Evaluation of Capiox RX25 and Quadrox-i Adult Hollow Fiber Membrane Oxygenators in a Simulated Cardiopulmonary Bypass Circuit

被引:8
|
作者
Wang, Shigang [1 ]
Kunselman, Allen R. [2 ]
Undar, Akif [1 ,3 ]
机构
[1] Penn State Hershey Childrens Hosp, Penn State Hershey Pediat Cardiovasc Res Ctr, Dept Pediat, Penn State Hershey Coll Med, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Publ Hlth & Sci, Penn State Hershey Coll Med, Hershey, PA USA
[3] Penn State Hershey Childrens Hosp, Surg & Bioengn, Penn State Milton S Hershey Med Ctr, Penn State Hershey Coll Med, Hershey, PA USA
关键词
Hollow fiber membrane oxygenator; Gaseous microemboli; Hemodynamic energy; Cardiopulmonary bypass; Adult; GASEOUS MICROEMBOLI; PRESSURE;
D O I
10.1111/aor.12633
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Capiox RX25 and Quadrox-i Adult oxygenators are commonly used in clinical adult cardiopulmonary bypass circuits. This study was designed to test the effectiveness of two adult oxygenators in order to evaluate gaseous microemboli (GME) trapping capability and hemodynamic performance. A simulated adult CPB circuit was used and primed with Ringer's lactate and packed red blood cells (hematocrit 25%). All trials were conducted at flow rates of 2-5 L/min (1 L/min increments) with a closed and open arterial filter purge line at 35 degrees C. The postcannula pressure was maintained at 100 mm Hg. After a 5 cc of bolus air was introduced into the venous line, an Emboli Detection and Classification system was used to detect and classify GME at the preoxygenator, postoxygenator, and precannula sites. At the same time, real-time pressure and flow data were recorded, and hemodynamic energy was calculated using a custom-made data acquisition system and Labview software. Our results showed that the oxygenator pressure drops of Quadrox-i Adult oxygenator were lower than Capiox RX25 at all flow rates. The Quadrox-i Adult oxygenator retained more hemodynamic energy across the oxygenator. Both oxygenators could trap the majority of GME, but Capiox RX25 did better than the Quadrox-i Adult oxygenator. No GME was delivered to the pseudo patient at all flow rates in the Capiox group. The Capiox RX25 venous reservoir could capture more GME at lower flow rates, while the Quadrox-i Adult venous reservoir performed better at higher flow rates. An open arterial filter purge line reduced GME slightly in the Capiox group, but GME increased in the Quadrox group. The Quadrox-i Adult oxygenator is a low-resistance, high-compliance oxygenator. The GME handling ability of Capiox RX25 performed well under our clinical setting. Further optimized design for the venous/cardiotomy reservoir is needed.
引用
收藏
页码:E69 / E78
页数:10
相关论文
共 4 条
  • [1] Evaluation of Quadrox-i and Capiox FX neonatal oxygenators with integrated arterial filters in eliminating gaseous microemboli and retaining hemodynamic properties during simulated cardiopulmonary bypass
    Lin, J.
    Dogal, N. M.
    Mathis, R. K.
    Qiu, F.
    Kunselman, A.
    Uendar, A.
    PERFUSION-UK, 2012, 27 (03): : 235 - 243
  • [2] Evaluation of three hollow-fiber membrane oxygenators without integrated arterial filters for neonatal cardiopulmonary bypass
    Dogal, N. M.
    Mathis, R. K.
    Lin, J.
    Qiu, F.
    Kunselman, A.
    Uendar, A.
    PERFUSION-UK, 2012, 27 (02): : 132 - 140
  • [3] In vitro evaluation of Capiox FX05 and RX05 oxygenators in neonatal cardiopulmonary bypass circuits with varying venous reservoir and vacuum-assisted venous drainage levels
    Sathianathan, Shyama
    Nasir, Rafay
    Wang, Shigang
    Kunselman, Allen R.
    Undar, Akif
    ARTIFICIAL ORGANS, 2020, 44 (01) : 28 - 39
  • [4] Pressure drop, shear stress, and activation of leukocytes during cardiopulmonary bypass: A comparison between hollow fiber and flat sheet membrane oxygenators
    Gu, YJ
    Boonstra, PW
    Graaff, R
    Rijnsburger, AA
    Mungroop, H
    van Oeveren, W
    ARTIFICIAL ORGANS, 2000, 24 (01) : 43 - 48