What is the optimal paddle force during paediatric external defibrillation?

被引:8
作者
Deakin, CD
Bennetts, SH
Petley, GW
Clewlow, F
机构
[1] Southampton Univ Hosp, NHS Trust, Shackleton Dept Anaesthet, Southampton SO16 6YD, Hants, England
[2] Southampton Univ Hosp, NHS Trust, Dept Med Phys & Bioengn, Southampton SO16 6YD, Hants, England
关键词
transthoracic impedance; defibrillation; paediatric; infant; paddle; force;
D O I
10.1016/S0300-9572(03)00173-4
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Transthoracic impedance (TTI) is a major determinant of transmyocardial current flow, and therefore, the success of defibrillation. European Resuscitation Council (ERC) paediatric guidelines recommend that 'firm' paddle force should be applied to the paddles during defibrillation. No study has yet established the optimal paddle force required to minimise TTI in children of different ages. Methods: Eighty patients aged 10 weeks to 17 yrs undergoing general anaesthesia for routine surgery were studied. Using defibrillation paddles placed in an anterior-apical position, TTI (Q) was measured for increasing values of force from 0.5 kgf (baseline) to 6.5 kgf. The optimal force, the force to achieve 95% of the overall reduction in TTI, was then determined. According to current guidelines, paediatric paddles (surface area 16 cm(2)) were used for infants ( :! 10 kg) and adult paddles (82 cm(2)) for older children. Optimal force was then calculated for infants less than or equal to 10 kg, children > 10 kg and less than or equal to 8 yrs and children 9-17 yrs age. Results: Increasing paddle force from 0.5 kgf progressively decreased TTI. Optimal force using paediatric paddles was 2.9 kgf in infants. Optimal force using adult paddles was 5.1 kgf in children > 10 kg but less than or equal to 8 yrs and 5.3 kgf in children aged 9-17 yrs. Conclusions: Force is an important determinant of TTI and therefore, outcome of defibrillation. It is recommended that a minimum of 3 kgf be applied to paddles when defibrillating infants with paediatric paddles, and a minimum of 5 kgf be applied to all older children when adult paddles are used. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:83 / 88
页数:6
相关论文
共 14 条
[1]  
[Anonymous], 2000, RESUSCITATION, V46, P109
[2]  
ATKINS DL, 1988, PEDIATRICS, V82, P914
[3]   Active compression cardioversion for refractory atrial fibrillation [J].
Cohen, TJ ;
Ibrahim, B ;
Denier, D ;
Haji, A ;
Quan, WL .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (03) :354-&
[4]   ACTIVE COMPRESSION-DECOMPRESSION DEFIBRILLATION PROVIDES EFFECTIVE DEFIBRILLATION DURING CARDIOPULMONARY-RESUSCITATION [J].
COHEN, TJ ;
NOUBANI, H ;
GOLDNER, BG ;
TUCKER, KJ ;
QUAN, WL ;
TRAZZERA, S ;
STERN, G .
AMERICAN HEART JOURNAL, 1995, 130 (01) :186-187
[5]  
DALZELL GWN, 1991, BRIT HEART J, V65, P311
[6]   ELECTRODE PAD SIZE, TRANS-THORACIC IMPEDANCE AND SUCCESS OF EXTERNAL VENTRICULAR DEFIBRILLATION [J].
DALZELL, GWN ;
CUNNINGHAM, SR ;
ANDERSON, J ;
ADGEY, AAJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (12) :741-744
[7]   A comparison of transthoracic impedance using standard defibrillation paddles and self-adhesive defibrillation pads [J].
Deakin, CD ;
McLaren, RM ;
Petley, GW ;
Clewlow, F ;
Dalrymple-Hay, MJR .
RESUSCITATION, 1998, 39 (1-2) :43-46
[8]   Determining the optimal paddle fore for external defibrillation [J].
Deakin, CD ;
Sado, DM ;
Petley, GW ;
Clewlow, F .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (07) :812-+
[9]   TRANS-THORACIC RESISTANCE IN HUMAN DEFIBRILLATION - INFLUENCE OF BODY-WEIGHT, CHEST SIZE, SERIAL SHOCKS, PADDLE SIZE AND PADDLE CONTACT PRESSURE [J].
KERBER, RE ;
GRAYZEL, J ;
HOYT, R ;
MARCUS, M ;
KENNEDY, J .
CIRCULATION, 1981, 63 (03) :676-682
[10]   ADVANCE PREDICTION OF TRANS-THORACIC IMPEDANCE IN HUMAN DEFIBRILLATION AND CARDIOVERSION - IMPORTANCE OF IMPEDANCE IN DETERMINING THE SUCCESS OF LOW-ENERGY SHOCKS [J].
KERBER, RE ;
KOUBA, C ;
MARTINS, J ;
KELLY, K ;
LOW, R ;
HOYT, R ;
FERGUSON, D ;
BAILEY, L ;
BENNETT, P ;
CHARBONNIER, F .
CIRCULATION, 1984, 70 (02) :303-308