Complete experimental and theoretical characterization of nonlinear concentration gradient generator microfluidic device for analytical purposes

被引:4
作者
Buzzetti, Paulo Henrique Maciel [1 ,2 ]
Taniguchi, Maiara Mitiko [1 ]
de Souza Mendes, Nayara [1 ]
Vicentino, Renata Correa [3 ]
de Oliveira, Jean Halison [1 ]
Junior, Bento Pereira Cabral [1 ]
de Souza, Marcos [3 ]
Monteiro, Johny Paulo [4 ]
Girotto, Emerson Marcelo [1 ]
机构
[1] Univ Estadual Maringa, Dept Chem, Mat Chem & Sensors Labs LMSEN, Colombo 5790, BR-87020900 Maringa, PR, Brazil
[2] Univ Grenoble Alpes, CNRS, DCM, UMR 5250, F-38000 Grenoble, France
[3] Univ Estadual Maringa, Dept Chem Engn, Colombo 5790, BR-87020900 Maringa, PR, Brazil
[4] Univ Tecnol Fed Parana, Dept Chem, Marcilio Dias 635, BR-86812460 Apucarana, PR, Brazil
关键词
Microfluidic devices; Automated calibration; Gradient generator; Optical biosensor; SPR; Nanohole arrays; OPTICAL-TRANSMISSION; CELL-CULTURE; SURFACE; ARRAY; CHIP; BIOSENSOR; FLOW;
D O I
10.1007/s00604-021-05110-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microfluidic devices that generate stable concentration gradients are efficient instruments for automated calibration for analytical and bioanalytical systems. However, little attention has been paid to the development of reusable microfluidic concentration gradient generators, which can be useful for a range of species through mathematical characterization. In this work, we develop a microfluidic device based on three steps of serial dilution that were able to generate nonlinear concentration gradient for dyes and biomolecules. The microfluidic device was described mathematically, statistically and was suitable for reusable analytical and bioanalytical analysis. The device reproducibility was assessed by experimental tests, which have shown the same gradient concentration profile for different dyes and statistical reproducibility with 95% confidence interval for bovine serum albumin (BSA). Moreover, the experimental data converged well with those obtained by computational fluid dynamics simulation. Applicability was verified by coupling the microfluidic device to a surface plasmon resonance (SPR) biosensor, based on nanohole arrays with sensitivity of 358.7 nm RIU-1 determined by white-light SPR excitation exposed to different D-(+)-glucose aqueous solutions with 1.3361-1.4035 refractive index interval. The transmission light intensities obtained by the array of images allowed to quantify a pseudo-unknown BSA sample (160 mu g mL(-1)) at 138 mu g mL(-1). The SPR analysis has been validated in parallel by fluorescence emissions, which showed a concentration of 154.8 +/- 16.6 mu g mL(-1).
引用
收藏
页数:9
相关论文
共 34 条
[11]   Logarithmic bacterial gradient chip for analyzing the effects of dietary restriction on C. elegans growth [J].
Ge, Anle ;
Hu, Liang ;
Wang, Xixian ;
Zhu, Jinchi ;
Feng, Xiaojun ;
Du, Wei ;
Liu, Bi-Feng .
SENSORS AND ACTUATORS B-CHEMICAL, 2018, 255 :735-744
[12]  
Genet C., 2009, REV NAT J, V445, P212, DOI [10.1142/9789814287005_0021, DOI 10.1142/9789814287005_0021]
[13]   Mixing characteristics of T-type microfluidic mixers [J].
Gobby, D ;
Angeli, P ;
Gavriilidis, A .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2001, 11 (02) :126-132
[14]  
Hu Chunfei., 2017, Biochem. Anal Biochem, V06, DOI [10.4172/2161-1009.1000320, DOI 10.4172/2161-1009.1000320]
[15]   SERS-based droplet microfluidics for high-throughput gradient analysis [J].
Jeon, Jinhyeok ;
Choi, Namhyun ;
Chen, Hao ;
Moon, Joung-Il ;
Chen, Lingxin ;
Choo, Jaebum .
LAB ON A CHIP, 2019, 19 (04) :674-681
[16]   IMMOBILIZATION OF PROTEINS TO A CARBOXYMETHYLDEXTRAN-MODIFIED GOLD SURFACE FOR BIOSPECIFIC INTERACTION ANALYSIS IN SURFACE-PLASMON RESONANCE SENSORS [J].
JOHNSSON, B ;
LOFAS, S ;
LINDQUIST, G .
ANALYTICAL BIOCHEMISTRY, 1991, 198 (02) :268-277
[17]   Quantitative analysis of molecular interaction in a microfluidic channel: The T-sensor [J].
Kamholz, AE ;
Weigl, BH ;
Finlayson, BA ;
Yager, P .
ANALYTICAL CHEMISTRY, 1999, 71 (23) :5340-5347
[18]   Neutrophil Chemotaxis within a Competing Gradient of Chemoattractants [J].
Kim, Donghyuk ;
Haynes, Christy L. .
ANALYTICAL CHEMISTRY, 2012, 84 (14) :6070-6078
[19]   Scalable static droplet array for biochemical assays based on concentration gradients [J].
Lee, Byungjin ;
Jin, Si Hyung ;
Noh, Young-Moo ;
Jeong, Seong-Geun ;
Jeong, Heon-Ho ;
Lee, Chang-Soo .
SENSORS AND ACTUATORS B-CHEMICAL, 2018, 273 :1572-1578
[20]   Generalized serial dilution module for monotonic and arbitrary microfluidic gradient generators [J].
Lee, Kangsun ;
Kim, Choong ;
Ahn, Byungwook ;
Panchapakesan, Rajagopal ;
Full, Anthony R. ;
Nordee, Ledum ;
Kang, Ji Yoon ;
Oh, Kwang W. .
LAB ON A CHIP, 2009, 9 (05) :709-717