Retrieval analysis of squeaking alumina ceramic-on-ceramic bearings

被引:43
作者
Walter, W. L. [1 ]
Kurtz, S. M. [1 ]
Esposito, C. [1 ]
Hozack, W. [1 ]
Holley, K. G. [1 ]
Garino, J. P. [1 ]
Tuke, M. A. [1 ]
机构
[1] Specialist Orthopaed Grp, Wollstonecraft, NSW 2065, Australia
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2011年 / 93B卷 / 12期
关键词
TOTAL HIP-ARTHROPLASTY; 1ST; 230; HIPS; ACETABULAR COMPONENT; POLYETHYLENE LINER; WEAR; FRACTURE; LUBRICATION; RESURFACINGS; ORIENTATION; PROSTHESES;
D O I
10.1302/0301-620X.93B12.27529
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This multicentre study analysed 12 alumina ceramic-on-ceramic components retrieved from squeaking total hip replacements after a mean of 23 months in situ (11 to 61). The rates and patterns of wear seen in these squeaking hips were compared with those seen in matched controls using retrieval data from 33 'silent' hip replacements with similar ceramic bearings. All 12 bearings showed evidence characteristic of edge-loading wear. The median rate of volumetric wear was 3.4 mm(3)/year for the acetabular component, 2.9 mm(3)/year on the femoral heads and 6.3 mm(3)/year for head and insert combined. This was up to 45 times greater than that of previously reported silent ceramic-on-ceramic retrievals. The rate of wear seen in ceramic components revised for squeaking hips appears to be much greater than in that seen in retrievals from 'silent' hips.
引用
收藏
页码:1597 / 1601
页数:5
相关论文
共 34 条
  • [1] Amstutz H. C., 1996, CLIN ORTHOP S, V329, P11
  • [2] AMSTUTZ HC, 1991, HIP ARTHROPLASTY, P1
  • [3] Early results of primary Birmingham hip resurfacings - An independent prospective study of the first 230 hips
    Back, DL
    Dalziel, R
    Young, D
    Shimmin, A
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (03): : 324 - 329
  • [4] BHUSHAN B, 2004, SPRINGER HDB NANOTEC, P883
  • [5] Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
  • [6] Chan FW, 1999, CLIN ORTHOP RELAT R, P10
  • [7] Charnley J, 1979, LOW FRICTION ARTHROP, P10
  • [8] Retrieval study of tibial baseplate fracture after total knee arthroplasty
    Chatterji, U
    Ashworth, MJ
    Smith, AL
    Brewster, N
    Lewis, PL
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (01) : 101 - 107
  • [9] The 2009 Frank Stinchfield Award "Hip Squeaking'' A Biomechanical Study of Ceramic-on-ceramic Bearing Surfaces
    Chevillotte, Christophe
    Trousdale, Robert T.
    Chen, Qingshan
    Guyen, Olivier
    An, Kai-Nan
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) : 345 - 350
  • [10] The results of primary Birmingham hip resurfacings at a mean of five years - An independent prospective review of the first 230 hips
    Hing, C. B.
    Back, D. L.
    Bailey, M.
    Young, D. A.
    Dalziel, R. E.
    Shimmin, A. J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (11): : 1431 - 1438