Development of an Immunity Test System With a Pseudo Biosignal Generator for Wearable Devices and Application to the ESD Test of an Artificial Hand

被引:6
|
作者
Wang, Jianqing [1 ]
Nakaya, Ryo [1 ,2 ]
Sato, Keisuke [1 ,3 ]
Anzai, Daisuke [1 ]
Fujiwara, Osamu [1 ]
Amemiya, Fujio [4 ]
机构
[1] Nagoya Inst Technol, Grad Sch Engn, Nagoya, Aichi 4668555, Japan
[2] Toyota Ind Corp, Kariya, Aichi 4488671, Japan
[3] Konica Minolta Inc, Tokyo 1007015, Japan
[4] NTT Adv Technol Corp, Kawasaki, Kanagawa 2120014, Japan
基金
日本学术振兴会;
关键词
Biosignal generator; electrostatic discharge (ESD); immunity test; myoelectric artificial hand; wearable device; ECG;
D O I
10.1109/TEMC.2017.2785353
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this study, we developed an immunity test system for wearable devices used for healthcare or assistance to people with disabilities. The immunity test system consists of a pseudo biosignal generator for generating various vital signals such as electrocardiogram or electromyography and a bioequivalent gel phantom for simulating a part of the human body. As an application to the electrostatic discharge (ESD) immunity test for a myoelectric artificial hand, we used the pseudo biosignal generator to generate various myoelectric signals and confirmed a high correlation between the generated and the real myoelectric signals. We then applied this system to the IEC61000-4-2 specified ESD test. Through a comparison between the ESD test results by using the immunity test system and the result for a real human body, we found a high similarity between them, which suggests that the proposed immunity test system is feasible for various immunity tests in wearable devices.
引用
收藏
页码:73 / 81
页数:9
相关论文
共 50 条
  • [1] A Proto-type ESD Generator for System Immunity Test of Wearable Devices
    Park, Junsik
    Lee, Jongsung
    Jo, Cheolgu
    Seol, Byongsu
    Kim, Jingook
    2018 40TH ELECTRICAL OVERSTRESS/ELECTROSTATIC DISCHARGE SYMPOSIUM (EOS/ESD), 2018,
  • [2] An ESD Immunity Test for Battery-Operated Control Circuit Board in Myoelectric Artificial Hand System
    Ji, Cheng
    Anzai, Daisuke
    Wang, Jianqing
    Mori, Ikuko
    Fujiwara, Osamu
    IEICE TRANSACTIONS ON COMMUNICATIONS, 2015, E98B (12) : 2477 - 2484
  • [3] The Repeatability of System Level ESD Test and Relevant ESD Generator Parameters
    Koo, Jayong
    Cai, Qing
    Pommerenke, David
    Wang, Kai
    Mass, John
    Hirata, Masayuki
    Martwick, Andy
    2008 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, VOLS 1-3, 2008, : 626 - +
  • [4] Development of Immunity Test Method for Long-Duration Induced Noise on a Wearable Devices by Electrostatic Discharge
    Tanaka, Musashi
    Yoshida, Takahiro
    PROCEEDINGS OF THE 2021 ASIA-PACIFIC INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (APEMC 2021), 2021,
  • [5] Development and application of geotechnical direct tension test devices
    Geotechnical and Structural Engineering Research Center, Shandong University, Jinan 250061, China
    不详
    Yantu Gongcheng Xuebao, 7 (1309-1315):
  • [6] Development of generator for voltage dips, short interruptions and voltage variations immunity test
    Yu, ZY
    Liu, GB
    Wu, H
    2002 3RD INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, 2002, : 67 - 70
  • [7] Application of Artificial Intelligence in the Driver Road Test System
    Wu, Yefu
    Zhang, Dan
    Yuan, Xiaoling
    PERVASIVE COMPUTING AND THE NETWORKED WORLD, 2014, 8351 : 686 - 697
  • [8] Development of an arbitrary function generator and test system for nuclear spectrometers
    Avci, O.
    Bayburt, S.
    Bayburt, M.
    Celiktas, C.
    RADIATION MEASUREMENTS, 2011, 46 (08) : 730 - 733
  • [9] COMPOSITE NOISE GENERATOR (CNG) WITH RANDOM PULSE STREAM (RPS) GENERATOR FOR IMMUNITY TEST IN DIGITAL SYSTEM
    SUZUKI, H
    SHIZUYA, H
    TAKAGI, T
    IEICE TRANSACTIONS ON COMMUNICATIONS, 1992, E75B (03) : 183 - 187
  • [10] Development and experimental application of substructure pseudo dynamic test platform NetSLabOSR
    Huang M.
    Guo Y.
    Guo, Yurong (yurongguo@hnu.edu.cn), 1600, Harbin Institute of Technology (48): : 89 - 94