Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates

被引:119
作者
Quinn, Brian [1 ]
Dauer, Zak [1 ]
Pandit-Taskar, Neeta [2 ]
Schoder, Heiko [2 ]
Dauer, Lawrence T. [1 ,2 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, Box 84,1275 York Ave, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiol, 1275 York Ave, New York, NY 10021 USA
来源
BMC MEDICAL IMAGING | 2016年 / 16卷
关键词
PET/CT; CT; Radiation exposure; Effective dose; 18F-FDG; TOMOGRAPHY; EXPOSURE; ADULT; CT;
D O I
10.1186/s12880-016-0143-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Whole body fluorine-18 fluorodeoxyglucose positron emission tomography/computed tomography (PET/CT) is the standard of care in oncologic diagnosis and staging, and patient radiation dose must be well understood to balance exam benefits with the risk from radiation exposure. Although reference PET/CT patient doses are available, the potential for widely varying total dose prompts evaluation of clinic-specific patient dose. The aims of this study were to use exam-specific information to characterize the radiation dosimetry of PET/CT exams that used two different CT techniques for adult oncology patients and evaluate the practicality of employing an exam-specific approach to dose estimation. Methods: Whole body PET/CT scans from two sets of consecutive adult patients were retrospectively reviewed. One set received a PET scan with a standard registration CT and the other a PET scan with a diagnostic quality CT. PET dose was calculated by modifying the standard reference phantoms in OLINDA/EXM 1.1 with patient-specific organ mass. CT dose was calculated using patient-specific data in ImPACT. International Commission on Radiological Protection publication 103 tissue weighting coefficients were used for effective dose. Results: One hundred eighty three adult scans were evaluated (95 men, 88 women). The mean patient-specific effective dose from a mean injected 18F-FDG activity of 450 +/- 32 MBq was 9.0 +/- 1.6 mSv. For all standard PET/CT patients, mean effective mAs was 39 +/- 11 mAs, mean CT effective dose was 5.0 +/- 1.0 mSv and mean total effective dose was 14 +/- 1.3 mSv. For all diagnostic PET/CT patients, mean effective mAs was 120 +/- 51 mAs, mean CT effective dose was 15.4 +/- 5.0 mSv and mean total effective dose was 24.4 +/- 4.3 mSv. The five organs receiving the highest organ equivalent doses in all exams were bladder, heart, brain, liver and lungs. Conclusions: Patient-specific parameters optimize the patient dosimetry utilized in the medical justification of whole body PET/CT referrals and optimization of PET and CT acquisition parameters. Incorporating patient-specific data into dose estimates is a worthwhile effort for characterizing patient dose, and the specific dosimetric information assists in the justification of risk and optimization of PET/CT.
引用
收藏
页数:11
相关论文
共 40 条
  • [1] AAPM, 2011, Size Specific Dose Estimate (SSDE) in Pediatric and Adult Body CT Examination (Task Group 204)
  • [2] [Anonymous], 1996, Ann ICRP, V26, P1
  • [3] [Anonymous], 1975, REPORT TASK GROUP RE, pxix
  • [4] Positron emission tomography/computed tomography-imaging protocols, artifacts, and pitfalls
    Bockisch, A
    Beyer, T
    Antoch, G
    Freudenberg, LS
    Kühl, H
    Debatin, JF
    Müller, SP
    [J]. MOLECULAR IMAGING AND BIOLOGY, 2004, 6 (04) : 188 - 199
  • [5] MIRD Pamphlet No. 21: A Generalized Schema for Radiopharmaceutical Dosimetry-Standardization of Nomenclature
    Bolch, Wesley E.
    Eckerman, Keith F.
    Sgouros, George
    Thomas, Stephen R.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (03) : 477 - 484
  • [6] Estimated radiation risks potentially associated with full-body CT screening
    Brenner, DJ
    Elliston, CD
    [J]. RADIOLOGY, 2004, 232 (03) : 735 - 738
  • [7] Image Wisely: A Campaign to Increase Awareness about Adult Radiation Protection
    Brink, James A.
    Amis, E. Stephen, Jr.
    [J]. RADIOLOGY, 2010, 257 (03) : 601 - 602
  • [8] Brix G, 2005, J NUCL MED, V46, P608
  • [9] Changes in Radiation Dose with Variations in Human Anatomy: Moderately and Severely Obese Adults
    Clark, Landon D.
    Stabin, Michael G.
    Fernald, Michael J.
    Brill, Aaron B.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (06) : 929 - 932
  • [10] Clement C. H., 2008, Annals of the ICRP, V38, P1, DOI 10.1016/j.icrp.2008.08.003