Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3

被引:19
作者
Ferraris, Arnaud [1 ]
Jacquet-Lagreze, Matthias [1 ]
Fellahi, Jean-Luc [1 ,2 ]
机构
[1] Hosp Civils Lyon, Hop Cardiovasc & Pneumol Louis Pradel, Serv Anesthesie Reanimat, 59 Blvd Pinel, F-69394 Lyon, France
[2] Univ Claude Bernard Lyon 1, INSERM, U1060, Lyon, France
关键词
Near-infrared spectroscopy; Regional oxygen saturation; EQUANOX monitor; O3; monitor; Cardiac surgery; HEALTHY-VOLUNTEERS; CONTROLLED-TRIAL; O-2; SATURATION; SPECTROSCOPY; OXYGENATION; HYPOXIA; NIRS;
D O I
10.1007/s10877-017-0025-z
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Near-infrared spectroscopy (NIRS) is a continuous and noninvasive technology that measures regional tissue oxygen saturation (rSO(2)). A new 4-wavelength generation of NIRS monitors is now available. We aimed to compare peripheral somatic rSO(2) values given by the 4-wavelength EQUANOX (TM) 7600 device (Nonin Medical Inc., Plymouth, Mn) and O3 (TM) device (Masimo Corporation, Irvine, CA). Twenty adult patients scheduled for conventional elective cardiac surgery with cardiopulmonary bypass over a 4-month period were included after local Ethics Committee approval. For each patient, 2 NIRS sensors (EQUANOX and O3) were placed over the medial part of the forearm. Thirteen couples of measurements were performed at predefined intraoperative time points. We compared 260 couples of absolute intraoperative rSO(2) values. No significant difference was found between both monitors: EQUANOX median rSO(2) 60% (95% CI 57-62) versus O3 median rSO(2) 62% (95% CI 61-64), P = 0.103. Bias was 4.0% and limits of agreement were +/- 26.3%. Significant correlations were evidenced between EQUANOX and O3 rSO(2) absolute values: rho = 0.758 (95% CI 0.701-0.806), P < 0.0001, and rSO(2) percent maximum difference versus baseline: rho = 0.582 (95% CI 0.188-0.815), P = 0.007. While absolute values of rSO(2) given by both devices were equivalent and well correlated, the clinical agreement is probably not acceptable, meaning that EQUANOX and O3 are not interchangeable in routine practice.
引用
收藏
页码:253 / 259
页数:7
相关论文
共 21 条
[1]   Multi-site and multi-depth near-infrared spectroscopy in a model of simulated (central) hypovolemia: lower body negative pressure [J].
Bartels, Sebastiaan A. ;
Bezemer, Rick ;
de Vries, Floris J. Wallis ;
Milstein, Dan M. J. ;
Lima, Alexandre ;
Cherpanath, Thomas G. V. ;
van den Meiracker, Anton H. ;
van Bommel, Jasper ;
Heger, Michal ;
Karemaker, John M. ;
Ince, Can .
INTENSIVE CARE MEDICINE, 2011, 37 (04) :671-677
[2]   Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers [J].
Bickler, Philip E. ;
Feiner, John R. ;
Rollins, Mark D. .
ANESTHESIA AND ANALGESIA, 2013, 117 (04) :813-823
[3]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[4]   Ability of Near Infrared Spectroscopy to Measure Oxygenation in Isolated Upper Extremity Muscle Compartments [J].
Cole, Ashley L. ;
Herman, Richard A., Jr. ;
Heimlich, Jonathan B. ;
Ahsan, Sahir ;
Freedman, Brett A. ;
Shuler, Michael S. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2012, 37A (02) :297-302
[5]   Reproducibility, interchangeability of measures, time to measure stabilization, and reference values of two tissue oximeters in healthy volunteers [J].
Cournoyer, Alexis ;
Denault, Andre ;
Cossette, Sylvie ;
Fortier, Annik ;
Daoust, Raoul ;
Iseppon, Massimiliano ;
Chauny, Jean-Marc ;
Notebaert, Eric .
JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (09)
[6]   Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation A Comparison of Three Cerebral Oximetty Technologies [J].
Davie, Sophie N. ;
Grocott, Hilary P. .
ANESTHESIOLOGY, 2012, 116 (04) :834-840
[7]   A systematic review of cerebral oxygenation-monitoring devices in cardiac surgery [J].
Douds, M. T. ;
Straub, E. J. ;
Kent, A. C. ;
Bistrick, C. H. ;
Sistino, J. J. .
PERFUSION-UK, 2014, 29 (06) :545-552
[8]  
Fellahi J-L, 2014, Ann Fr Anesth Reanim, V33, pe9, DOI 10.1016/j.annfar.2013.11.014
[9]   Dynamic evaluation of near-infrared peripheral oximetry in healthy volunteers: A comparison between INVOS and EQUANOX [J].
Fellahi, Jean-Luc ;
Butin, Geraldine ;
Fischer, Marc-Olivier ;
Zamparini, Guillaume ;
Gerard, Jean-Louis ;
Hanouz, Jean-Luc .
JOURNAL OF CRITICAL CARE, 2013, 28 (05) :881.e1-881.e6
[10]   Use of non-invasive NIRS during a vascular occlusion test to assess dynamic tissue O2 saturation response [J].
Gomez, Hernando ;
Torres, Andres ;
Polanco, Patricio ;
Kim, Hyung Kook ;
Zenker, Sven ;
Puyana, Juan Carlos ;
Pinsky, Michael R. .
INTENSIVE CARE MEDICINE, 2008, 34 (09) :1600-1607