Comparison of implant stability measurement devices for bone-anchored hearing aid systems

被引:0
作者
Westover, Lindsey [1 ]
Faulkner, Gary [2 ]
Hodgetts, William [3 ]
Raboud, Don [1 ]
机构
[1] Univ Alberta, Dept Mech Engn, Edmonton, AB, Canada
[2] Glenrose Rehabil Hosp, Rehabil Res & Technol Dev, Edmonton, AB, Canada
[3] Univ Alberta, Commun Sci & Disorders, Rehabil Med, Edmonton, AB, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
RESONANCE FREQUENCY-ANALYSIS; OSSEOINTEGRATED IMPLANTS; CONDUCTION DEVICES; SURVIVAL; COMPLICATIONS; TOLERABILITY;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. The success of implants for bone-anchored hearing aids (BAHA) relies on proper osseointegration at the bone-implant interface. Clinical evaluation of implant stability is important in prescribing loading, identifying the risk of failure, and monitoring the long-term health of the implant. Purpose. The purpose of this in vitro study was to evaluate 2 measurement systems for BAHA implant stability: the most commonly used, Osstell implant stability quotient (ISQ), and a newly developed advance system for implant stability testing (ASIST). Material and methods. BAHA implants (Oticon Medical Ponto and Cochlear BAHA Connect systems) were installed in plastic materials with adhesive to simulate implants integrated in bone with varying levels of interface stiffness. Different lengths of BAHA abutments were used with each implant specimen, and stability measurements were obtained with both the Osstell ISQ and the ASIST systems. The measurement systems were evaluated in terms of sensitivity to differences in interface stiffness and the effect of abutment length on the stability measurement. Repeated measures ANOVA followed by post hoc t tests were used for the comparisons with a Bonferroni adjusted alpha value of.05/15 = .003 to control for potential type 1 errors. Results. Changing the abutment length of a single implant installation had minimal effect on the ASIST stability coefficient, whereas large variations were observed in the Osstell implant stability quotient (ISQ). The Osstell showed a clear relationship of decreasing ISQ with increasing abutment length for both the Oticon Medical and the Cochlear implant systems. Both the ASIST and the Osstell were found to be sensitive to changes in interface properties, with the ASIST being more sensitive to these changes. Conclusions. The ASIST system is more sensitive to changes in interface properties and shows smaller variation because of changes in abutment length than the Osstell ISQ system.
引用
收藏
页码:178 / 184
页数:7
相关论文
共 50 条
  • [21] First Report: Linear Incision for Placement of a Magnetically Coupled Bone-Anchored Hearing Implant
    Barry, Jonnae Y.
    Reghunathan, Saranya
    Jacob, Abraham
    OTOLOGY & NEUROTOLOGY, 2017, 38 (02) : 221 - 224
  • [22] Clinical Outcome of a Wide-diameter Bone-anchored Hearing Implant and a Surgical Technique With Tissue Preservation
    Mowinckel, Marius S.
    Moller, Martin N.
    Wielandt, Kirsten N.
    Foghsgaard, Soren
    OTOLOGY & NEUROTOLOGY, 2016, 37 (04) : 374 - 379
  • [23] The Birmingham Pediatric Bone-Anchored Hearing Aid Program: A 15-Year Experience
    McDermott, Ann-Louise
    Williams, Jo
    Kuo, Michael
    Reid, Andrew
    Proops, David
    OTOLOGY & NEUROTOLOGY, 2009, 30 (02) : 178 - 183
  • [24] Percutaneous bone-anchored hearing implant surgery: dermatome versus linear incision technique
    Strijbos, Ruben M.
    Bom, Steven J. H.
    Zwerver, Stefan
    Hol, Myrthe K. S.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2017, 274 (01) : 109 - 117
  • [25] A Retrospective Review of Temporal Bone Imaging With Respect to Bone-Anchored Hearing Aid Placement
    Baker, Aaron R.
    Fanelli, David G.
    Kanekar, Sangam
    Isildak, Huseyin
    OTOLOGY & NEUROTOLOGY, 2017, 38 (01) : 86 - 88
  • [26] A New Wide-Diameter Bone-Anchored Hearing Implant-Prospective 1-Year Data on Complications, Implant Stability, and Survival
    Foghsgaard, Slren
    Caye-Thomasen, Per
    OTOLOGY & NEUROTOLOGY, 2014, 35 (07) : 1238 - 1241
  • [27] Benefit and Quality of Life After Bone-Anchored Hearing Aid Fitting in Children With Unilateral or Bilateral Hearing Impairment
    de Wolf, Maarten J. F.
    Hol, Myrthe K. S.
    Mylanus, Emmanuel A. M.
    Snik, Ad F. M.
    Cremers, Cor W. R. J.
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2011, 137 (02) : 130 - 138
  • [28] The Viability of Percutaneous Bone-Anchored Hearing Systems in the HIV Population
    Kuschke, Silva
    Honnet, Simon
    Scott, Alex J.
    Labuschagne, Sherene
    Pithey, Sharon
    Solomons, Zane
    Meyer, Estie
    OTOLOGY & NEUROTOLOGY, 2018, 39 (10) : E1064 - E1068
  • [29] A new bone-anchored hearing implant: short-term retrospective data on implant survival and subjective benefit
    Nelissen, Rik C.
    Mylanus, Emmanuel A. M.
    Kunst, Henricus P. M.
    Pennings, Ronald J. E.
    Snik, Ad F. M.
    Hol, Myrthe K. S.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2013, 270 (12) : 3019 - 3025
  • [30] Percutaneous bone-anchored hearing implant surgery: inside or outside the line of incision?
    Strijbos, Ruben M.
    den Besten, Christine A.
    Mylanus, Emmanuel A. M.
    Hol, Myrthe K. S.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2016, 273 (11) : 3713 - 3722