Activity determination without tracers in high-resolution alpha-particle spectrometry

被引:6
作者
Pollanen, Roy [1 ]
Siiskonen, Teemu [1 ]
Ihantola, Sakari [1 ]
Toivonen, Harri [1 ]
Pelikan, Andreas [2 ]
Inn, Kenneth [3 ]
La Rosa, Jerome [3 ]
Bene, Balazs [3 ]
机构
[1] STUK Radiat & Nucl Safety Author, Helsinki 00881, Finland
[2] Dienstleitungen Automat Datenverarbeitung & Infor, Osterreich, Austria
[3] NIST, Ionizing Radiat Div, Radioact Grp, Gaithersburg, MD 20899 USA
关键词
Alpha spectrometry; Spectrum unfolding; Tracer; Plutonium; SIMULATION; CODE;
D O I
10.1007/s10967-011-1379-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-resolution alpha spectrometry was applied for the activity determination of 10 reference sources containing different amounts of (238)Pu, (239)Pu, (240)Pu and (242)Pu. They were analyzed as blind sources using a novel spectrum analysis tool ADAM. The information needed in the spectrum unfolding was taken only from the spectrum under investigation, and no tracers were applied. Therefore, a Monte Carlo program AASI was used to compute geometrical detection efficiency of the measurement setup. All reported activities corresponded to those of the reference sources within expanded uncertainty. The developed tools can be used for the activity determination in nondestructive alpha spectrometry or when the radionuclide composition does not change during the sample processing.
引用
收藏
页码:551 / 555
页数:5
相关论文
共 12 条
[1]   SOLANG: A user-friendly code to calculate the geometry factor using Monte Carlo simulations. Application to alpha-particle spectrometry [J].
Cornejo Diaz, N. A. ;
Martin Sanchez, A. ;
de la Torre Perez, J. .
APPLIED RADIATION AND ISOTOPES, 2011, 69 (05) :822-824
[2]   Advanced alpha spectrum analysis based on the fitting and covariance analysis of dependent variables [J].
Ihantola, S. ;
Pelikan, A. ;
Pollanen, R. ;
Toivonen, H. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 656 (01) :55-60
[3]   ELECTRODEPOSITION OF PLUTONIUM AND AMERICIUM FOR HIGH-RESOLUTION ALPHA SPECTROMETRY [J].
KRESSIN, IK .
ANALYTICAL CHEMISTRY, 1977, 49 (06) :842-846
[4]   High-resolution alpha spectrometry from thick sources [J].
Pöllänen, R ;
Siiskonen, T ;
Vesterbacka, P .
RADIATION MEASUREMENTS, 2005, 39 (05) :565-568
[5]  
POLLANEN R, 2011, DETERMINATION UNPUB
[6]  
Pollanen R., 2005, P INT C MON ASS UNC
[7]  
Pollanen R., 2005, P 2 INT C RAD ENV 2, P193
[8]  
Pommé S, 2008, ACTA CHIM SLOV, V55, P111
[9]   Advanced simulation code for alpha spectrometry [J].
Siiskonen, T ;
Pöllänen, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 550 (1-2) :425-434
[10]   Simulation of alpha particle spectra from aerosol samples [J].
Siiskonen, T ;
Pöllänen, R .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (06) :947-953