Energy dependence of TLD 100 and MCP-N detectors

被引:18
作者
Carinoua, E. [1 ]
Boziari, A. [1 ]
Askounisa, P. [1 ]
Mikulis, A. [2 ]
Kamenopoulou, V. [1 ]
机构
[1] Ag Paraskevi, Greek Atom Energy Commiss, Athens, Greece
[2] Radiat Safety Ctr, Riga, Latvia
关键词
energy dependence; TLD; 100H; MCP-N; RQR; narrow spectrum series; ring dosemeters;
D O I
10.1016/j.radmeas.2007.12.042
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
One of the most important characteristics of thermoluminescent detectors, if they are intended to be used in a wide range of energies, is their energy response. Thermoluminescent detectors (TLD) 100H and MCP-N, which are the commercial names of the LiF:Mg,Cu,P material, are being examined for their energy response against the well known and widely used material LiF:Mg,Ti, TLD 100. In this study the energy dependence of these materials is examined either in air or on a rod phantom. The dosemeters' energy dependence has been determined by using the ISO 4037-1 narrow spectrum series (NSS), Cs-137 and Co-60 as well as the RQR series according to IEC 61267. The relative response of TLD 100H and MCP-N at all energies is similar, indicating that the two TLD types have similar energy characteristics. The TLD 100 detectors present the same energy behaviour at the RQR series (in air and phantom configuration) with the TLD 100H and MCP-N. However, at the nearly monoenergetic NSS their relative energy response is higher showing an over response of nearly 60% at the NSS40. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:599 / 602
页数:4
相关论文
共 9 条
[1]   THERMOLUMINESCENCE EFFICIENCY OF LIFMG,CU,P (MCP-N) DETECTORS TO PHOTONS, BETA-ELECTRONS, ALPHA-PARTICLES AND THERMAL-NEUTRONS [J].
BILSKI, P ;
OLKO, P ;
BURGKHARDT, B ;
PIESCH, E ;
WALIGORSKI, MPR .
RADIATION PROTECTION DOSIMETRY, 1994, 55 (01) :31-38
[2]   Variations in dose response with x-ray energy of LiF:Mg,Cu,P thermoluminescence dosimeters:: implications for clinical dosimetry [J].
Duggan, L ;
Hood, C ;
Warren-Forward, H ;
Haque, M ;
Kron, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17) :3831-3845
[3]   The low energy X-ray response of the LiF:Mg:Cu:P thermoluminescent dosemeter:: a comparison with LiF:Mg:Ti [J].
Edwards, CR ;
Mountford, PJ ;
Green, S ;
Palethorpe, JE ;
Moloney, AJ .
BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (930) :543-547
[4]   ANGULAR-DEPENDENCE FACTORS AND AIR KERMA TO DOSE-EQUIVALENT CONVERSION COEFFICIENTS FOR CYLINDRICAL PHANTOMS IRRADIATED BY PLANE-PARALLEL EXTENDED MONOENERGETIC PHOTON BEAMS [J].
GROSSWENDT, B .
RADIATION PROTECTION DOSIMETRY, 1995, 59 (03) :165-179
[5]   Determination of LiF:Mg,Ti and LiF:Mg,Cu,P TL efficiency for X-rays and their application to Monte Carlo simulations of dosemeter response [J].
Hranitzky, C. ;
Stadtmann, H. ;
Olko, P. .
RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) :483-486
[6]  
IEC, 2005, IEC 612672005 MEDICA
[7]  
*ISO, 2000, 12794 ISO INT ORG ST
[8]  
*ISO, 1999, 40373 ISO INT ORG ST
[9]  
*ISO, 1996, 40371 ISO INT ORG ST