Performance of the round window soft coupler for the backward stimulation of the cochlea in a temporal bone model

被引:7
作者
Gostian, Antoniu-Oreste [1 ]
Schwarz, David [1 ]
Mandt, Philipp [1 ]
Anagiotos, Andreas [1 ]
Ortmann, Magdalene [2 ]
Pazen, David [1 ]
Beutner, Dirk [1 ]
Huettenbrink, Karl-Bernd [1 ]
机构
[1] Univ Cologne, Fac Med, Dept Otorhinolaryngol Head & Neck Surg, Kerpener Str 62, D-50937 Cologne, Germany
[2] Jean Uhrmacher Inst Clin ENT Res, Geibelstr 29-31, D-50931 Cologne, Germany
关键词
Middle ear implant; Hearing loss; Round window vibroplasty; Soft round window coupler; Backward stimulation; FLOATING MASS TRANSDUCER; MIXED HEARING LOSSES; VIBRANT SOUNDBRIDGE; VIBROPLASTY; RECONSTRUCTION; CONDUCTION; MEMBRANE; IMPACT;
D O I
10.1007/s00405-016-3997-2
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The round window vibroplasty is a feasible option for the treatment of conductive, sensorineural and mixed hearing loss. Although clinical data suggest a satisfying clinical outcome with various coupling methods, the most efficient coupling technique of the floating mass transducer to the round window is still a matter of debate. For this, a soft silicone-made coupler has been developed recently that aims to ease and optimize the stimulation of the round window membrane of this middle ear implant. We performed a temporal bone study evaluating the performance of the soft coupler compared to the coupling with individually shaped cartilage, perichondrium and the titanium round window coupler with loads up to 20 mN at the unaltered and fully exposed round window niche. The stimulation of the cochlea was measured by the volume velocities of the stapes footplate detected by a laser Doppler vibrometer. The coupling method was computed as significant factor with cartilage and perichondrium allowing for the highest volume velocities followed by the soft and titanium coupler. Exposure of the round window niche allowed for higher volume velocities while the applied load did not significantly affect the results. The soft coupler allows for a good contact to the round window membrane and an effective backward stimulation of the cochlea. Clinical data are mandatory to evaluate performance of this novel coupling method in vivo.
引用
收藏
页码:3651 / 3661
页数:11
相关论文
共 26 条
[11]  
ISHII T, 1995, ACTA OTO-LARYNGOL, P78
[12]   Experience with the Vibrant Soundbridge RW-Coupler for round window Vibroplasty with tympanosclerosis [J].
Iwasaki, Satoshi ;
Suzuki, Hiroaki ;
Moteki, Hideaki ;
Miyagawa, Maiko ;
Takumi, Yutaka ;
Usami, Shin-ichi .
ACTA OTO-LARYNGOLOGICA, 2012, 132 (06) :676-682
[13]   Round window stimulation with an implantable hearing aid (Soundbridge®) combined with autogenous reconstruction of the auricle -: A new approach [J].
Kiefer, Jan ;
Arnold, Wolfgang ;
Staudenmaier, Rainer .
ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 2006, 68 (06) :378-385
[14]  
Luers JC, 2013, OTOL NEUROTOL, V34, P1005, DOI 10.1097/MAO.0b013e3182990d2b
[15]   The effect of static force on round window stimulation with the direct acoustic cochlea stimulator [J].
Maier, Hannes ;
Salcher, Rolf ;
Schwab, Burkard ;
Lenarz, Thomas .
HEARING RESEARCH, 2013, 301 :115-124
[16]  
Müller S, 2001, J AUDIO ENG SOC, V49, P443
[17]   Evaluation of Round Window Stimulation Using the Floating Mass Transducer by Intracochlear Sound Pressure Measurements in Human Temporal Bones [J].
Nakajima, Hideko Heidi ;
Dong, Wei ;
Olson, Elizabeth S. ;
Rosowski, John J. ;
Ravicz, Michael E. ;
Merchant, Saumil N. .
OTOLOGY & NEUROTOLOGY, 2010, 31 (03) :506-511
[18]   Analysis of Vibrant Soundbridge Placement Against the Round Window Membrane in a Human Cadaveric Temporal Bone Model [J].
Pennings, Ronald J. E. ;
Ho, Allan ;
Brown, Jeremy ;
van Wijhe, Rene G. ;
Bance, Manohar .
OTOLOGY & NEUROTOLOGY, 2010, 31 (06) :998-1003
[19]   Impact of Floating Mass Transducer Coupling and Positioning in Round Window Vibroplasty [J].
Rajan, Gunesh P. ;
Lampacher, Peter ;
Ambett, Ranjeeta ;
Dittrich, Gregor ;
Kuthubutheen, Jafri ;
Wood, Bradley ;
McArthur, Anne ;
Marino, Roberta .
OTOLOGY & NEUROTOLOGY, 2011, 32 (02) :271-277
[20]   Round window stimulation with the floating mass transducer at constant pretension [J].
Salcher, Rolf ;
Schwab, Burkard ;
Lenarz, Thomas ;
Maier, Hannes .
HEARING RESEARCH, 2014, 314 :1-9