Axial and Rotational Malreduction (Golf Club Deformity) in Distal Femur Fractures

被引:9
作者
Lowe, Jason [1 ,2 ]
Alhandi, Ali [3 ]
Manoharan, Aditya [1 ,2 ]
Ouellette, Elizabeth [4 ]
Kaimrajh, David [5 ]
Milne, Edward [5 ]
Latta, Loren [5 ,6 ]
机构
[1] Univ Arizona, Dept Orthopaed Surg, 1501 N Campbell,Room 8401, Tucson, AZ 85724 USA
[2] Banner Univ, Med Ctr, Dept Orthopaed Surg, Tucson, AZ USA
[3] King Abdul Aziz Med City, Riyadh, Saudi Arabia
[4] Baptist Hlth South Florida, Miami Orthoped & Sports Med Grp, Coral Gables, FL USA
[5] Mt Sinai, Max Biedermann Inst Biomech, Miami Beach, FL USA
[6] Univ Miami, Dept Orthopaed, Miami, FL USA
关键词
golf club deformity; distal femur; malreduction; INVASIVE STABILIZATION SYSTEM; PLATES; FIXATION; LISS;
D O I
10.1097/BOT.0000000000002373
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To describe malreduction of supracondylar distal femur fractures stabilized with lateral femoral locking plates and determine whether a mismatch in axial lateral distal femur anatomy and lateral distal femoral plate design contributes to supracondylar distal femoral fracture malreduction. Materials and Methods: OTA/AO 33A were simulated in 7 cadaver femurs and fixed with a lateral distal locking femoral plate placed flush to the lateral femoral condyle (group 1). In group 2, the anterior flange of the plate was externally rotated 10 degrees in relation to the lateral condyle. A motion capture system measured translation and rotation of the articular segment as shaft screws were applied, reducing plate to femoral diaphysis. Articular segment movement was compared between groups using paired Student t test, P < 0.05. A large database of 3D scans of 800 femurs was used to define the relationship of the lateral femoral condyle to the lateral cortical surface of the human femur. Results: Malreduction was observed with anatomic plate application results from medial translation (17 mm) and external rotation (12.2 degrees) (group 1). Modifying plate geometry to match lateral femur anatomy (group 2) improved medial translation by 46% and external rotation by 80%. An analysis of the shape of the 800 distal femurs showed that the average posterior anterior inclination is 16.5 degrees. Conclusions: Anatomic application of distal femoral plates results in significant malreduction. Modifying the plate design to accommodate 10 degrees slope of lateral distal femur results in partial correction of deformity. Future studies should investigate other means of correcting the malalignment, especially considering the 16.5-degree posterior anterior inclination of the condyle.
引用
收藏
页码:515 / 518
页数:4
相关论文
共 12 条
[1]   Analysis of the fit of modern pre-contoured distal femur plates: Expect an imperfect contour [J].
Bedard, Julia C. ;
Tanner, Stephanie ;
Cameron, James ;
Jeray, Kyle J. ;
Adams, John D. , Jr. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2020, 51 (03) :719-722
[2]   Distal Femur Locking Plates Fit Poorly Before and After Total Knee Arthroplasty [J].
Campbell, Sean T. ;
Bosch, Liam C. ;
Swinford, Steven ;
Amanatullah, Derek F. ;
Bishop, Julius A. ;
Gardner, Michael J. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2019, 33 (05) :239-243
[3]   Pitfalls in the Application of Distal Femur Plates for Fractures [J].
Collinge, Cory A. ;
Gardner, Michael J. ;
Crist, Brett D. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (11) :695-706
[4]   Locking Plates for Distal Femur Fractures: Is There a Problem With Fracture Healing? [J].
Henderson, Christopher E. ;
Kuhl, Lori L. ;
Fitzpatrick, Daniel C. ;
Marsh, J. L. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 :S8-S14
[5]   Avoiding malunion with 95° fixed-angle distal femoral implants [J].
Karunakar, MA ;
Kellam, JF .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (07) :443-445
[6]   Malalignment after minimally invasive plate osteosynthesis in distal femoral fractures [J].
Kim, Joon-Woo ;
Oh, Chang-Wug ;
Oh, Jong-Keon ;
Park, Il-Hyung ;
Kyung, Hee-Soo ;
Park, Kyeong-Hyeon ;
Yoon, Seong-Dae ;
Kim, Seong-Min .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (03) :751-757
[7]   Distal femoral fixation: A biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate [J].
Koval, KJ ;
Hoehl, JJ ;
Kummer, FJ ;
Simon, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (07) :521-524
[8]   Distal femoral fracture fixation utilizing the Less Invasive Stabilization System (LISS): The technique and early results [J].
Kregor, PJ ;
Stannard, J ;
Zlowodzki, M ;
Cole, PA ;
Alonso, J .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :32-47
[9]   Treatment of distal femur fractures using the Less Invasive Stabilization System - Surgical experience and early clinical results in 103 fractures [J].
Kregor, PJ ;
Stannard, JA ;
Zlowodzki, M ;
Cole, PA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :509-520
[10]   Use of the Less Invasive Stabilization System (LISS) in patients with distal femoral (AO33) fractures:: a prospective multicenter study [J].
Schütz, M ;
Müller, M ;
Regazzoni, P ;
Höntzsch, D ;
Krettek, C ;
Van der Werken, C ;
Haas, N .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (02) :102-108