Assessment of corneal curvature using verion optical imaging system: a comparative study

被引:10
作者
Lauschke, Jenny L. [1 ,2 ]
Lawless, Michael [1 ,3 ]
Sutton, Gerard [1 ,3 ]
Roberts, Timothy V. [1 ,3 ]
Hodge, Christopher [1 ]
机构
[1] Vis Eye Inst Chatswood, Sydney, NSW, Australia
[2] Prince Wales Hosp, Dept Ophthalmol, Sydney, NSW, Australia
[3] Univ Sydney, Save Sight Inst, Sydney, NSW, Australia
关键词
keratometry; verion optical imaging; corneal curvature; iolmaster; topcon; pentacam; LENS POWER CALCULATION; AUTOMATED KERATOMETRY; AXIS LOCATION; NORMAL EYES; TOPOGRAPHY; ASTIGMATISM; BIOMETRY; DEVICE;
D O I
10.1111/ceo.12687
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
BackgroundKeratometry is a critical determinant of the postoperative refractive outcome. This study evaluates the comparability of keratometry measurements obtained using the Verion Optical Imaging System with devices used in current clinical practice. Further, it determines the interobserver reliability of the Verion Optical Imaging System. DesignRetrospective analysis of patient data ParticipantsKeratometric data was obtained from 100 patients presenting for pre-operative evaluation. Furthermore, repeatability and reproducibility were assessed using 15 healthy volunteers. MethodsCorneal curvature values acquired by the Verion Optical Imaging System were compared with existing keratometry devices including autokeratometry, partial coherence interferometry (IOLMaster) and Scheimpflug corneal topography (Pentacam). Main Outcome MeasuresDirect comparison of corneal power, cylinder, axis of astigmatism and vector analysis were performed using correlation and Bland-Altman analyses. Reproducibility and reliability of the device were assessed using within-subject standard deviation and intraclass correlation coefficients for experienced and inexperienced technicians. ResultsThere was no statistically significant difference between Verion values and those obtained by other methods in relation to mean keratometry, corneal astigmatism, steep meridian and vector analyses (P>0.05). Bland-Altman plots showed narrow limits of agreement for keratometry and astigmatism and wider agreement for steep meridian. Intraoperator reliability for both experienced and inexperienced operators and interoperator reproducibility showed no statistically significant differences between values. ConclusionMeasurements show no significant difference from those obtained by the other devices. This suggests the Verion system is en par with instruments used in current clinical practice. The Verion Optical Imaging System produces repeatable data with no difference related to operator experience.
引用
收藏
页码:369 / 376
页数:8
相关论文
共 22 条
[1]   Evaluation of a Risk Factor Scoring System for Corneal Ectasia After LASIK in Eyes With Normal Topography [J].
Binder, Perry S. ;
Trattler, William B. .
JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (04) :241-250
[2]   Progression of Keratoconus by Longitudinal Assessment with Corneal Topography [J].
Choi, Jin A. ;
Kim, Man-Soo .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (02) :927-935
[3]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[4]   Intraocular Lens Power Measured by Partial Coherence Interferometry [J].
Hsieh, Yi-Ting ;
Wang, I-Jong .
OPTOMETRY AND VISION SCIENCE, 2012, 89 (12) :1697-1701
[5]   An evaluation of keratometry in 6-year-old children [J].
Huynh, Son C. ;
Mai, Tung Q. ;
Kifley, Annette ;
Wang, Jie Jin ;
Rose, Kathiyn A. ;
Mitchell, Paul .
CORNEA, 2006, 25 (04) :383-387
[6]   Biommetry and intraocular lens power calculation results with a new optical biometry device: Comparison with the gold standard [J].
Kaswin, Godefroy ;
Rousseau, Antoine ;
Mgarrech, Mohamed ;
Barreau, Emmanuel ;
Labetoulle, Marc .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (04) :593-600
[7]   Comparison of corneal power, corneal astigmatism, and axis location in normal eyes obtained from an autokeratometer and a corneal topographer [J].
Kobashi, Hidenaga ;
Kamiya, Kazutaka ;
Igarashi, Akihito ;
Ishii, Rie ;
Sato, Nobuyuki ;
Wang, Guoqin ;
Shimizu, Kimiya .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04) :648-654
[8]   Comparison of Automated and Partial Coherence Keratometry and Resulting Choice of Toric IOL [J].
Magar, Jit B. Ale ;
Cunningham, Frank ;
Brian, Garry .
OPTOMETRY AND VISION SCIENCE, 2013, 90 (04) :385-391
[9]   Keratometry Evaluations With the Pentacam High Resolution in Comparison With the Automated Keratometry and Conventional Corneal Topography [J].
Modis, Laszlo, Jr. ;
Szalai, Eszter ;
Kolozsvari, Bence ;
Nemeth, Gabor ;
Vajas, Attila ;
Berta, Andras .
CORNEA, 2012, 31 (01) :36-41
[10]   Sources of error in intraocular lens power calculation [J].
Norrby, Sverker .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (03) :368-376