Pulmonary Perfusion Scans Following Transcatheter Patent Ductus Arteriosus Closure Using the Amplatzer Devices

被引:17
作者
Kharouf, Rami [1 ]
Heitschmidt, Mary [1 ]
Hijazi, Ziyad M. [1 ]
机构
[1] Rush Univ, Med Ctr, Rush Ctr Congenital & Struct Heart Dis, Chicago, IL 60612 USA
关键词
patent ductus arteriosus; lung perfusion; transcatheter closure; Amplatzer device; LUNG PERFUSION; COIL OCCLUSION; BLOOD-FLOW; CATHETER CLOSURE; GIANTURCO COILS; OCCLUDER DEVICE; OBSTRUCTION; EXPERIENCE; IMMEDIATE; INFANTS;
D O I
10.1002/ccd.22917
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Obstruction to flow in the left pulmonary artery (LPA) is a well-known complication after transcatheter device closure of patent ductus arteriosus (PDA). This complication has been studied for different devices using lung perfusion radionuclide scintigraphy (LPRS), but not for Amplatzer devices. This study was performed to evaluate the effect of such devices on lung perfusion using LPRS. Methods: This is a retrospective study that looked at all patients who had PDA closure using different Amplatzer devices at our center between July 1999 and January 2007. All patients underwent LPRS within 24 hr of the procedure. We compared LPRS with other hemodynamic data obtained by cardiac catheterization and echocardiography. Results are presented as mean +/- SD or median and ranges. Results: A total of 70 patients had PDA closure using an Amplatzer device; median age was 1.8 years (4 months to 75 years) and median weight was 12 kg (5-112 Kg). Nine patients had associated cardiac anomalies. Sixty eight patients had available LPRS. The mean percent of left lung perfusion (LLP) was 42.7% (+/-6.7%). Excluding patients with pre-existing LPA stenosis, 17% had abnormally decreased LLP. On hemodynamic measurements, 62 patients had available direct pressure measurements following PDA closure. None had significant increase. No correlation was found with echocardiographic data. Conclusion: PDA closure with Amplatzer family of devices is associated with a relatively significant risk of decreased perfusion to the left lung, mostly mild abnormalities. Comparison with catheterization and echocardiographic measurements showed lack of correlation with LPRS findings. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:664 / 670
页数:7
相关论文
共 50 条
  • [21] Transcatheter closure of the patent ductus arteriosus
    Grifka, RG
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (04) : 554 - 570
  • [22] Patent ductus arteriosus closure using an Amplatzer™ ventricular septal defect closure device
    Fernando, Rajeev
    Koranne, Ketan
    Loyalka, Pranav
    Kar, Biswajit
    Gregoric, Igor
    EXPERIMENTAL & CLINICAL CARDIOLOGY, 2013, 18 (01): : E50 - E54
  • [23] Is the new Occlutech duct occluder an appropriate device for transcatheter closure of patent ductus arteriosus?
    Godart, Francois
    Houeijeh, Ali
    Domanski, Olivia
    Guillaume, Marie-Paule
    Brard, Melanie
    Lucron, Hugues
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2018, 261 : 54 - 57
  • [24] Percutaneous Closure of Patent Ductus Arteriosus in Adults Using Different Devices
    Gamboa, Ricardo
    Rios-Mendez, Raul E.
    Mollon, Francisco P.
    Arroyo, Graciela M.
    Gutierrez, Diego F.
    REVISTA ESPANOLA DE CARDIOLOGIA, 2010, 63 (06): : 726 - 729
  • [25] Utility of the Amplatzer Vascular Plug Type II™ for transcatheter closure of patent ductus arteriosus in adults
    Blount, Thomas
    Yeung, Michael
    Frantz, Elman
    PROGRESS IN PEDIATRIC CARDIOLOGY, 2019, 53 : 47 - 50
  • [26] Percutaneous Closure of Severe Calcified Patent Ductus Arteriosus With Two Amplatzer Devices
    Takafuji, Hiroya
    Muraishi, Makio
    Obunai, Kotaro
    JOURNAL OF INVASIVE CARDIOLOGY, 2023, 35 (07)
  • [27] Transcatheter closure of patent ductus arteriosus and ductus in children
    Irdem, Ahmet
    Kervancioglu, Mehmet
    Baspinar, Osman
    Kilinc, Metin
    EUROPEAN JOURNAL OF THERAPEUTICS, 2012, 18 (03): : 132 - 138
  • [28] Transcatheter closure of persistent ductus arteriosus in infants using the Amplatzer duct occluder
    Fischer, G
    Stieh, J
    Uebing, A
    Grabitz, R
    Kramer, HH
    HEART, 2001, 86 (04) : 444 - 447
  • [29] Transcatheter closure of a 22 mm patent ductus arteriosus with an amplatzer atrial septal occluder
    Spies, C
    Ujivari, F
    Schräder, R
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (03) : 352 - 355
  • [30] Transcatheter Closure of a Patent Ductus Arteriosus Using a Piccolo Duct Occluder
    Deb, Pranjit
    Benerjee, Anindya
    Som, Tapas
    Barik Sr, Ramachandra
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (08)