A Novel Non-Invasive Thermometer for Continuous Core Body Temperature: Comparison with Tympanic Temperature in an Acute Stroke Clinical Setting

被引:23
作者
Ajcevic, Milos [1 ,2 ]
Stella, Alex Buoite [1 ]
Furlanis, Giovanni [1 ]
Caruso, Paola [1 ]
Naccarato, Marcello [1 ]
Accardo, Agostino [2 ]
Manganotti, Paolo [1 ]
机构
[1] Univ Trieste, Cattinara Univ Hosp ASUGI, Dept Med Surg & Hlth Sci, Clin Unit Neurol, Str Fiume, I-44734149 Trieste, Italy
[2] Univ Trieste, Dept Engn & Architecture, Via A Valerio, I-1034127 Trieste, Italy
关键词
continuous non-invasive temperature monitoring; core temperature; clinical thermometry; acute stroke; heat flux; wearable device; CRITICALLY-ILL PATIENTS; TEMPORAL-ARTERY; FEVER; CARE; ACCURACY; BLADDER; ESOPHAGEAL; AXILLARY; SENSOR; ADULT;
D O I
10.3390/s22134760
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is a growing research interest in wireless non-invasive solutions for core temperature estimation and their application in clinical settings. This study aimed to investigate the use of a novel wireless non-invasive heat flux-based thermometer in acute stroke patients admitted to a stroke unit and compare the measurements with the currently used infrared (IR) tympanic temperature readings. The study encompassed 30 acute ischemic stroke patients who underwent continuous measurement (Tcore) with the novel wearable non-invasive CORE device. Paired measurements of Tcore and tympanic temperature (Ttym) by using a standard IR-device were performed 3-5 times/day, yielding a total of 305 measurements. The predicted core temperatures (Tcore) were significantly correlated with Ttym (r = 0.89, p < 0.001). The comparison of the Tcore and Ttym measurements by Bland-Altman analysis showed a good agreement between them, with a low mean difference of 0.11 +/- 0.34 degrees C, and no proportional bias was observed (B = -0.003, p = 0.923). The Tcore measurements correctly predicted the presence or absence of Ttym hyperthermia or fever in 94.1% and 97.4% of cases, respectively. Temperature monitoring with a novel wireless non-invasive heat flux-based thermometer could be a reliable alternative to the Ttym method for assessing core temperature in acute ischemic stroke patients.
引用
收藏
页数:10
相关论文
共 65 条
[1]   e-Health solution for home patient telemonitoring in early post-acute TIA/ Minor stroke during COVID-19 pandemic [J].
Ajcevic, Milos ;
Furlanis, Giovanni ;
Naccarato, Marcello ;
Caruso, Paola ;
Polverino, Paola ;
Marsich, Alessandro ;
Accardo, Agostino ;
Manganotti, Paolo .
INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2021, 152
[2]   Perioperative measurement of core body temperature using an unobtrusive passive heat flow sensor [J].
Atallah, Louis ;
Ciuhu, Calina ;
Paulussen, Igor ;
Bongers, Edwin ;
Blom, Antoon H. M. ;
Idrissi, Akram ;
Noordergraaf, Gerrit .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2020, 34 (06) :1351-1359
[3]   Unobtrusive Monitoring of Neonatal Brain Temperature Using a Zero-Heat-Flux Sensor Matrix [J].
Atallah, Louis ;
Bongers, Edwin ;
Lamichhane, Bishal ;
Bambang-Oetomo, Sidarto .
IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2016, 20 (01) :100-107
[4]  
Bindu Barkha, 2017, J Anaesthesiol Clin Pharmacol, V33, P306, DOI 10.4103/joacp.JOACP_334_16
[5]   Accuracy of zero-heat-flux thermometry and bladder temperature measurement in critically ill patients [J].
Braeuer, Anselm ;
Fazliu, Albulena ;
Perl, Thorsten ;
Heise, Daniel ;
Meissner, Konrad ;
Brandes, Ivo Florian .
SCIENTIFIC REPORTS, 2020, 10 (01)
[6]   Heat risk exacerbation potential for neurology patients during the COVID-19 pandemic and related isolation [J].
Buoite Stella, Alex ;
Filingeri, Davide ;
Ravanelli, Nicholas ;
Morrison, Shawnda A. ;
Ajcevic, Milos ;
Furlanis, Giovanni ;
Manganotti, Paolo .
INTERNATIONAL JOURNAL OF BIOMETEOROLOGY, 2021, 65 (04) :627-630
[7]   Return to school in the COVID-19 era: considerations for temperature measurement [J].
Buoite Stella A. ;
Manganotti P. ;
Furlanis G. ;
Accardo A. ;
Ajčević M. .
Journal of Medical Engineering and Technology, 2020, 44 (08) :468-471
[8]   Measurement of Core Body Temperature Using Graphene-Inked Infrared Thermopile Sensor [J].
Chaglla, Jorge S. E. ;
Celik, Numan ;
Balachandran, Wamadeva .
SENSORS, 2018, 18 (10)
[9]   Investigation of the Impact of Infrared Sensors on Core Body Temperature Monitoring by Comparing Measurement Sites [J].
Chen, Hsuan-Yu ;
Chen, Andrew ;
Chen, Chiachung .
SENSORS, 2020, 20 (10)
[10]  
Chen WX, 2019, BIOMED ENG LETT, V9, P3, DOI 10.1007/s13534-019-00102-2