Automatic anesthesia depth staging using entropy measures and relative power of electroencephalogram frequency bands

被引:2
作者
Jahanseir, Mercedeh [1 ]
Setarehdan, Seyed Kamaledin [2 ]
Momenzadeh, Sirous [3 ]
机构
[1] Islamic Azad Univ, Dept Biomed Engn, Sci & Res Branch, Tehran, Iran
[2] Univ Tehran, Coll Engn, Sch Elect & Comp Engn, Control & Intelligent Proc Ctr Excellence, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Funct Neurosurg Res Ctr, Tehran, Iran
关键词
Electroencephalogram; Anesthesia; Entropy; Power spectra; LS-SVM; DETRENDED FLUCTUATION ANALYSIS; PERMUTATION ENTROPY; EEG; SEVOFLURANE; INDUCTION; AWARENESS; DYNAMICS; SIGNALS;
D O I
10.1007/s13246-018-0688-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many of the surgeries performed under general anesthesia are aided by electroencephalogram (EEG) monitoring. With increased focus on detecting the anesthesia states of patients in the course of surgery, more attention has been paid to analyzing the power spectra and entropy measures of EEG signal during anesthesia. In this paper, by using the relative power of EEG frequency bands and the EEG entropy measures, a new method for detecting the depth of anesthesia states has been presented based on the least squares support vector machine (LS-SVM) classifiers. EEG signals were recorded from 20 patients before, during and after general anesthesia in the operating room at a sampling rate of 200Hz. Then, 12 features were extracted from each EEG segment, 10s in length, which are used for anesthesia state monitoring. No significant difference was observed (p>0.05) between these features and the bispectral index (BIS), which is the commonly used measure of anesthetic effect. The used LS-SVM classifier based method is able to identify the anesthesia states with an accuracy of 80% with reference to the BIS index. Since the underlying equation of the utilized LS-SVM is linear, the computational time of the algorithm is not significant and therefore it can be used for online application in operation rooms.
引用
收藏
页码:919 / 929
页数:11
相关论文
共 35 条
  • [1] Automatic anesthesia depth staging using entropy measures and relative power of electroencephalogram frequency bands
    Mercedeh Jahanseir
    Seyed Kamaledin Setarehdan
    Sirous Momenzadeh
    Australasian Physical & Engineering Sciences in Medicine, 2018, 41 : 919 - 929
  • [2] Depth of Anesthesia Indicator Using combination of Complexity and frequency Measures
    Shalbaf, R.
    Mehrnam, A. H.
    Behnam, H.
    2014 21TH IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING (ICBME), 2014, : 156 - 160
  • [3] Measuring the Depth of Anesthesia Using Ordinal Power Spectral Density of Electroencephalogram
    Choi, Ho-Ik
    Noh, Gyu-Jeong
    Shin, Hyun-Chool
    IEEE ACCESS, 2020, 8 : 50431 - 50438
  • [4] Monitoring the depth of anesthesia using entropy features and an artificial neural network
    Shalbaf, Reza
    Behnam, Hamid
    Sleigh, Jamie W.
    Steyn-Ross, Alistair
    Voss, Logan J.
    JOURNAL OF NEUROSCIENCE METHODS, 2013, 218 (01) : 17 - 24
  • [5] The Brain Function Index as a Depth of Anesthesia Indicator Using Complexity Measures
    Shalbaf, R.
    Behnam, H.
    Moghadam, H. Jelveh
    Mehrnam, A.
    Sadaghiani, M.
    2013 IEEE CONFERENCE ON SYSTEMS, PROCESS & CONTROL (ICSPC), 2013, : 68 - 72
  • [6] Classification of Depth of Coma Using Complexity Measures and Nonlinear Features of Electroencephalogram Signals
    Altintop, Cigdem Guluzar
    Latifoglu, Fatma
    Akin, Aynur Karayol
    Bayram, Adnan
    Ciftci, Murat
    INTERNATIONAL JOURNAL OF NEURAL SYSTEMS, 2022, 32 (05)
  • [7] The effect of age on electroencephalogram measures of anesthesia hypnosis: A comparison of BIS, Alpha Power, Lempel-Ziv complexity and permutation entropy during propofol induction
    Biggs, Daniela
    Boncompte, Gonzalo
    Pedemonte, Juan C. C.
    Fuentes, Carlos
    Cortinez, Luis I. I.
    FRONTIERS IN AGING NEUROSCIENCE, 2022, 14
  • [8] Changes in complexities and power spectrum of rat electroencephalogram under various Anesthesia depth
    Feng, ZY
    Zheng, XX
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 145 - 146
  • [9] A COMPARISON OF MEDIAN FREQUENCY, SPECTRAL EDGE FREQUENCY, A FREQUENCY BAND POWER RATIO, TOTAL POWER, AND DOMINANCE SHIFT IN THE DETERMINATION OF DEPTH OF ANESTHESIA
    DRUMMOND, JC
    BRANN, CA
    PERKINS, DE
    WOLFE, DE
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1991, 35 (08) : 693 - 699
  • [10] Measuring and Reflecting Depth of Anesthesia Using Wavelet and Power Spectral Density
    Tai Nguyen-Ky
    Wen, Peng
    Li, Yan
    Gray, Robert
    IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2011, 15 (04): : 630 - 639