Automated peritoneal dialysis with 'on-line'-prepared bicarbonate-buffered dialysate: technique and first clinical experiences

被引:15
|
作者
Brunkhorst, R [1 ]
Fromm, S
Wrenger, E
Berke, A
Petersen, R
Riede, G
Westphale, J
Zamore, E
Ledebo, I
机构
[1] Hannover Med Sch, Div Nephrol, Ctr Internal Med, D-30623 Hannover, Germany
[2] Otto Von Guericke Univ, Div Nephrol, Magdeburg, Germany
[3] Gambro, Lund, Sweden
关键词
automated peritoneal dialysis; bicarbonate dialysate; on-line-prepared dialysate;
D O I
10.1093/ndt/13.12.3189
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
Background. Automated peritoneal dialysis (APD) has the possibility of increasing the dialysis efficacy by using higher fill volumes, frequent dialysate exchanges, and tidal techniques. It is then possible to treat patients adequately without residual renal function. The drawbacks of the required high amounts of dialysis solution of up to 30 litres per session are the high costs of lactate-based dialysate bags and difficulties for the patients in handling these bags. So far, bicarbonate-based peritoneal dialysate, which may be more biocompatible, is only available for CAPD in double-chamber bags. In APD this could be overcome by 'on-line' preparation of bicarbonate-buffered dialysate using advanced technologies originally designed for on-line preparation of substitution fluid for haemofiltration. Methods. Four patients without residual renal function were treated with APD five times weekly in a crossover study design. Patients received standard lactate-based (35 mmol/l) treatment (25 litres per session each) in weeks 1 and 3. In week 2 on-line-produced bicarbonate-buffered (37 mmol/l) dialysate was used. This dialysate was prepared by an AK 100 Ultra haemodialysis machine. The machine was modified for adding glucose from a 50% concentrate to the desired concentration of 1.7%. Electrolytes, pH, pCO(2), and dialysis efficacy parameters were measured. Microbiological testing was carefully performed. Results. Creatinine clearances, Kt/V, and pCO(2) did not vary between the different treatment phases, whereas the pH showed a distinct increase during the bicarbonate phase. Repeated determinations of endotoxins and culturing showed no contamination of the dialysate. The composition of the produced dialysate was reproducible with respect to pH, pCO(2), sodium, calcium and bicarbonate, whereas the glucose concentration varied by +/-20%. Conclusions. On-line preparation of PD fluid with the AK 100 Ultra is easy and safe to handle. APD with dialysate containing 37 mmol/l bicarbonate provides improved acid-base balance and possibly improved biocompatibility, and may lead to a significant cost reduction. Further development in order to provide smaller machines and more precise ways of achieving a desired dialysate glucose concentration is necessary.
引用
收藏
页码:3189 / 3192
页数:4
相关论文
共 36 条
  • [21] In vitro biocompatibility evaluation of a novel bicarbonate-buffered amino-acid solution for peritoneal dialysis
    Jorres, A
    Gahl, GM
    Ludat, K
    Frei, U
    PasslickDeetjen, J
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 1997, 12 (03) : 543 - 549
  • [22] Effects of pH-neutral, bicarbonate-buffered dialysis fluid on peritoneal transport kinetics in children
    Schmitt, CP
    Haraldsson, B
    Doetschmann, R
    Zimmering, M
    Greiner, C
    Böswald, M
    Klaus, G
    Passlick-Deetjen, J
    Schaefer, F
    KIDNEY INTERNATIONAL, 2002, 61 (04) : 1527 - 1536
  • [23] PERITONEAL-DIALYSIS USING BICARBONATE-CONTAINING DIALYSATE PRODUCED BY AUTOMATED DIALYSATE DELIVERY MACHINE
    ING, TS
    GANDHI, VC
    DAUGIRDAS, JT
    HUNT, J
    QUON, MJ
    ROPLI, S
    ARTIFICIAL ORGANS, 1982, 6 (01) : 67 - 69
  • [24] IN-VITRO-BIOCOMPATIBILITY EVALUATION OF A NOVEL BICARBONATE-BUFFERED AMINO-ACID SOLUTION FOR PERITONEAL-DIALYSIS
    JORRES, A
    GAHL, GM
    LUDAT, K
    FREI, U
    PASSLICKDEETJEN, J
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1994, 5 (03): : 457 - 457
  • [25] ON LINE PREPARATION OF BICARBONATE-CONTAINING DIALYSATE FOR USE IN PERITONEAL-DIALYSIS
    ING, TS
    QUON, MJ
    DAUGIRDAS, JT
    LIU, P
    GANDHI, VC
    REID, RR
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1981, 4 (06): : 308 - 309
  • [26] Bicarbonate/lactate and bicarbonate-buffered peritoneal dialysis fluids improve ex vivo peritoneal macrophage TNF-α secretion (vol 9, pg 1499, 1998)
    Mackenzie, RK
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2000, 11 (01):
  • [27] Clinical experience with a 39 mmol/L bicarbonate buffered peritoneal dialysis solution
    Feriani, M
    Carobi, C
    LaGreca, G
    Buoncristiani, U
    PasslickDeetjen, J
    PERITONEAL DIALYSIS INTERNATIONAL, 1997, 17 (01): : 17 - 21
  • [28] Clinical experience with two physiologic bicarbonate/lactate peritoneal dialysis solutions in automated peritoneal dialysis
    Dratwa, M
    Wilkie, M
    Ryckelynck, JP
    ter Wee, PM
    Rutherford, P
    Michel, C
    Hopwood, A
    Curtis, L
    Denys, N
    Divino, JC
    Faict, D
    KIDNEY INTERNATIONAL, 2003, 64 : S105 - S113
  • [29] DIALYSATE CONSUMPTION AND DIALYSIS DOSE IN ON-LINE HEMODIAFILTRATION WITH AUTOMATED COUPLING DIALYSATE/BLOOD FLOW FUNCTION (AUTOFLOW FUNCTION)
    Park, Yongki
    Jeon, Jimin
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 206 - 207
  • [30] Clinical experience with two physiological bicarbonate/lactate dialysis solutions in automated peritoneal dialysis.
    Rutherford, P
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2003, 14 : 747A - 747A