Residual astigmatism produced by toric intraocular lens rotation

被引:109
|
作者
Felipe, Adelina [1 ,2 ]
Artigas, Jose M. [1 ,2 ]
Diez-Ajenjo, Amparo [1 ,2 ]
Garcia-Domene, Carmen [2 ]
Alcocer, Pablo [2 ]
机构
[1] Univ Valencia, Fac Fis, Dept Opt, E-46100 Burjassot, Valencia, Spain
[2] Univ Valencia, Fac Fis, Fdn Oftalmol Mediterraneo, E-46100 Burjassot, Valencia, Spain
关键词
CORRECTING ASTIGMATISM; CATARACT PATIENTS; STABILITY; IMPLANTATION; EXPERIENCE; CENTRATION; EYES;
D O I
10.1016/j.jcrs.2011.04.036
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To analyze changes in the eye's refractive properties when a toric intraocular lens (IOL) rotates. SETTING: Fundacion Oftalmologica del Mediterraneo, Valencia, Spain. DESIGN: Experimental study. METHODS: The matrix definition of astigmatism was used in this theoretical study and compared with another vector representation. Two methods were compared: (1) The cylinder, C, resulting from the addition of 2 cylinders C(1) and C(2) whose axes form an angle a, is obtained by the addition of 2 vectors of values C(1) and C(2) forming an angle 2a; (2) the power matrix, F, of a thin astigmatic dioptric system that decomposes naturally into 3 orthogonal components: the purely spherical part F(nes), the ortho-astigmatism F(or), and oblique astigmatism F(ob). RESULTS: The residual cylinder was one third of the corneal astigmatism when a toric IOL rotated +/- 10 degrees when the cylinder values for the cornea (C(1)) and IOL (C(2)) were equal. Nevertheless, in most cases C(1) is greater than C(2); therefore, the residual astigmatism did not change noticeably with small rotations. The angle of rotation, b, which annuls the astigmatism correction, could be obtained from the following: cos(pi + 2b) = -r/2, with r being the ratio between the IOL and cornea cylinders. CONCLUSIONS: The 2 methods gave equivalent results. When the IOL cylinder had a value different from that of the corneal astigmatism, a better choice would be a lower, rather than higher, cylinder value to reduce residual astigmatism. In general, toric IOL rotations less than 10 degrees changed the eye's refraction less than 0.50 diopter. Thus, small axis rotations are not an obstacle for satisfactory astigmatism correction with toric IOLs.
引用
收藏
页码:1895 / 1901
页数:7
相关论文
共 50 条
  • [1] Toric intraocular lens in asymmetric astigmatism
    Kwitko, Sergio
    Marafon, Samara Barbara
    Stolz, Andressa Prestes
    INTERNATIONAL OPHTHALMOLOGY, 2020, 40 (05) : 1291 - 1298
  • [2] Transscleral suture fixation following recurrent toric intraocular lens rotation
    Arjmand, Parnian
    Chan, Toby Y. B.
    Ahmed, Iqbal Ike K.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) : 912 - 917
  • [3] Software-based assessment of postoperative rotation of toric intraocular lens
    Shah, Gauri D.
    Praveen, Mamidipudi R.
    Vasavada, Abhay R.
    Rampal, Naveen V.
    Vasavada, Viraj A.
    Asnani, Priyadarshi K.
    Pandita, Deepa. K.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (03) : 413 - 418
  • [4] Novel custom designed toric piggyback intraocular lens for the correction of residual postoperative astigmatism
    Chatterjee, Debapriya
    Mittal, Vikas
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 2247 - 2250
  • [5] Objective analysis of toric intraocular lens rotation and centration
    Wolffsohn, James S.
    Buckhurst, Phillip J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (05) : 778 - 782
  • [6] Iris-fixated toric phakic intraocular lens for myopic astigmatism
    Munoz, Gonzalo
    Cardoner, Antoni
    Albarran-Diego, Cesar
    Ferrer-Blasco, Teresa
    Belda-Salmeron, Lurdes
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07) : 1166 - 1175
  • [7] Optimized keratometry and total corneal astigmatism for toric intraocular lens calculation
    Savini, Giacomo
    Naeser, Kristian
    Schiano-Lomoriello, Domenico
    Ducoli, Pietro
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (09) : 1140 - 1148
  • [8] Cataract surgery with toric intraocular lens for correction of high corneal astigmatism
    Iovieno, Alfonso
    Yeung, Sonia N.
    Lichtinger, Alejandro
    Alangh, Manreet
    Slomovic, Allan R.
    Rootman, David S.
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2013, 48 (04): : 246 - 250
  • [9] Effects of the toric intraocular lens on correction of preexisting corneal astigmatism
    Sasaki, Hiroshi
    Yoshida, Motoaki
    Manabe, Shin-ichi
    Yoshimura, Koichi
    Hayashi, Ken
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2012, 56 (05) : 445 - 452
  • [10] AcrySof toric intraocular lens for post-keratoplasty astigmatism
    Gupta, Nishant
    Ram, Jagat
    Chaudhary, Manish
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2012, 60 (03) : 213 - 215