Impedance characterization and modeling of electrodes for biomedical applications

被引:495
作者
Franks, W [1 ]
Schenker, I
Schmutz, P
Hierlemann, A
机构
[1] ETH, Phys Elect Lab, CH-8093 Zurich, Switzerland
[2] Swiss Fed Inst Mat Testing & Res, EMPA, Lab Corros & Mat Integr, CH-8600 Dubendorf, Switzerland
关键词
electrochemical impedance spectroscopy; Pt; black; and TiN bioelectrodes;
D O I
10.1109/TBME.2005.847523
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A low electrode-electrolyte impedance interface is critical in the design of electrodes for biomedical applications. To design low-impedance interfaces a complete understanding of the physical processes contributing to the impedance is required. In this work a model describing these physical processes is validated and extended to quantify the effect of organic coatings and incubation time. Electrochemical impedance spectroscopy has been used to electrically characterize the interface for various electrode materials: platinum, platinum black, and titanium nitride; and varying electrode sizes: 1 cm(2), and 900 mu m(2). An equivalent circuit model comprising an interface capacitance, shunted by a charge transfer resistance, in series with the solution resistance has been fitted to the experimental results. Theoretical equations have been used to calculate the interface capacitance impedance and the solution resistance, yielding results that correspond well with the fitted parameter values, thereby confirming the validity of the equations. The effect of incubation time, and two organic cell-adhesion promoting coatings, poly-L-lysine and laminin, on the interface impedance has been quantified using the model. This demonstrates the benefits of using this model in developing a better understanding of the physical processes occurring at the interface in more complex, biomedically relevant situations.
引用
收藏
页码:1295 / 1302
页数:8
相关论文
共 35 条