Pulmonary Perfusion Scans Following Transcatheter Patent Ductus Arteriosus Closure Using the Amplatzer Devices

被引:17
作者
Kharouf, Rami [1 ]
Heitschmidt, Mary [1 ]
Hijazi, Ziyad M. [1 ]
机构
[1] Rush Univ, Med Ctr, Rush Ctr Congenital & Struct Heart Dis, Chicago, IL 60612 USA
关键词
patent ductus arteriosus; lung perfusion; transcatheter closure; Amplatzer device; LUNG PERFUSION; COIL OCCLUSION; BLOOD-FLOW; CATHETER CLOSURE; GIANTURCO COILS; OCCLUDER DEVICE; OBSTRUCTION; EXPERIENCE; IMMEDIATE; INFANTS;
D O I
10.1002/ccd.22917
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Obstruction to flow in the left pulmonary artery (LPA) is a well-known complication after transcatheter device closure of patent ductus arteriosus (PDA). This complication has been studied for different devices using lung perfusion radionuclide scintigraphy (LPRS), but not for Amplatzer devices. This study was performed to evaluate the effect of such devices on lung perfusion using LPRS. Methods: This is a retrospective study that looked at all patients who had PDA closure using different Amplatzer devices at our center between July 1999 and January 2007. All patients underwent LPRS within 24 hr of the procedure. We compared LPRS with other hemodynamic data obtained by cardiac catheterization and echocardiography. Results are presented as mean +/- SD or median and ranges. Results: A total of 70 patients had PDA closure using an Amplatzer device; median age was 1.8 years (4 months to 75 years) and median weight was 12 kg (5-112 Kg). Nine patients had associated cardiac anomalies. Sixty eight patients had available LPRS. The mean percent of left lung perfusion (LLP) was 42.7% (+/-6.7%). Excluding patients with pre-existing LPA stenosis, 17% had abnormally decreased LLP. On hemodynamic measurements, 62 patients had available direct pressure measurements following PDA closure. None had significant increase. No correlation was found with echocardiographic data. Conclusion: PDA closure with Amplatzer family of devices is associated with a relatively significant risk of decreased perfusion to the left lung, mostly mild abnormalities. Comparison with catheterization and echocardiographic measurements showed lack of correlation with LPRS findings. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:664 / 670
页数:7
相关论文
共 50 条
  • [41] Thrombocytopenia associated with transcatheter closure of giant patent ductus arteriosus
    Liu, Lei
    Yue, Peng
    Zheng, Xiaolan
    Liao, Hongyu
    Li, Jiawen
    Hua, Yimin
    Li, Yifei
    Zhou, Kaiyu
    CARDIOLOGY IN THE YOUNG, 2020, 30 (12) : 1896 - 1901
  • [42] Overview of transcatheter patent ductus arteriosus closure in preterm infants
    Almeida-Jones, Myriam
    Tang, Nai Yu
    Reddy, Aneela
    Zahn, Evan
    CONGENITAL HEART DISEASE, 2019, 14 (01) : 60 - 64
  • [43] Transcatheter closure of patent ductus arteriosus using Gianturco coils in adolescents and adults
    Wang, JK
    Liau, CS
    Huang, JJ
    Hsu, KL
    Lo, PH
    Hung, JS
    Wu, MH
    Lee, YT
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2002, 55 (04) : 513 - 518
  • [44] Lung perfusion studies after percutaneous closure of patent ductus arteriosus using the Amplatzer Duct Occluder in children
    Rafael Parra-Bravo, Jose
    Apolonio-Martinez, Adriana
    de Jesus Estrada-Loza, Maria
    Gracia Beirana-Palencia, Luisa
    Ivan Ramirez-Portillo, Cesar
    ARCHIVOS DE CARDIOLOGIA DE MEXICO, 2015, 85 (03): : 188 - 194
  • [45] A hybrid approach for patent ductus arteriosus closure using the Amplatzer Duct Occluder
    Gu, Qun
    Ma, Luyao
    Gu, Haitao
    Meng, Yaling
    Shen, Chen
    Kong, Xiangyin
    Shao, Yongfeng
    Zhang, Shijiang
    JOURNAL OF CARDIAC SURGERY, 2018, 33 (10) : 684 - 687
  • [46] Simultaneous Transcatheter Closure of Intralobar Pulmonary Sequestration and Patent Ductus Arteriosus in an Infant
    Zhi-Wei Zhang
    Yu-Mei Xie
    Yu-Fen Li
    Pediatric Cardiology, 2010, 31 : 159 - 160
  • [47] Patent ductus arteriosus with persistent pulmonary artery hypertension after transcatheter closure
    Feng, Jianqi
    Kong, Xiangqing
    Sheng, Yanhui
    Yang, Rong
    THERAPEUTICS AND CLINICAL RISK MANAGEMENT, 2016, 12 : 1609 - 1613
  • [48] Transcatheter closure of the patent ductus arteriosus in adults using the Gianturco coil
    Ing, FF
    Mullins, CE
    Rose, M
    Shapir, Y
    Bierman, FZ
    CLINICAL CARDIOLOGY, 1996, 19 (11) : 875 - 879
  • [49] Complications after transcatheter closure of patent ductus arteriosus
    Jang, Gi Young
    Son, Chang Sung
    Lee, Joo Won
    Lee, Jae Young
    Kim, Soo Jin
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2007, 22 (03) : 484 - 490
  • [50] DuctOcclud® -: Implant for transcatheter closure of patent Ductus Arteriosus
    Moore, JW
    Boosfeld, C
    BIOMEDIZINISCHE TECHNIK, 1999, 44 (03): : 46 - 51