Function of an Implanted Tissue Glucose Sensor for More than 1 Year in Animals

被引:131
作者
Gough, David A. [1 ]
Kumosa, Lucas S. [1 ]
Routh, Timothy L. [2 ]
Lin, Joe T. [2 ]
Lucisano, Joseph Y. [2 ]
机构
[1] Univ Calif San Diego, Dept Bioengn, La Jolla, CA 92093 USA
[2] GlySens Inc, San Diego, CA 92121 USA
关键词
BLOOD-GLUCOSE; OXYGEN SENSORS; INTENSIVE TREATMENT; ENZYME ELECTRODE; DYNAMICS; CATALASE; OXIDASE; SYSTEM; ERROR;
D O I
10.1126/scitranslmed.3001148
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
An implantable sensor capable of long-term monitoring of tissue glucose concentrations by wireless telemetry has been developed for eventual application in people with diabetes. The sensor telemetry system functioned continuously while implanted in subcutaneous tissues of two pigs for a total of 222 and 520 days, respectively, with each animal in both nondiabetic and diabetic states. The sensor detects glucose via an enzyme electrode that is based on differential electrochemical oxygen detection, which reduces the sensitivity of the sensor to encapsulation by the body, variations in local microvascular perfusion, limited availability of tissue oxygen, and inactivation of the enzymes. After an initial 2-week stabilization period, the implanted sensors maintained stability of calibration for extended periods. The lag between blood and tissue glucose concentrations was 11.8 +/- 5.7 and 6.5 +/- 13.3 minutes (mean +/- standard deviation), respectively, for rising and falling blood glucose challenges. The lag resulted mainly from glucose mass transfer in the tissues, rather than the intrinsic response of the sensor, and showed no systematic change over implant test periods. These results represent a milestone in the translation of the sensor system to human applications.
引用
收藏
页数:8
相关论文
共 38 条
[1]  
[Anonymous], NEW ENGL J MED
[2]  
*ANSI, 141601998 ANSIAAMIIS
[3]   APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS [J].
ARMOUR, JC ;
LUCISANO, JY ;
MCKEAN, BD ;
GOUGH, DA .
DIABETES, 1990, 39 (12) :1519-1526
[4]   Dynamic delay and maximal dynamic error in continuous biosensors [J].
Baker, DA ;
Gough, DA .
ANALYTICAL CHEMISTRY, 1996, 68 (08) :1292-1297
[5]   A SIMPLE TECHNIQUE FOR PRESERVATION OF FIXATION-SENSITIVE ANTIGENS IN PARAFFIN-EMBEDDED TISSUES [J].
BECKSTEAD, JH .
JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1994, 42 (08) :1127-1134
[6]   Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor [J].
Boyne, MS ;
Silver, DM ;
Kaplan, J ;
Saudek, CD .
DIABETES, 2003, 52 (11) :2790-2794
[7]   Is blood glucose predictable from previous values? A solicitation for data [J].
Bremer, T ;
Gough, DA .
DIABETES, 1999, 48 (03) :445-451
[8]   Real-Time Hypoglycemia Prediction Suite Using Continuous Glucose Monitoring A safety net for the artificial pancreas [J].
Dassau, Eyal ;
Cameron, Fraser ;
Lee, Hyunjin ;
Bequette, B. Wayne ;
Zisser, Howard ;
Jovanovic, Lois ;
Chase, H. Peter ;
Wilson, Darrell M. ;
Buckingham, Bruce A. ;
Doyle, Francis J., III .
DIABETES CARE, 2010, 33 (06) :1249-1254
[9]  
Garg S, 2006, DIABETES CARE, V29, P44, DOI 10.2337/diacare.29.01.06.dc05-1686
[10]  
Glanz S.A., 1992, PRIMER BIOSTATISTICS, V3rd