Method Verification of the Caretium XC-A30 Automated Erythrocyte Sedimentation Rate Analyser for Erythrocyte Sedimentation Rate

被引:0
|
作者
Khongwichit, Soemwit [1 ]
Saelim, Menapha [2 ]
Na Songkhla, Yanisa [2 ]
Buncherd, Hansuk [1 ]
Nopparatana, Chawadee [2 ]
Srinoun, Kanitta [1 ,3 ]
机构
[1] Prince Songkla Univ, Fac Med Technol, Hat Yai, Thailand
[2] Prince Songkla Univ, Fac Med, Dept Pathol, Hat Yai, Thailand
[3] Prince Songkla Univ, Fac Med Technol, 15 Kanjanavanit Rd, Hat Yai 90110, Songkhla, Thailand
来源
MALAYSIAN JOURNAL OF MEDICAL SCIENCES | 2022年 / 29卷 / 05期
关键词
erythrocyte sedimentation rate; automated ESR analyser; method verification; MATIC CUBE 200; WESTERGREN METHOD;
D O I
10.21315/mjms2022.29.5.5
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: The erythrocyte sedimentation rate (ESR) analyser is widely used in haematological testing. In addition to the Westergren method, new automatic methods for ESR measurements have been developed. We aimed to study the reliability, precision, accuracy and stability of the Caretium XC-A30 automated ESR analyser.Methods: Ethylenediamine tetraacetic acid (EDTA)-treated blood samples were analysed via the Caretium XC-A30 automated ESR analyser and the Westergren method to compare accuracy. Precision was assessed using control samples and patient samples were classified into three groups-low, medium and high-according to their rates of sedimentation. Moreover, a stability test was performed.Results: The correlation coefficient of the results of the Caretium XC-A30 and Westergren analyses was 0.97. The correlation coefficient of ESR values obtained from the two methods assessed in the low, medium and high groups were r = 0.80, r = 0.68 and r = 0.74, respectively. The coefficient of variation of within-run (%CVw) and between-run (%CVb), with replicates performed with commercial controls samples, were 7.54% and 8.04% for the normal control and 4.68% and 3.50% for abnormal control, respectively. The %CVw obtained with patient samples in the low, medium and high groups were 10.68%, 13.13% and 4.45%, respectively. The Caretium XC-A30 measurements were stable for up to 24 h when samples were stored at 4 degrees C.Conclusion: The Caretium XC-A30 ESR analyser proved to be a suitable instrument for routine analysis of ESR.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [1] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [2] The comparison of erythrocyte sedimentation rate (ESR) modify Westergren Caretium Xc-A30 and Westergren Manual in Clinical Pathology Laboratory, Sanglah General Hospital, Denpasar, Bali
    Dewi, Made Minarti Witarini
    Herawati, Sianny
    Mulyantari, Ni Kadek
    Prabawa, I. Putu Yuda
    BALI MEDICAL JOURNAL, 2019, 8 (02) : 396 - 399
  • [3] Automated measurement of the erythrocyte sedimentation rate: method validation and comparison
    Lapic, Ivana
    Piva, Elisa
    Spolaore, Federica
    Tosato, Francesca
    Pelloso, Michela
    Plebani, Mario
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (09) : 1364 - 1373
  • [4] ERYTHROCYTE SEDIMENTATION RATE
    Besisik, Sevgi Kalayoglu
    NOBEL MEDICUS, 2005, 1 (01): : 4 - 8
  • [5] Analytical validation of the iSED automated analyzer for erythrocyte sedimentation rate
    Lapic, Ivana
    Milos, Marija
    Tosato, Francesca
    Piva, Elisa
    Zadro, Renata
    Rogic, Dunja
    Plebani, Mario
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (02) : 109 - 115
  • [6] Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch
    Lapic, Ivana
    Rade, Anamarija
    Kraljevic, Andela
    Milos, Marija
    Herak, Desiree Coen
    Daskijevic, Lucija
    Cerovac, Paula
    Rogic, Dunja
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (08) : 1463 - 1469
  • [7] VALIDATION OF AN ERYTHROCYTE SEDIMENTATION RATE ANALYZER WITH MODIFIED WESTERGREN METHOD
    Patil, Asha
    Nayak, Deepak M.
    Belurkar, Sushma V.
    Verma, Seemitr
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2013, 2 (03): : 284 - 289
  • [8] The TEST 1 automated system - A new method for measuring the erythrocyte sedimentation rate
    Plebani, M
    De Toni, S
    Sanzari, MC
    Bernardi, D
    Stockreiter, E
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1998, 110 (03) : 334 - 340
  • [9] Micromethod of an estimate of erythrocyte sedimentation rate
    Aristov, Alexandr A.
    Evtushenko, Gennadly S.
    Ermolovich, Dmitriy G.
    LASERS FOR MEASUREMENTS AND INFORMATION TRANSFER 2007, 2008, 7006
  • [10] EDTA samples for the measurement of erythrocyte sedimentation rate using Sedimatic 100 analyser
    Horsti, J
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2001, 39 (10) : 1001 - 1003