Clinical validation of a GPU-based Monte Carlo dose engine of a commercial treatment planning system for pencil beam scanning proton therapy

被引:28
作者
Fracchiolla, Francesco [1 ]
Engwall, Erik [2 ]
Janson, Martin [2 ]
Tamm, Fredrik [2 ]
Lorentini, Stefano [1 ]
Fellin, Francesco [1 ]
Bertolini, Mattia [1 ]
Algranati, Carlo [1 ]
Righetto, Roberto [1 ]
Farace, Paolo [1 ]
Amichetti, Maurizio [1 ]
Schwarz, Marco [1 ,3 ]
机构
[1] Trento Hosp, Protontherapy Dept, Trento, Italy
[2] RaySearch Labs AB, Stockholm, Sweden
[3] TIFPA Trento Inst Fundamental Phys & Applicat, 14 Via Sommar, I-38123 Povo, TN, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2021年 / 88卷
关键词
Proton therapy; GPU monte carlo; Clinical validation; Dosimetric validation; Calculation time; MULTIPLE-SCATTERING; BENCHMARKING; OPTIMIZATION; ACCURACY; CODE;
D O I
10.1016/j.ejmp.2021.07.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To perform the validation of the GPU-based (Graphical Processing Unit based) proton Monte Carlo (MC) dose engine implemented in a commercial TPS (RayStation 10B) and to report final dose calculation times for clinical cases. Materials and Methods: 440 patients treated at the Proton Therapy Center of Trento, Italy, between 2018 and 2019 were selected for this study. 636 approved plans with 3361 beams computed with the clinically implemented CPU-MC dose engine (version 4.2 and 4.5), were used for the validation of the new algorithm. For each beam, the dose was recalculated using the new GPU-MC dose engine with the initial CPU computation settings and compared to the original CPU-MC dose. Beam dose difference distributions were studied to ensure that the two dose distributions were equal within the expected fluctuations of the MC statistical uncertainty (s) of each computation. Plan dose distributions were compared with respect to the dosimetric indices D98, D50 and D1 of all ROIs defined as targets. A complete assessment of the computation time as a function of s and dose grid voxel size was done. Results: The median over all mean beam dose differences between CPU- and GPU-MC was -0.01% and the median of the corresponding standard deviations was close to (root 2s) both for simulations with an s of 0.5% and 1.0% per beam. This shows that the two dose distributions can be considered equal. All the DVH indices showed an average difference below 0.04%. About half of the plans were computed with 1.0% statistical uncertainty on a 2 mm dose calculation grid, for which the median computation time was 5.2 s. The median computational speed for all plans in the study was 8.4 million protons/second. Conclusion: A validation of a clinical MC algorithm running on GPU was performed on a large pool of patients treated with pencil beam scanning proton therapy. We demonstrated that the differences with the previous CPUbased MC were only due to the intrinsic statistical fluctuations of the MC method, which translated to insignificant differences on plan dose level. The significant increase in dose calculation speed is expected to facilitate new clinical workflows.
引用
收藏
页码:226 / 234
页数:9
相关论文
共 33 条
[1]   Comprehensive clinical commissioning and validation of the RayStation treatment planning system for proton therapy with active scanning and passive treatment techniques [J].
Baeumer, C. ;
Geismar, D. ;
Koska, B. ;
Kramer, P. H. ;
Lambert, J. ;
Lemke, M. ;
Plaude, S. ;
Pschichholz, L. ;
Qamhiyeh, S. ;
Schiemann, A. ;
Timmermann, B. ;
Vermeren, X. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 43 :15-24
[2]  
Beltran Chris, 2016, Int J Part Ther, V3, P312, DOI 10.14338/IJPT-16-00011.1
[3]  
Bohr K. D. V. N, 1948, KGL DANSKE VIDENSKAB, V18
[4]   Commissioning of pencil beam and Monte Carlo dose engines for non-isocentric treatments in scanned proton beam therapy [J].
Carlino, A. ;
Boehlen, T. ;
Vatnitsky, S. ;
Grevillot, L. ;
Osorio, J. ;
Dreindl, R. ;
Palmans, H. ;
Stock, M. ;
Kragl, G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (17)
[5]   A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems [J].
Chang, Chih-Wei ;
Huang, Sheng ;
Harms, Joseph ;
Zhou, Jun ;
Zhang, Rongxiao ;
Dhabaan, Anees ;
Slopsema, Roelf ;
Kang, Minglei ;
Liu, Tian ;
McDonald, Mark ;
Langen, Katja ;
Lin, Liyong .
MEDICAL PHYSICS, 2020, 47 (04) :1545-1557
[6]   Including anatomical variations in robust optimization for head and neck proton therapy can reduce the need of adaptation [J].
Cubillos-Mesias, Macarena ;
Troost, Esther G. C. ;
Lohaus, Fabian ;
Agolli, Linda ;
Rehm, Maximilian ;
Richter, Christian ;
Stuetzer, Kristin .
RADIOTHERAPY AND ONCOLOGY, 2019, 131 :127-134
[7]   A pre-absorber optimization technique for pencil beam scanning proton therapy treatments [J].
Fracchiolla, Francesco ;
Fellin, Francesco ;
Innocenzi, Marta ;
Lipparini, Mirko ;
Lorentini, Stefano ;
Widesott, Lamberto ;
Farace, Paolo ;
Schwarz, Marco .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 :145-152
[8]   Commissioning of GPU-Accelerated Monte Carlo Code FRED for Clinical Applications in Proton Therapy [J].
Gajewski, Jan ;
Garbacz, Magdalena ;
Chang, Chih-Wei ;
Czerska, Katarzyna ;
Durante, Marco ;
Krah, Nils ;
Krzempek, Katarzyna ;
Kopec, Renata ;
Lin, Liyong ;
Mojzeszek, Natalia ;
Patera, Vincenzo ;
Pawlik-Niedzwiecka, Monika ;
Rinaldi, Ilaria ;
Rydygier, Marzena ;
Pluta, Elzbieta ;
Scifoni, Emanuele ;
Skrzypek, Agata ;
Tommasino, Francesco ;
Schiavi, Angelo ;
Rucinski, Antoni .
FRONTIERS IN PHYSICS, 2021, 8
[9]   Validation of a GPU-based Monte Carlo code (gPMC) for proton radiation therapy: clinical cases study [J].
Giantsoudi, Drosoula ;
Schuemann, Jan ;
Jia, Xun ;
Dowdell, Stephen ;
Jiang, Steve ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (06) :2257-2269
[10]  
Goudsmit S, 1940, PHYS REV, V57, P24, DOI 10.1103/PhysRev.57.24