Validation of an HPLC/MS/MS method with isotopic dilution for quantitative determination of trans, trans-muconic acid in urine samples of workers exposed to low benzene concentrations

被引:29
|
作者
Tranfo, Giovanna [1 ]
Paci, Enrico [1 ]
Sisto, Renata [1 ]
Pigini, Daniela [1 ]
机构
[1] Inst Occupat Safety & Prevent ISPESL, I-00141 Rome, Italy
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2008年 / 867卷 / 01期
关键词
biological monitoring; trans; trans-muconic acid; HPLC/MS/MS; method validation; oil refinery workers;
D O I
10.1016/j.jchromb.2008.03.004
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Urinary trans,trans-muconic acid (t,t-MA), a biomarker of benzene exposure, is usually determined by HPLC methods with detection by either UV or, more recently, electrospray tandem mass spectrometry. However, not all these methods have been fully validated for quantitative analysis. This paper presents an HPLC/MS/MS method for reliable quantitative determination of t,t-MA that uses a commercial deuterium-labeled isotope as internal standard; the matrix effect has been evaluated and LOD is 0.22 mu g/L. We used this method to test 200 urine samples, 175 of them collected at end-of-shift from workers in an oil refinery. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:26 / 31