A fatal complication of a vena caval filter associated with pulmonary thromboembolism

被引:4
作者
Cunliffe, Clare H. [1 ]
Denton, J. Scott [1 ]
机构
[1] Off Med Examiner Cook Cty, Chicago, IL USA
关键词
vena caval filters; pulmonary thromboembolism; death;
D O I
10.1097/PAF.0b013e318173f00a
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Vena caval filters are increasingly used for the prevention of pulmonary thromboembolism in patients with deep vein thrombosis ever since the introduction of the stainless steel Greenfield filter in the 1970s. Although complications associated with their use are rare, they can be fatal, and therefore it is important for forensic pathologists to be aware of their occurrence. Because of the ramifications of a medical device failure, a thorough documentation of the autopsy findings is essential. We describe an unusual case of a 75-year-old woman who developed a fatal pulmonary thromboembolism with concurrent migration of a lesser known type of vena caval filter to the right ventricle occurring 19 days after filter placement. A review of the types of vena caval filters in current use, and their complications, are discussed.
引用
收藏
页码:173 / 176
页数:4
相关论文
共 14 条
[1]   VENA TECH-LGM FILTER - LONG-TERM RESULTS OF A PROSPECTIVE-STUDY [J].
CROCHET, DP ;
STORA, O ;
FERRY, D ;
GROSSETETE, R ;
LEURENT, B ;
BRUNEL, P ;
NGUYEN, JM .
RADIOLOGY, 1993, 188 (03) :857-860
[2]   LATE COMPLICATION OF A GREENFIELD FIL TER ASSOCIATING CAUDAL MIGRATION AND PERFORATION OF THE ABDOMINAL-AORTA BY A RUPTURED STRUT [J].
DABBAGH, A ;
CHAKFE, N ;
KRETZ, JG ;
DEMRI, B ;
NICOLINI, P ;
FUENTES, C ;
METTAUER, B ;
EPAILLY, E ;
MUSTER, D ;
EISENMANN, B .
JOURNAL OF VASCULAR SURGERY, 1995, 22 (02) :182-187
[3]   Duodenal perforation with an inferior vena cava filter: An unusual cause of abdominal pain [J].
Feezor, RJ ;
Huber, TS ;
Welborn, MB ;
Schell, SR .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1010-1012
[4]  
GATES J, CURRENT INDICATORS I
[5]   EXTENDED EVALUATION OF THE TITANIUM GREENFIELD VENA-CAVAL FILTER [J].
GREENFIELD, LJ ;
PROCTOR, MC ;
CHO, KJ ;
CUTLER, BS ;
FERRIS, EJ ;
MCFARLAND, D ;
SOBEL, M ;
TISNADO, J .
JOURNAL OF VASCULAR SURGERY, 1994, 20 (03) :458-465
[6]  
GREENFIELD LJ, 1991, J VASC SURG, V14, P253
[7]  
Izutani H, 2004, CAN J CARDIOL, V20, P233
[8]   Tricuspid insufficiency after intracardiac migration of a Greenfield filter: Case report and review of the literature [J].
James, KV ;
Sobolewski, AP ;
Lohr, JM ;
Welling, RE .
JOURNAL OF VASCULAR SURGERY, 1996, 24 (03) :494-498
[9]  
Joels CS, 2003, AM SURGEON, V69, P654
[10]  
MAGNANT JG, 1992, J VASC SURG, V16, P701