Validating the Use of the Mimic dV-trainer for Robotic Surgery Skill Acquisition Among Urology Residents

被引:55
作者
Korets, Ruslan
Mues, Adam C.
Graversen, Joseph A.
Gupta, Mantu
Benson, Mitchell C.
Cooper, Kimberly L.
Landman, Jaime
Badani, Ketan K. [1 ]
机构
[1] Columbia Univ, Coll Phys & Surg, Dept Urol, New York, NY 10032 USA
关键词
SIMULATION;
D O I
10.1016/j.urology.2011.07.1426
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE To compare robotic surgery skill acquisition of residents trained with Mimic dVTrainer (MdVT) and da Vinci Surgical System (dVSS) console. No standardized curriculum currently exists for robotic surgical education. The MdVT is a compact hardware platform that closely reproduces the experience of the dVSS. METHODS Sixteen urology trainees were randomized into 3 groups. A baseline evaluation using dVSS was performed and consisted of 2 exercises requiring endowrist manipulation (EM), camera movement and clutching (CC), needle control (NC), and knot-tying (KT). Groups 1 and 2 completed a standardized training curriculum on MdVT and dVSS, respectively. Group 3 received no additional training. After completion of the training phase, all trainees completed a secondary evaluation on dVSS consisting of the same exercises performed during baseline evaluation. RESULTS There was no difference in baseline performance scores across the 3 groups. Although Group 3 showed no significant improvement in EM/CC domain (P = .15), Groups 1 and 2 had statistically significant improvement in EM/CC domain (P = .039 and P = .007, respectively). The difference in improvement between Groups 1 and group 2 was not statistically different (P = .21). Only Group 2 trainees showed significant improvement in the NC and KT domains during secondary evaluation (P = .02). CONCLUSION Curriculum-based training with MdVT or dVSS significantly improves robotic surgery aptitude. Similar improvements are seen for exercise domains shared between MdVT and dVSS groups. Follow-up studies are necessary to assess the efficacy of MdVT over a wider spectrum of domains. UROLOGY 78: 1326-1331, 2011. (C) 2011 Elsevier Inc.
引用
收藏
页码:1326 / 1330
页数:5
相关论文
共 19 条
[1]   Virtual reality-assisted robotic surgery simulation [J].
Albani, Justin M. ;
Lee, David I. .
JOURNAL OF ENDOUROLOGY, 2007, 21 (03) :285-287
[2]   Robotics in urologic surgery: An evolving new technology [J].
Atug, Fatih ;
Castle, Erik P. ;
Woods, Michael ;
Davis, Rodney ;
Thomas, Raju .
INTERNATIONAL JOURNAL OF UROLOGY, 2006, 13 (07) :857-863
[3]   Hybrid augmented reality simulator: Preliminary construct validation of laparoscopic smoothness in a urology residency program [J].
Feifer, Andrew ;
Delisle, Josee ;
Anidjar, Maurice .
JOURNAL OF UROLOGY, 2008, 180 (04) :1455-1459
[4]   Qualitative and quantitative analysis of the learning curve of a simulated surgical task on the da Vinci system [J].
Hernandez, JD ;
Bann, SD ;
Munz, Y ;
Moorthy, K ;
Datta, V ;
Martin, S ;
Dosis, A ;
Bello, F ;
Darzi, A ;
Rockall, T .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2004, 18 (03) :372-378
[5]   FACE, CONTENT, AND CONSTRUCT VALIDITY OF A NOVEL DA VINCI® SURGICAL SIMULATOR [J].
Hung, Andrew J. ;
Zehnder, Pascal ;
Patil, Mukul B. ;
Ng, Casey ;
Gill, Inderbir S. ;
Desai, Mihir M. .
JOURNAL OF UROLOGY, 2011, 185 (04) :E595-E595
[6]   ProMIS™ Can Serve as a da Vinci® Simulator-A Construct Validity Study [J].
Jonsson, Martin N. ;
Mahmood, Mahmood ;
Askerud, Tomas ;
Hellborg, Henrik ;
Ramel, Stig ;
Wiklund, N. Peter ;
Kjellman, Magnus ;
Ahlberg, Gunnar .
JOURNAL OF ENDOUROLOGY, 2011, 25 (02) :345-350
[7]   Face, Content, and Construct Validity of dV-Trainer, a Novel Virtual Reality Simulator for Robotic Surgery [J].
Kenney, Patrick A. ;
Wszolek, Matthew F. ;
Gould, Justin J. ;
Libertino, John A. ;
Moinzadeh, Alireza .
UROLOGY, 2009, 73 (06) :1288-1292
[8]   The current role of medical simulation in American urological residency training programs: An assessment by program directors [J].
Le, Carter Q. ;
Lightner, Deborah J. ;
VanderLei, Laura ;
Segura, Joseph W. ;
Gettman, Matthew T. .
JOURNAL OF UROLOGY, 2007, 177 (01) :288-291
[9]   Best Practices for Robotic Surgery Training and Credentialing [J].
Lee, Jason Y. ;
Mucksavage, Phillip ;
Sundaram, Chandru P. ;
McDougall, Elspeth M. .
JOURNAL OF UROLOGY, 2011, 185 (04) :1191-1197
[10]   Does Training on a Virtual Reality Robotic Simulator Improve Performance on the da Vinci® Surgical System? [J].
Lerner, Michelle A. ;
Ayalew, Mikias ;
Peine, William J. ;
Sundaram, Chandru P. .
JOURNAL OF ENDOUROLOGY, 2010, 24 (03) :467-472