Microtensile strain on the corrosion performance of diamond-like carbon coating

被引:4
作者
Lee, Seung-Hwan [1 ]
Kim, Jung-Gu [1 ]
Choi, Heon-Woong [2 ]
Lee, Kwang-Ryeol [2 ]
机构
[1] Sungkyunkwan Univ, Dept Adv Mat Engn, Suwon 440746, South Korea
[2] Korea Inst Sci & Technol, Future Technol Res Div, Seoul 130650, South Korea
关键词
diamond-like carbon; stent; microtensile test; electrochemical impedance spectroscopy; potentiodynamic polarization test;
D O I
10.1002/jbm.a.31597
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hydrogenated diamond-like carbon films (a-C:H DLC) were deposited on STS 304 substrates for the fabrication of vascular stents by means of the r.f. plasmaassisted chemical vapor deposition technique. This study provides reliable and quantitative data for the assessment of the effect of strain on the corrosion performance of DLC-coated systems in the simulated body fluid obtained through electrochemical techniques (potentiodynamic polarization test and electrochemical impedance spectroscopy) and surface analysis (scanning electron microscopy). The electrolyte used in this test was 0.89% NaCl solution at pH 7.4 and 37 degrees C. It was found that the corrosion resistance of the plastically deformed DLC coating was insufficient for use as a protective film in a corrosive body environment. This is due to the increase in the delammation area and degradation of the substrate's corrosion properties with increasing tensile deformation. (c) 2007 Wiley Periodicals, Inc.
引用
收藏
页码:808 / 814
页数:7
相关论文
共 18 条
[1]   The effects of diamond-like carbon coatings on macrophages, fibroblasts and osteoblast-like cells in vitro [J].
Allen, M. ;
Law, F. ;
Rushton, N. .
Clinical Materials, 1994, 17 (01) :1-10
[2]   IN-VITRO STUDIES OF DLC COATINGS WITH SILICON INTERMEDIATE LAYER [J].
BUTTER, R ;
ALLEN, M ;
CHANDRA, L ;
LETTINGTON, AH ;
RUSHTON, N .
DIAMOND AND RELATED MATERIALS, 1995, 4 (5-6) :857-861
[3]   Interfacial studies for improving the adhesion of diamond-like carbon films on steel [J].
Chen, CC ;
Hong, FCN .
APPLIED SURFACE SCIENCE, 2005, 243 (1-4) :296-303
[4]   Fracture behavior of diamond-like carbon films on stainless steel under a micro-tensile test condition [J].
Choi, HW ;
Lee, KR ;
Wang, RZ ;
Oh, KH .
DIAMOND AND RELATED MATERIALS, 2006, 15 (01) :38-43
[5]   Diamond-like carbon coatings as biocompatible materials - an overview [J].
Grill, A .
DIAMOND AND RELATED MATERIALS, 2003, 12 (02) :166-170
[6]   A review of modified DLC coatings for biological applications [J].
Hauert, R .
DIAMOND AND RELATED MATERIALS, 2003, 12 (3-7) :583-589
[7]   An overview on tailored tribological and biological behavior of diamond-like carbon [J].
Hauert, R ;
Müller, U .
DIAMOND AND RELATED MATERIALS, 2003, 12 (02) :171-177
[8]  
Jones MI, 2000, J BIOMED MATER RES, V52, P413, DOI 10.1002/1097-4636(200011)52:2<413::AID-JBM23>3.0.CO
[9]  
2-U
[10]   Tribological behavior of DLC coatings with functionally gradient interfaces [J].
Liu, Y ;
Meletis, EI .
SURFACE & COATINGS TECHNOLOGY, 2002, 153 (2-3) :178-183