Analysis of BTEX and other substituted benzenes in water using headspace SPME-GC-FID: method validation

被引:53
作者
Almeida, CMM
Boas, LV
机构
[1] Univ Lisbon, Fac Farm, Lab Hidrol & Anal Hidrol, P-1649019 Lisbon, Portugal
[2] Inst Tecnol Quim & Biol, P-2784505 Oeiras, Portugal
[3] Inst Super Tecn, P-1049001 Lisbon, Portugal
来源
JOURNAL OF ENVIRONMENTAL MONITORING | 2004年 / 6卷 / 01期
关键词
D O I
10.1039/b307053k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The analysis of BTEX and other substituted benzenes in water samples using solid phase microextraction ( SPME) and quantification by gas chromatography with flame ionization detection (GC-FID) was validated. The best analytical conditions were obtained using PDMS/DVB/CAR fibre using headspace extraction (HS-SPME) at 50 degreesC for 20 min without stirring. The linear range for each compound by HS-SPME with GC/FID was defined. The detection limits for these compounds obtained with PDMS/DVB/CAR fibre and GC/FID were: benzene (15 ng L-1), toluene (160 ng L-1), monochlorobenzene (54 ng L-1), ethylbenzene (32 ng L-1), m-xylene (56 ng L-1), p-xylene (69 ng L-1), styrene (35 ng L-1), o-xylene (42 ng L-1), m-dichlorobenzene (180 ng L-1), p-dichlorobenzene (230 ng L-1), o-dichlorobenzene (250 ng L-1) and trichlorobenzene (260 ng L-1). This headspace SPME-GC-FID method was compared with a previously validated method of analysis using closed-loop-stripping analysis (CLSA). The headspace SPME-GC-FID method is suitable for monitoring the production and distribution of potable water and was used, in field trials, for the analysis of samples from main intakes of water ( surface or underground) and from the water supply system of a large area ( Lisbon and neighbouring municipalities).
引用
收藏
页码:80 / 88
页数:9
相关论文
共 32 条
  • [1] ALMEIDA C, 1996, REV IND AGUA, V21, P14
  • [2] ALMEIDA C, 1998, OPTIMIZACAO CONDICOE
  • [3] Alpendurada MD, 2000, J CHROMATOGR A, V889, P3
  • [4] *APHA AWWA WEF, 1995, STAND METH EX W 6000
  • [5] *AWWA, 1999, WAT QUAL TREATM HDB, pCH2
  • [6] Calibration of a commercial solid-phase microextraction device for measuring headspace concentrations of organic volatiles
    Bartelt, RJ
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (03) : 364 - 372
  • [7] Monitoring of ambient BTX at Monterotondo (Rome) and indoor-outdoor evaluation in school and domestic sites
    Bertoni, G
    Ciuchini, C
    Pasini, A
    Tappa, R
    [J]. JOURNAL OF ENVIRONMENTAL MONITORING, 2002, 4 (06): : 903 - 909
  • [8] DAVID WP, 1992, J CHROMATOGR, V625, P247
  • [9] Effect of the various parameters governing solid-phase microextraction for the trace-determination of pesticides in water
    Dugay, J
    Miège, C
    Hennion, MC
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1998, 795 (01) : 27 - 42
  • [10] FABRICIO VP, 2002, J CHROMATOGR SCI, V3, P122