echo and multishot spin-echo echoplanar imaging [J].
Hori, M ;
Murakami, T ;
Kim, T ;
Kanematsu, M ;
Tsuda, K ;
Takahashi, S ;
Takamura, M ;
Hoshi, H ;
Nakamura, H .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (05) :1423-1431
[7]  
Kanematsu M, 1999, JMRI-J MAGN RESON IM, V10, P25, DOI 10.1002/(SICI)1522-2586(199907)10:1<25::AID-JMRI4>3.3.CO
[8]  
2-I
[9]  
KIEFER B, 1998, ECHO PLANAR IMAGING, P582
[10]   MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES [J].
MANSFIELD, P .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03) :L55-L58