Creation and characterization of japanese standards for myocardial perfusion SPECT: Database from the Japanese society of nuclear medicine working group

被引:47
作者
Nakajima, Kenichi [1 ]
Kumita, Shinichiro [1 ]
Ishida, Yoshio [1 ]
Momose, Mitsuru [1 ]
Hashimoto, Jun [1 ]
Morita, Koichi [1 ]
Taki, Junichi [1 ]
Yamashina, Shohei [1 ]
Maruno, Hirotaka [1 ]
Ogawa, Masami [1 ]
Kubota, Masahiro [1 ]
Takahashi, Munehiro [1 ]
Odagawa, Tetsuro [1 ]
Yokozuka, Koichi [1 ]
机构
[1] Kanazawa Univ Hosp, Kanazawa, Ishikawa 9208641, Japan
关键词
myocardial perfusion imaging; SPECT; standard; Japanese population; working group of the Japanese Society of Nuclear Medicine;
D O I
10.1007/s12149-007-0059-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Standards for myocardial single- photon emission computed tomography (SPECT) adapted for a Japanese population were not available. The purpose of this study was to create standard fi les approved by the Japanese Society of Nuclear Medicine and to make known the characteristics of the myocardial perfusion pattern of this population. Methods With the collaboration of nine hospitals, a total of 326 sets of exercise - rest myocardial perfusion images were accumulated from subjects with a low likelihood of cardiac diseases. The normal database included a Tc-99m- MIBI/ tetrofosmin myocardial perfusion study with 360 (n = 80) and 180 (= 56) rotations, Tl-201 study with 360 (n = 115) and 180 rotations (n = 54) and a dual- isotope study with 360 rotation (n = 27). The projection images were transferred by digital imaging and communications in medicine (DICOM) format and reconstructed and analyzed with polar maps. Results The projection data from multiple centers were successfully transferred to a common format for SPECT reconstruction. When the average values were analyzed using a 17-segment model, myocardial counts in the septal segment differed signifi cantly between 180 and 360 rotation acquisitions. Regional differences were observed between men and women in the inferior and anterior regions. A tracer difference between Tc-99m and Tl-201 was also observed in some segments. The attenuation patterns differed signifi cantly between subjects from the United States and those from Japan. Conclusions Myocardial perfusion data that were specifi c for the Japanese population were generated. The normal database can serve a standard for nuclear cardiology work conducted in Japan.
引用
收藏
页码:505 / 511
页数:7
相关论文
共 14 条
[1]  
Faber TL, 1999, J NUCL MED, V40, P650
[2]  
Faber TL, 2004, J NUCL MED, V45, P745
[3]  
GERMANO G, 1995, J NUCL MED, V36, P2138
[4]  
GERMANO G, 2003, ATLAS NUCL CARDIOLOG, P167
[5]   ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging - Executive summary - A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to revise the 1995 guidelines for the clinical use of cardiac radionuclide imaging) [J].
Klocke, FJ ;
Baird, MG ;
Lorell, BH ;
Bateman, TM ;
Messer, JV ;
Berman, DS ;
O'Gara, PT ;
Carabello, BA ;
Russell, RO ;
Cerqueira, MD ;
Sutton, MGS ;
DeMaria, AN ;
Udelson, JE ;
Kennedy, JW ;
Verani, MS ;
Williams, KA ;
Antman, EM ;
Smith, SC ;
Alpert, JS ;
Gregoratos, G ;
Anderson, JL ;
Hiratzka, LF ;
Faxon, DP ;
Hunt, SA ;
Fuster, V ;
Jacobs, AK ;
Gibbons, RJ ;
Russell, RO .
CIRCULATION, 2003, 108 (11) :1404-1418
[6]   Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET:: comparison with cardiac magnetic resonance imaging [J].
Lipke, CSA ;
Kühl, HP ;
Nowak, B ;
Kaiser, HJ ;
Reinartz, P ;
Koch, KC ;
Buell, U ;
Schaefer, WM .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (04) :482-490
[7]   Comparison of automatic quantification software for the measurement of ventricular volume and ejection fraction in gated myocardial perfusion SPECT [J].
Lum, DP ;
Coel, MN .
NUCLEAR MEDICINE COMMUNICATIONS, 2003, 24 (03) :259-266
[8]   Inter-institution preference-based variability of ejection fraction and volumes using quantitative gated SPECT with 99mTc-tetrofosmin:: a multicentre study involving 106 hospitals [J].
Nakajima, K ;
Nishimura, T .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 (02) :127-133
[9]  
Nakajima K, 2001, J NUCL MED, V42, P1571
[10]   Normal limits of ejection fraction and volumes determined by gated SPECT in clinically normal patients without cardiac events: a study based on the J-ACCESS database [J].
Nakajima, Kenichi ;
Kusuoka, Hideo ;
Nishimura, Shigeyuki ;
Yamashina, Akira ;
Nishimura, Tsunehiko .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (07) :1088-1096