Asymmetric vascular stent feasibility study of a new low-porosity patch-containing stent

被引:26
作者
Ionita, Ciprian N. [2 ,3 ,5 ]
Paciorek, Ann M. [5 ]
Hoffmann, Kenneth R. [2 ,5 ]
Bednarek, Daniel R. [3 ,5 ]
Yamamoto, Junichi [1 ,2 ,5 ]
Kolega, John [4 ]
Levy, Elad I. [1 ,2 ,3 ,5 ]
Hopkins, L. Nelson [1 ,2 ,3 ,5 ]
Rudin, Stephen [2 ,3 ,5 ]
Mocco, J. [1 ,2 ,5 ]
机构
[1] Univ Buffalo Neurosurg, Millard Fillmore Gates Hosp, Dept Neurosurg, Buffalo, NY 14209 USA
[2] SUNY Buffalo, Sch Med & Biomed Sci, Dept Neurosurg, Buffalo, NY 14260 USA
[3] SUNY Buffalo, Sch Med & Biomed Sci, Dept Radiol, Buffalo, NY 14260 USA
[4] SUNY Buffalo, Sch Med & Biomed Sci, Dept Pathol, Buffalo, NY 14260 USA
[5] SUNY Buffalo, Sch Med & Biomed Sci, Toshiba Stroke Res Ctr, Buffalo, NY 14260 USA
关键词
aneurysm; stent; canine model; time-density curve;
D O I
10.1161/STROKEAHA.107.503862
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and Purpose - Intracranial aneurysm (IA) treatment through hemodynamic modification with novel stent designs is a burgeoning area of research. We present a feasibility study for a new low-porosity patch-containing stent designed to treat intracranial aneurysms. The device is deployed so the patch covers the aneurysm neck ensuring strong flow diversion away from the aneurysm while keeping a low probability of occlusion of perforating vessels. Methods - We created 17 side-wall aneurysms in 6 dogs, 2 per carotid artery if animal size permitted. Twelve proximal aneurysms were treated with AVSs: 5 distal aneurysms were untreated, serving as controls against self-thrombosis; 7 treated aneurysms were fully-covered; and 5 were partially-covered. After 4 weeks, a final angiogram was performed and aneurysms were explanted. Angiograms acquired pre- and posttreatment and at 4-week follow-up were analyzed quantitatively using normalized time-density curves (NTDC). Cone-beam micro-CT and histological specimen analysis were then performed. Results - Posttreatment, NTDC average peaks dropped to 45% of initial values for the partially-covered aneurysms and 78% for the fully-covered aneurysms. Cone-beam micro-CT imaging performed at 4 weeks posttreatment showed partial thrombosis in 4 of 5 partially-covered aneurysms and complete thrombosis in all fully-covered aneurysms. Histology revealed neointimal coverage of all asymmetrical patch regions and thrombus formation in both fully-and partially-covered aneurysms. Four-week follow-up was not done for 1 animal ( 2 controls, 2 treated) that expired because of groin hemorrhage and for another animal ( 1 aneurysm) with an occluded carotid. Conclusions - We demonstrate aneurysmal blood flow diversion using a new low-porosity patch-containing asymmetrical vascular stent in a canine side-wall aneurysm model. Overall results are encouraging and support continued AVS development.
引用
收藏
页码:2105 / 2113
页数:9
相关论文
共 28 条
[1]  
Cebral JR, 2005, AM J NEURORADIOL, V26, P2550
[2]   Double-stent method: therapeutic alternative for small wide-necked aneurysms [J].
Doerfler, A ;
Wanke, I ;
Egelhof, T ;
Stolke, D ;
Forsting, M .
JOURNAL OF NEUROSURGERY, 2004, 100 (01) :150-154
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]   MEASUREMENT OF ABSOLUTE FLOW-RATE IN VESSELS USING A STEREOSCOPIC DSA SYSTEM [J].
FENCIL, LE ;
DOI, K ;
CHUA, KG ;
HOFFMAN, KR .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) :659-671
[5]   Method to rotate an endovascular device around the axis of a vessel using an external magnetic field [J].
Florin, E. ;
Rangwala, H. S. ;
Rudin, S. .
MEDICAL PHYSICS, 2007, 34 (01) :328-333
[6]  
HOI Y, 2006, P SPIE, V6143
[7]   Microangiographic image guided localization of a new asymmetric stent for treatment of cerebral aneurysms [J].
Ionita, CN ;
Rudin, S ;
Hoffmann, KR ;
Bednarek, DR .
Medical Imaging 2005: Visualization, Image-Guided Procedures, and Display, Pts 1 and 2, 2005, 5744 :354-365
[8]   Particle image velocimetry (PIV) evaluation of flow modification in aneurysm phantoms using asymmetric stents [J].
Ionita, CN ;
Hoi, Y ;
Meng, H ;
Rudin, S .
MEDICAL IMAGING 2004: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2004, 5 (23) :295-306
[9]  
IONITA CN, 2007, J DIGIT IMAGING
[10]  
IONITA CN, 2006, TREATMENT CEREBROVAS