Automatic spectral calibration for polarization-sensitive optical coherence tomography

被引:8
作者
Chen, Yan [1 ,2 ]
Li, Zhongliang [1 ,2 ]
Nan, Nan [1 ,2 ]
Bu, Yang [1 ,2 ]
Wang, Xuan [1 ,2 ]
Pan, Liuhua [1 ,2 ]
Wang, Xiangzhao [1 ,2 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Opt & Fine Mech, Lab Informat Opt & Optoelect Technol, Shanghai 201800, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
关键词
HUMAN SKIN; BIREFRINGENCE; RESOLUTION; INTERFEROMETRY; RANGE; OCT;
D O I
10.1364/OE.25.023605
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Accurate wavelength assignment is important for Fourier domain polarization-sensitive optical coherence tomography. Incorrect wavelength mapping between the orthogonal horizontal (H) and vertical (V) polarization channels leads to broadening the axial point spread function and generating polarization artifacts. To solve the problem, we propose an automatic spectral calibration method by seeking the optimal calibration coefficient between wavenumber k(H) and k(V). The method first performs a rough calibration to get the relationship between the wavelength. and the pixel number x of the CCD for each channel. And then a precise calibration is taken to bring both polarization interferograms in the same k range through the optimal calibration coefficient. The optimal coefficient is automatically obtained by evaluating the cross-correlation of A-line signals. Simulations and experiments are implemented to demonstrate the performance of the proposed method. The results show that, compared to the peaks method, the proposed method is suitable in both Gaussian and non-Gaussian spectrums with a higher calibration accuracy. (C) 2017 Optical Society of America
引用
收藏
页码:23605 / 23618
页数:14
相关论文
共 21 条
[21]   Large-field high-speed polarization sensitive spectral domain OCT and its applications in ophthalmology [J].
Zotter, Stefan ;
Pircher, Michael ;
Torzicky, Teresa ;
Baumann, Bernhard ;
Yoshida, Hirofumi ;
Hirose, Futoshi ;
Roberts, Philipp ;
Ritter, Markus ;
Schuetze, Christopher ;
Goetzinger, Erich ;
Trasischker, Wolfgang ;
Vass, Clemens ;
Schmidt-Erfurth, Ursula ;
Hitzenberger, Christoph K. .
BIOMEDICAL OPTICS EXPRESS, 2012, 3 (11) :2720-2732