Clinical accuracy and agreement between tympanic and forehead body temperature measurements for screening of patients with COVID-19

被引:6
|
作者
Nia, Hamid Sharif [1 ]
Chong, Pei Pei [2 ]
Huak, Chan Yiong [3 ]
Gorgulu, Ozkan [4 ]
Taghipour, Behzad [5 ]
Froelicher, Erika Sivarajan [6 ,7 ]
Sharif, Saeed Pahlevan [8 ]
Rahmatpour, Pardis [9 ]
机构
[1] Mazandaran Univ Med Sci, Sch Nursing & Midwifery Amol, Sari, Iran
[2] Taylors Univ, Fac Hlth & Med Sci, Sch Biosci, Subang Jaya, Malaysia
[3] Yong Loo Lin Sch Med, Singapore, Singapore
[4] Kirsehir Ahi Evran Univ, Fac Med, Dept Biostat & Med Informat Kirsehir, Kirsehir, Turkey
[5] Mazandaran Univ Med Sci, Imam Khomeinis Hosp Amol, Sari, Iran
[6] Univ Calif San Francisco, Sch Nursing, Dept Physiol Nursing, San Francisco, CA 94143 USA
[7] Univ Calif San Francisco, Sch Med, Dept Epidemiol & Biostat, San Francisco, CA USA
[8] Taylors Univ, Taylors Business Sch, Lakeside Campus, Subang Jaya, Malaysia
[9] Alborz Univ Med Sci, Dept Nursing, Karaj, Iran
关键词
agreement; body temperature; fever; forehead; hospital infection; nursing; SARS-COV-2; screening; thermometer; tympanic; CONCORDANCE CORRELATION-COEFFICIENT; AXILLARY; THERMOMETER; PRECISION;
D O I
10.1111/jocn.16166
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Aim To investigate the accuracy, reliability and agreement between infrared forehead thermometers versus infrared tympanic thermometers temperature, a cross-sectional study was conducted in April 2020. Methods The forehead and tympanic temperatures of 615 subjects were measured simultaneously in three exposed SARS-COV-2 groups at one hospital in Iran, during April 2020. These comparisons were evaluated by Bland-Altman Plot, repeatability, Passing-Bablok regression and Lin's concordance correlation coefficient. The receiver operating characteristic (ROC) analysis was done to describe the discrimination accuracy of a diagnostic test. The study adhered to STROBE checklist for cross-sectional studies. Results A Bland-Altman plot indicated that the limits of agreement between the forehead and tympanic temperature were -0.259 to +0.19 degrees C. Passing-Bablok regression analysis illustrated that the infrared forehead was not linearly related to tympanic temperatures (reference method), with a slope estimate that was significantly different from 1.00. The infrared forehead thermometer showed poor precision and lower accuracy than the tympanic. The forehead temperature readings had 60.0% sensitivity and 44.4% specificity (p > .05) to predict disease. Conclusion According to the results of study, there is no evidence that the assessment of temperature by infrared forehead thermometer could discriminate between the two groups (positive and negative).
引用
收藏
页码:3272 / 3285
页数:14
相关论文
共 50 条
  • [1] Agreement between axillary, tympanic, and mid-forehead body temperature measurements in adult emergency department patients
    Sener, Serkan
    Karcioglu, Ozgur
    Eken, Cenker
    Yaylaci, Serpil
    Ozsarac, Murat
    EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2012, 19 (04) : 252 - 256
  • [2] Covid-19 screening: are forehead temperature measurements during cold outdoor temperatures really helpful?
    Cornelius Dzien
    Wolfgang Halder
    Hannes Winner
    Monika Lechleitner
    Wiener klinische Wochenschrift, 2021, 133 : 331 - 335
  • [3] Covid-19 screening: are forehead temperature measurements during cold outdoor temperatures really helpful?
    Dzien, Cornelius
    Halder, Wolfgang
    Winner, Hannes
    Lechleitner, Monika
    WIENER KLINISCHE WOCHENSCHRIFT, 2021, 133 (7-8) : 331 - 335
  • [4] Limitations of temporal (forehead) temperature readings as a screening method for covid-19
    Fernando Aragon-Vargas, Luis
    PENSAR EN MOVIMIENTO-REVISTA DE CIENCIAS DEL EJERCICIO Y LA SALUD, 2020, 18 (01)
  • [5] Is body temperature mass screening a reliable and safe option for preventing COVID-19 spread?
    Lippi, Giuseppe
    Nocini, Riccardo
    Mattiuzzi, Camilla
    Henry, Brandon Michael
    DIAGNOSIS, 2022, 9 (02) : 195 - 198
  • [6] Clinical accuracy of non-contact forehead infrared thermometer and infrared tympanic thermometer in postoperative adult patients: A comparative study.
    Kameda, Norihiro
    JOURNAL OF PERIOPERATIVE PRACTICE, 2022, 32 (06) : 142 - 148
  • [7] Accuracy of the volume-viscosity swallow test for clinical screening of dysphagia in post COVID-19 patients
    Guillen-Sola, Anna
    Ramirez, Cindry
    Bel-Franquesa, Helena
    Duarte, Esther
    Grillo, Christian
    Duran, Xavier
    Boza, Roser
    CLINICAL NUTRITION ESPEN, 2023, 58 : 295 - 300
  • [8] Clinical Pediatric Screening for COVID-19
    Shayganmehr, Aref
    Dorosti, Abbas Ali
    Saboktakin, Lida
    Abbasi, Madineh
    Khaiatzadeh, Simin
    Khoshmaram, Najibeh
    Emrouzi, Robabeh
    Raeisi, Mohammadreza
    Allahyari, Ali
    IRANIAN JOURNAL OF PEDIATRICS, 2021, 31 (01) : 1 - 7
  • [9] Feasibility and effectiveness of daily temperature screening to detect COVID-19 in a prospective cohort at a large public university
    Facente, Shelley N.
    Hunter, Lauren A.
    Packel, Laura J.
    Li, Yi
    Harte, Anna
    Nicolette, Guy
    McDevitt, Shana
    Petersen, Maya
    Reingold, Arthur L.
    BMC PUBLIC HEALTH, 2021, 21 (01)
  • [10] Feasibility and effectiveness of daily temperature screening to detect COVID-19 in a prospective cohort at a large public university
    Shelley N. Facente
    Lauren A. Hunter
    Laura J. Packel
    Yi Li
    Anna Harte
    Guy Nicolette
    Shana McDevitt
    Maya Petersen
    Arthur L. Reingold
    BMC Public Health, 21