Method Verification of the Caretium XC-A30 Automated Erythrocyte Sedimentation Rate Analyser for Erythrocyte Sedimentation Rate

被引:0
|
作者
Khongwichit, Soemwit [1 ]
Saelim, Menapha [2 ]
Na Songkhla, Yanisa [2 ]
Buncherd, Hansuk [1 ]
Nopparatana, Chawadee [2 ]
Srinoun, Kanitta [1 ,3 ]
机构
[1] Prince Songkla Univ, Fac Med Technol, Hat Yai, Thailand
[2] Prince Songkla Univ, Fac Med, Dept Pathol, Hat Yai, Thailand
[3] Prince Songkla Univ, Fac Med Technol, 15 Kanjanavanit Rd, Hat Yai 90110, Songkhla, Thailand
来源
MALAYSIAN JOURNAL OF MEDICAL SCIENCES | 2022年 / 29卷 / 05期
关键词
erythrocyte sedimentation rate; automated ESR analyser; method verification; MATIC CUBE 200; WESTERGREN METHOD;
D O I
10.21315/mjms2022.29.5.5
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: The erythrocyte sedimentation rate (ESR) analyser is widely used in haematological testing. In addition to the Westergren method, new automatic methods for ESR measurements have been developed. We aimed to study the reliability, precision, accuracy and stability of the Caretium XC-A30 automated ESR analyser.Methods: Ethylenediamine tetraacetic acid (EDTA)-treated blood samples were analysed via the Caretium XC-A30 automated ESR analyser and the Westergren method to compare accuracy. Precision was assessed using control samples and patient samples were classified into three groups-low, medium and high-according to their rates of sedimentation. Moreover, a stability test was performed.Results: The correlation coefficient of the results of the Caretium XC-A30 and Westergren analyses was 0.97. The correlation coefficient of ESR values obtained from the two methods assessed in the low, medium and high groups were r = 0.80, r = 0.68 and r = 0.74, respectively. The coefficient of variation of within-run (%CVw) and between-run (%CVb), with replicates performed with commercial controls samples, were 7.54% and 8.04% for the normal control and 4.68% and 3.50% for abnormal control, respectively. The %CVw obtained with patient samples in the low, medium and high groups were 10.68%, 13.13% and 4.45%, respectively. The Caretium XC-A30 measurements were stable for up to 24 h when samples were stored at 4 degrees C.Conclusion: The Caretium XC-A30 ESR analyser proved to be a suitable instrument for routine analysis of ESR.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [31] Erythrocyte sedimentation rate as a marker for coronary heart disease
    Yayan, Josef
    VASCULAR HEALTH AND RISK MANAGEMENT, 2012, 8 : 219 - 223
  • [32] Erythrocyte sedimentation rate and C-reactive protein
    Harrison, Michael
    AUSTRALIAN PRESCRIBER, 2015, 38 (03) : 93 - 94
  • [33] Erythrocyte sedimentation rate measurements using MIX-RATE® X20 and VISION A automated analyzers: Method validation and comparison study
    Prompetchara, Eakachai
    Nowaratsopon, Sunudda
    Wongkamchai, Suppakorn
    Srieakpanit, Jittra
    Ketloy, Chutitorn
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (05) : 846 - 853
  • [34] Measurement of Erythrocyte Sedimentation Rate and its Comparison by Microsed System and Westergren Method
    Rehman, Asma
    Farrukh, Ambreen
    Choudary, Munazza
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (02): : 243 - 245
  • [35] Determination of the length of sedimentation reaction (erythrocyte sedimentation rate) in non-anticoagulated blood with the Microtest 1
    Piva, E
    Fassina, P
    Plebani, M
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (07) : 713 - 717
  • [36] Evaluation of Ves-Matic Cube 200-an automated system for the measurement of the erythrocyte sedimentation rate
    Perovic, E.
    Bakovic, L.
    Valcic, A.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2010, 32 (01) : 88 - 94
  • [37] The Relationship between Erythrocyte Sedimentation Rate and Erythrocyte Sedimentation/Red Blood Cell Ratio and Disease Activity in Systemic Lupus Erythematosus
    Xiao, Li-jia
    Liu, Jin-lian
    Pan, Nan-nan
    Cui, Sheng-jin
    He, Hai-hong
    Ren, Ting-yu
    Lan, Xi
    Li, Qing
    Zhou, Yi-wen
    CLINICAL LABORATORY, 2020, 66 (03) : 309 - 317
  • [38] Laser-induced changes of in vitro erythrocyte sedimentation rate
    Mustafa S. Al Musawi
    M. S. Jaafar
    B. Al-Gailani
    Naser M. Ahmed
    Fatanah M. Suhaimi
    Lasers in Medical Science, 2017, 32 : 2089 - 2095
  • [39] Polymyalgia rheumatica with normal erythrocyte sedimentation rate:: Clinical aspects
    Martínez-Taboada, VM
    Blanco, R
    Rodríguez-Valverde, V
    CLINICAL AND EXPERIMENTAL RHEUMATOLOGY, 2000, 18 (04) : S34 - S37
  • [40] Haemoconcentration by gelatin-induced acceleration of erythrocyte sedimentation rate
    Sümpelmann, R
    Günther, A
    Zander, R
    ANAESTHESIA, 2000, 55 (03) : 217 - 220