PERFORMANCE OF SCINTILLATION PIXEL DETECTORS WITH MPPC READ-OUT AND DIGITAL SIGNAL PROCESSING

被引:3
作者
Makek, M. [1 ]
Bosnar, D. [1 ]
Gacic, V. [1 ]
Pavelic, L. [1 ]
Senjug, P. [1 ]
Zugec, P. [1 ]
机构
[1] Univ Zagreb, Fac Sci, Dept Phys, Bijenicka 32, Zagreb 10000, Croatia
来源
ACTA PHYSICA POLONICA B | 2017年 / 48卷 / 10期
关键词
D O I
10.5506/APhysPolB.48.1721
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We have assembled and tested a system of 2 detector modules containing Lutetium Fine Silicate scintillator pixels in 4 x 4 matrix, read out by multi-pixel photon counter arrays with the matching element size. The amplified signals were acquired using fast digitizers and stored for offline analysis. Using two different approaches, we have determined the single crystal relative energy resolution at 511 keV to be 14% and 12%. The coincidence time resolution has been tested using the digital constant fraction triggering and the leading edge method, and the resulting resolutions are 1.6 ns and 0.5 ns, respectively.
引用
收藏
页码:1721 / 1726
页数:6
相关论文
共 4 条
[1]  
[Anonymous], [No title captured]
[2]   Comparative timing measurements of LYSO and LFS-3 to achieve the best time resolution for TOF-PET [J].
Doroud, K. ;
Williams, M. C. S. ;
Zichichi, A. ;
Zuyeuski, R. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 793 :57-61
[3]   Novel type of avalanche photodetector with Geiger mode operation [J].
Golovin, V ;
Saveliev, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 518 (1-2) :560-564
[4]   Next-generation PET capability with lutetium fine silicate and multi-pixel photon counter [J].
Yamazaki, M. ;
Takeshita, T. ;
Hasegawa, Y. .
JOURNAL OF INSTRUMENTATION, 2012, 7