A central positron source to perform the timing alignment of detectors in a PET scanner

被引:38
作者
Thompson, CJ [1 ]
Camborde, ML
Casey, ME
机构
[1] Montreal Neurol Inst, Montreal, PQ H3A 2B4, Canada
[2] UBC Triumf PET Grp, Brain Res Ctr, Vancouver, BC V6T 2B5, Canada
[3] CPS Innovat, Knoxville, TN 37932 USA
基金
加拿大自然科学与工程研究理事会;
关键词
calibration; central source; positron emission tomography; time of flight; timing alliance;
D O I
10.1109/TNS.2005.858234
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Accurate timing alignment and stability are important to maximize the true counts and minimize the random counts in positron emission tomography. Its importance increases in time-of-flight (TOF) scanners. We propose using a central positron emitting source enclosed in a detector which detects the excess energy of the positron before it annihilates as a timing reference. All crystals can be time-aligned with respect to this central source. We evaluated 10 mu Ci Na-22 and Ge-68 sources embedded in cylinders of plastic scintillator coupled to a fast PMT. Light flashes. produced after the parent isotope emits positrons are detected, and the anode signals from the PMT are the reference time for each positron decay. The time delay before the gamma ray is detected by the scanner's conventional gamma ray detectors is the time offset to be applied to that crystal. Since all detectors are almost the same distance from the central source, TOF errors are minimized. Preliminary results show a mean signal amplitude of >0.5 V from Na-22 at 1000-V PMT bias, a timing FWHM of 850 ps with respect to a small LSO crystal. This suggests it could be useful to align both conventional and TOF PET scanners.
引用
收藏
页码:1300 / 1304
页数:5
相关论文
共 13 条
  • [1] Characterization of a single LSO crystal layer high resolution research tomograph
    Boellaard, R
    Buijs, F
    de Jong, HWAM
    Lenox, M
    Gremillion, T
    Lammertsma, AA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (04) : 429 - 448
  • [2] Brix G, 1997, J NUCL MED, V38, P1614
  • [3] A positron-decay triggered transmission source for positron emission tomography
    Camborde, ML
    Thompson, CJ
    Togane, D
    Zhang, N
    Reader, A
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 53 - 57
  • [4] CASEY ME, 2004, J NUCL MED JUN
  • [5] DEGRADO TR, 1994, J NUCL MED, V35, P1398
  • [6] Performance evaluation of the microPET R4 PET scanner for rodents
    Knoess, C
    Siegel, S
    Smith, A
    Newport, D
    Richerzhagen, R
    Winkeler, A
    Jacobs, A
    Goble, RN
    Graf, R
    Wienhard, K
    Heiss, WD
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) : 737 - 747
  • [7] LINES KS, 1978, J PHYS B ATOM MOL PH, V11, P510
  • [8] LUO D, IEEE MIC C REC CD 20
  • [9] Time of flight in PET revisited
    Moses, WW
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1325 - 1330
  • [10] A method for determination of the timing stability of PET scanners
    Thompson, CJ
    Goertzen, AL
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (08) : 1053 - 1057