Actuation of Novel Blood Pump by Direct Application of Rotating Magnetic Field

被引:13
作者
Kim, S. H. [1 ]
Hashi, S. [1 ]
Ishiyama, K. [1 ]
机构
[1] Tohoku Univ, Elect Commun Res Inst, Aoba Ku, Katahira 9808577, Japan
基金
日本学术振兴会;
关键词
Battery-free; direct application of rotating magnetic field; implantable blood pump; magnetic torque; wireless; SUPPORT; SYSTEM;
D O I
10.1109/TMAG.2011.2177527
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents a magnetic centrifugal pump, driven by an external rotating magnetic field, for use in artificial heart-assist blood pumps. The magnet rotor in the pump is synchronized to a rotating magnetic field, which then causes the fabricated impeller to produce a centrifugal force in the pump. Because the pump is driven by the direct application of an external rotating magnetic field, the proposed pump does not require a shaft or mechanical bearings. Thus, the proposed mechanism and external driving source give the pump the advantages of compact size; wireless, battery-free operation; heat-free operation; and minimal risk of bacterial infections. The pump weight, length, and diameter are 34 g, 35 mm, and 20 mm, respectively. The compact size and performance make the device suitable for use in pediatric cardiac assistance pumps.
引用
收藏
页码:1869 / 1874
页数:6
相关论文
共 19 条
[1]   The National Heart, Lung, and Blood Institute Pediatric Circulatory Support Program A Summary of the 5-Year Experience [J].
Baldwin, J. Timothy ;
Borovetz, Harvey S. ;
Duncan, Brian W. ;
Gartner, Mark J. ;
Jarvik, Robert K. ;
Weiss, William J. .
CIRCULATION, 2011, 123 (11) :1233-1240
[2]   The HeartMate II: Design and development of a fully sealed axial flow left ventricular assist system [J].
Burke, DJ ;
Burke, E ;
Parsaie, F ;
Poirier, V ;
Butler, K ;
Thomas, D ;
Taylor, L ;
Maher, T .
ARTIFICIAL ORGANS, 2001, 25 (05) :380-385
[3]   Current status of the total artificial heart [J].
Gray, Norman A., Jr. ;
Selzman, Craig H. .
AMERICAN HEART JOURNAL, 2006, 152 (01) :4-10
[4]   A magnetically levitated centrifugal blood pump with a simple-structured disposable pump head [J].
Hijikata, Wataru ;
Shinshi, Tadahiko ;
Asama, Junichi ;
Li, Lichuan ;
Hoshi, Hideo ;
Takatani, Setsuo ;
Shimokohbe, Akira .
ARTIFICIAL ORGANS, 2008, 32 (07) :531-540
[5]   Disposable MagLev Centrifugal Blood Pump Utilizing a Cone-Shaped Impeller [J].
Hijikata, Wataru ;
Sobajima, Hideo ;
Shinshi, Tadahiko ;
Nagamine, Yasuyuki ;
Wada, Suguru ;
Takatani, Setsuo ;
Shimokohbe, Akira .
ARTIFICIAL ORGANS, 2010, 34 (08) :669-U3
[6]   Magnetic Drive System for a New Centrifugal Rotary Blood Pump [J].
Hilton, Andrew ;
Tansley, Geoff .
ARTIFICIAL ORGANS, 2008, 32 (10) :772-777
[7]   A Method for Acquiring the Torque of a Magnetic Pump [J].
Kim, S. H. ;
Hashi, S. ;
Ishiyama, K. .
IEEE TRANSACTIONS ON MAGNETICS, 2011, 47 (10) :3971-3974
[8]   Centrifugal Force Based Magnetic Micro-Pump Driven by Rotating Magnetic Fields [J].
Kim, S. H. ;
Hashi, S. ;
Ishiyama, K. .
2ND INTERNATIONAL SYMPOSIUM ON ADVANCED MAGNETIC MATERIALS AND APPLICATIONS (ISAMMA 2010), 2011, 266
[9]  
Kitao T, 2011, ARTIF ORGANS, V35, P543, DOI [10.1111/j.1525-1594.2011.01321.x, 10.1111/j.1525-1594.2010.01192.x]
[10]   EXPERIMENTAL-STUDY OF PHYSIOLOGICAL ADVANTAGES OF ASSIST CIRCULATION USING OSCILLATED BLOOD-FLOW [J].
KOBAYASHI, S ;
NITTA, S ;
YAMBE, T ;
NAGANUMA, S ;
HASHIMOTO, H ;
FUKUJU, T ;
TABAYASHI, K .
ARTIFICIAL ORGANS, 1995, 19 (07) :704-707