Biomechanical strength of the Peri-Loc® proximal tibial plate:: A comparison of all-locked versus hybrid locked/nonlocked screw configurations

被引:16
作者
Estes, Chris [1 ]
Rhee, Peter [1 ]
Shrader, M. Wade [1 ]
Csavina, Kristine [1 ]
Jacofsky, Marc C. [1 ]
Jacofsky, David J. [1 ]
机构
[1] Ctr Orthoped Res, Sun Hlth Res Inst, Sun City W, AZ 85375 USA
关键词
tibia; fracture; plate; strength; biomechanics;
D O I
10.1097/BOT.0b013e31817279b8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: The purpose of this study was to compare the biomechanical properties of a contoured locking plate instrumented with either an all-locked or hybrid locked/nonlocked screw construct in a proximal metaphyseal fracture of the tibia (AO 41-A3.2). Methods: A standardized proximal metaphyseal wedge osteotomy (AO 41-A3.2) was created in five pairs of cadaveric tibia. Each pair was randomly instrumented with either an all-locked or combination locked/nonlocked screw construct using a locked contoured periarticular plate (Peri-Loc (R) periarticular locked plating system, Smith & Nephew, Memphis, TN). Vertical subsidence (irreversible deformation) and deflection (reversible deformation) in each pair were analyzed and compared. Load to failure, defined by complete fracture gap closure, was also determined. Results: There was no statistically significant difference in vertical subsidence (P = 0.19) or deflection (P = 0.19) of the proximal tibia between the all-locked and combination locked/nonlocked screw construct with increasing levels of cyclical axial load from 200 to 1200 N. Failure occurred at a mean value of 2160 N in the locked group and 1760 N in the hybrid group (P = 0.19); the failure mode was plate bending in all specimens. Conclusions: The results indicate that the use of compression screws with locked screws in this particular construct allows a similar amount of irreversible and reversible deformation in response to an axial load when compared to an all-locked screw construct. This suggests that there is no statistically significant difference in the stability in fixation between the two methods, allowing the surgeon the freedom to choose the appropriate screw combination unique to each fracture.
引用
收藏
页码:312 / 316
页数:5
相关论文
共 18 条
  • [1] Extraosseous blood supply of the tibia and the effects of different plating techniques: A human cadaveric study
    Borrelli, J
    Prickett, W
    Song, E
    Becker, D
    Ricci, W
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (10) : 691 - 695
  • [2] Less Invasive Stabilization System (LISS) for fractures of the proximal tibia: indications, surgical technique and preliminary results of the UMC clinical trial
    Cole, PA
    Zlowodzki, M
    Kregor, PJ
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 : S16 - S29
  • [3] Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and bone
    Cordey, J
    Borgeaud, M
    Perren, SM
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 : 21 - 28
  • [4] Biomechanics of locked plates and screws
    Egol, KA
    Kubiak, EN
    Fulkerson, E
    Kummer, FJ
    Koval, KJ
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) : 488 - 493
  • [5] Minimally invasive plate osteosynthesis: Does percutaneous plating disrupt femoral blood supply less than the traditional technique?
    Farouk, O
    Krettek, C
    Miclau, T
    Schandelmaier, P
    Guy, P
    Tscherne, H
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1999, 13 (06) : 401 - 406
  • [6] Development of the Locking Compression Plate
    Frigg, R
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 : 6 - 10
  • [7] Fixation of diaphyseal fractures with a segmental defect: A biomechanical comparison of locked and conventional plating techniques
    Fulkerson, E
    Egol, KA
    Kubiak, EN
    Liporace, F
    Kummer, FJ
    Koval, KJ
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 60 (04): : 830 - 835
  • [8] Guidelines for the clinical application of the LCP
    Gautier, E
    Sommer, C
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 : 63 - 76
  • [9] Gray F B, 1998, Am J Orthop (Belle Mead NJ), V27, P558
  • [10] Biomechanical evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures
    Marti, A
    Fankhauser, C
    Frenk, A
    Cordey, J
    Gasser, B
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (07) : 482 - 487