Optimizing Pacemaker Longevity with Pacing Mode and Settings Programming: Results from a Pacemaker Multicenter Registry

被引:14
作者
Benkemoun, Henri [1 ]
Sacrez, Jean [1 ]
Lagrange, Philippe [1 ]
Amie, Alain [2 ]
Prakash, Atul [3 ]
Himmrich, Ewald [4 ]
Aime, Ezio [5 ]
Mairesse, Georges H. [6 ]
Guenon, Christine [7 ]
Sbragia, Pascal [8 ]
机构
[1] Clin St Pierre, Perpignan, France
[2] Univ Hosp, Poitiers, France
[3] St Marys Hosp, Passaic, NJ USA
[4] Univ Med Ctr, Mainz, Germany
[5] IRCCS Policlin San Donato, Milan, Italy
[6] Clin Sud Luxembourg, Arlon, Belgium
[7] Sorin Grp France, Clamart, France
[8] Hop Nord Marseille, APM, Marseille, France
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2012年 / 35卷 / 04期
关键词
cardiac pacing; longevity; programming mode; reprogramming; DUAL-CHAMBER; GENERATOR LONGEVITY; OUTPUT; IMPLANTATION; ATRIAL; TRIAL; LEADS;
D O I
10.1111/j.1540-8159.2011.03318.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study aimed to describe the influence on dual-chamber devices' expected longevity of devices' settings. Methods: Data from patients implanted with dual chamber devices (Symphony (TM), SORIN CRM SAS, Clamart, France) from 2003 to 2006 were collected in registries. Programmer files were retrieved: device-estimated longevity, assessed through algorithm prediction, was analyzed according to device settings. Results: One thousand sixty-eight recipients of dual chamber pacemaker in sinus rhythm (75.3 +/- 11.1 years, 54.5% male, ventricular block 30%, brady-tachy syndrome 21%, and sinus node dysfunction 49%) were followed up to 14.2 +/- 12.1 months (ranging from first quartile Q1: 2.9 months to fourth quartile Q4: 49.3 months) after implantation. DDD with automatic mode conversion and minimized ventricular pacing (SafeR) modes were programmed in 34.3%, 2.9%, and 62.8% of the patients, respectively. The mean total longevity estimated by the device was 134.1 +/- 31.5 months (11.2 +/- 2.6 years). Significant increase in longevity was observed in devices undergoing at least one reprogramming (134.4 +/- 31.4 months) versus device presenting no reprogramming (103.4 +/- 32.3 months, P = 0.0005). The parameters associated with the major increase in mean longevity were the mode (mean longevity increase of + 23.9 months in SafeR as compared to DDD mode, P < 0.0001) and the atrial (A) and ventricular (V) amplitudes (mean longevity increase of + 29.6 and + 26.9 months for a decrease of less than 1V in A and V outputs respectively, P < 0.0001). Conclusion: This study provides information on dual chamber pacemakers' longevity and highlights the impact of devices' reprogramming on expected longevities. (PACE 2012; 35: 403-408)
引用
收藏
页码:403 / 408
页数:6
相关论文
共 21 条
  • [1] BAROLD SS, 1997, HEART DIS, P705
  • [2] Long-term survival after pacemaker implantation - Prognostic importance of gender and baseline patient characteristics
    Brunner, M
    Olschewski, M
    Geibel, A
    Bode, C
    Zehender, M
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 (01) : 88 - 95
  • [3] Automatic adjustment of pacemaker stimulation output correlated with continuously monitored capture thresholds:: A multicenter study
    Clarke, M
    Liu, B
    Schüller, H
    Binner, L
    Kennergren, C
    Guerola, M
    Weinmann, P
    Ohm, OJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (08): : 1567 - 1575
  • [4] Crossley GH, 1996, CIRCULATION, V94, P245
  • [5] Effect of modern pacing algorithms on generator longevity: A predictive analysis
    Gelvan, D
    Crystal, E
    Dokumaci, B
    Goldshmid, Y
    Ovsyshcher, IE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (09): : 1796 - 1802
  • [6] Late complications following permanent pacemaker implantation or elective unit replacement
    Harcombe, AA
    Newell, SA
    Ludman, PF
    Wistow, TE
    Sharples, LD
    Schofield, PM
    Stone, DL
    Shapiro, LM
    Cole, T
    Petch, MC
    [J]. HEART, 1998, 80 (03) : 240 - 244
  • [7] Longevity of dual chamber pacemakers:: Device and patient related determinants
    Kindermann, M
    Schwaab, B
    Berg, M
    Fröhlig, G
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (05): : 810 - 815
  • [8] Levine P., 2000, CLIN CARDIAC PACING, P827
  • [9] A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome - Echocardiographic and clinical outcome
    Nielsen, JC
    Kristensen, L
    Andersen, HR
    Mortensen, PT
    Pedersen, OL
    Pedersen, AK
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (04) : 614 - 623
  • [10] Ribeiro A L, 2001, Arq Bras Cardiol, V76, P437