Impact of a Reduced Tube Voltage on CT Angiography and Radiation Dose Results of the PROTECTION I Study

被引:174
作者
Bischoff, Bernhard [1 ]
Hein, Franziska [1 ]
Meyer, Tanja [1 ]
Hadamitzky, Martin [1 ]
Martinoff, Stefan [2 ]
Schoemig, Albert [1 ]
Hausleiter, Joerg [1 ]
机构
[1] Klin TU Munchen, Klin Herz & Kreislauferkrankungen, Munich, Germany
[2] Klin TU Munchen, Deutsch Herzzentrum Munchen, Inst Radiol & Nukl Med, Munich, Germany
关键词
computed tomography; coronary artery disease; tube voltage; MULTISLICE COMPUTED-TOMOGRAPHY; CORONARY-ANGIOGRAPHY; QUANTIFICATION; CARDIOLOGY; RADIOLOGY; COMMITTEE;
D O I
10.1016/j.jcmg.2009.02.015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The aim of this study was to determine the impact of a reduced 100-kV tube voltage on image quality and radiation exposure in a pre-defined subgroup analysis of the international, multicenter radiation dose survey PROTECTION I (Prospective Multicenter Study on RadiaTion Dose Estimates Of Cardiac CT AngIOgraphy I) study. BACKGROUND Cardiac computed tomography angiography (CCTA) has become a frequently used diagnostic tool in clinical practice. Despite continually improving CT technology, there remain concerns regarding the associated radiation exposure. A reduced tube voltage of 100 kV has been proposed as an effective means for dose reduction in nonobese patients. METHODS The study assessed the relevant radiation dose parameters as well as quantitative and qualitative diagnostic image quality data in a subgroup of 321 patients (100 kV: 82 patients; 120 kV: 239 patients), who were scanned at study sites that applied a 100-kV tube voltage in at least 1 patient. Diagnostic image quality was assessed by an experienced CCTA investigator with a 4-point score (1: nondiagnostic to 4: excellent image quality). Effective radiation dose was estimated from the dose-length-product of each CCTA study. RESULTS The use of the 100-kV scan protocol was associated with 53% reduction in CCTA median radiation dose estimates, when compared with the conventional 120-kV scan protocol (p < 0.001). Although image noise significantly increased by 26.3% with 100 kV, signal-as well as contrast-to-noise ratios also increased by 7.9% (p = 0.254) and 10.8% (p = 0.027), respectively. Reduction of tube voltage did not impair diagnostic image quality (median diagnostic score: 3.5 [3.25 to 3.75] vs. 3.5 [3.0 to 3.75] for 100 kV vs. 120 kV; p = 0.22). \ CONCLUSIONS In this nonrandomized PROTECTION I dose survey, reducing the CCTA tube voltage to 100 kV in nonobese patients is associated with a significant reduction in radiation exposure while maintaining diagnostic image quality. Thus, the 100-kV scan technique should be considered for CCTA dose reduction in adequately selected patients. (J Am Coll Cardiol Img 2009;2:940-6) (C) 2009 by the American College of Cardiology Foundation
引用
收藏
页码:940 / 946
页数:7
相关论文
共 14 条
  • [1] MDCT of the coronary arteries: Feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose
    Abada, Hicham T.
    Larchez, Christophe
    Daoud, Beatrice
    Sigal-Cinqualbre, Anne
    Paul, Jean-Francois
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) : S387 - S390
  • [2] [Anonymous], 2004, EUROPEAN GUIDELINES
  • [3] Assessment of coronary artery disease by cardiac computed tomography - A scientific statement from the American Heart Association committee on cardiovascular imaging and intervention, council on cardiovascular radiology and intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology
    Budoff, Matthew J.
    Achenbach, Stephan
    Blumenthal, Roger S.
    Carr, J. Jeffrey
    Goldin, Jonathan G.
    Greenland, Philip
    Guerci, Alan D.
    Lima, Joao A. C.
    Rader, Daniel J.
    Rubin, Geoffrey D.
    Shaw, Leslee J.
    Wiegers, Susan E.
    [J]. CIRCULATION, 2006, 114 (16) : 1761 - 1791
  • [4] Coronary artery stenosis quantification using multislice computed tomography
    Dewey, Marc
    Rutsch, Wofgang
    Schnapauff, Dirk
    Teige, Florian
    Hamm, Bernd
    [J]. INVESTIGATIVE RADIOLOGY, 2007, 42 (02) : 78 - 84
  • [5] Guidelines on the management of stable angina pectoris: executive summary
    Fox, Kim
    Angeles Alonso Garcia, Maria
    Ardissino, Diego
    Buszman, Pawel
    Katowice
    Camici, Paolo G.
    Crea, Filippo
    Daly, Caroline
    De Backer, Guy
    Ghent
    Hjemdahl, Paul
    Lopez-Sendon, Jose
    Marco, Jean
    Morais, Joao
    Leiria
    Pepper, John
    Sechtem, Udo
    Simoons, Maarten
    Thygesen, Kristian
    Priori, Silvia G.
    Blanc, Jean-Jacques
    Budaj, Andrzej
    Camm, John
    Dean, Veronica
    Deckers, Jaap
    Dickstei, Kenneth
    Lekakis, John
    McGregor, Keith
    Metra, Marco
    Morais, Joao
    Osterspey, Ady
    Tamargo, Juan
    Zamorano, Jose L.
    Andreotti, Felicita
    Becher, Harald
    Dietz, Rainer
    Fraser, Alan
    Hernandez Antolin, Rosa Ana
    Huber, Kurt
    Kremastinos, Dimitris T.
    Maseri, Attilio
    Nesser, Hans-Joachim
    Pasierski, Tomasz
    Sigwart, Ulrich
    Tubaro, Marco
    Weis, Michael
    [J]. EUROPEAN HEART JOURNAL, 2006, 27 (11) : 1341 - 1381
  • [6] Radiation dose estimates from cardiac multislice computed tomography in daily practice -: Impact of different scanning protocols on effective dose estimates
    Hausleiter, J
    Meyer, T
    Hadamitzky, M
    Huber, E
    Zankl, M
    Martinoff, S
    Kastrati, A
    Schömig, A
    [J]. CIRCULATION, 2006, 113 (10) : 1305 - 1310
  • [7] Estimated Radiation Dose Associated With Cardiac CT Angiography
    Hausleiter, Joerg
    Meyer, Tanja
    Hermann, Franziska
    Hadamitzky, Martin
    Krebs, Markus
    Gerber, Thomas C.
    McCollough, Cynthia
    Martinoff, Stefan
    Kastrati, Adnan
    Schoemig, Albert
    Achenbach, Stephan
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (05): : 500 - 507
  • [8] ACCF/ACR/SCCT/SCMR/ASNC/NASCI/SCAI/SIR 2006 appropriateness criteria for cardiac computed tomography and cardiac magnetic resonance imaging - A report of the American College of Cardiology Foundation Quality Strategic Directions Committee Appropriateness Criteria Working Group, American College of Radiology, Society of Cardiovascular Computed Tomography, Society for Cardiovascular Magnetic Resonance, American Society of Nuclear Cardiology, North American Society for Cardiac Imaging, Society for Cardiovascular Angiography and Interventions, and Society of Interventional Radiology
    Hendel, Robert C.
    Patel, Manesh R.
    Kramer, Christopher M.
    Poon, Michael
    Hendel, Robert C.
    Brindis, Ralph G.
    Hendel, Robert C.
    Douglas, Pamela S.
    Peterson, Eric D.
    Wolk, Michael J.
    Allen, Joseph M.
    Patel, Manesh R.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (07) : 1475 - 1497
  • [9] Johns H.E., 1983, The Physics of Radiology, V4th ed.
  • [10] Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography - A comparative study with quantitative coronary angiography and intravascular ultrasound
    Leber, AW
    Knez, A
    von Ziegler, F
    Becker, A
    Nikolaou, K
    Paul, S
    Wintersperger, B
    Reiser, M
    Becker, CR
    Steinbeck, G
    Boekstegers, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (01) : 147 - 154