Current Preclinical Testing of New Hip Arthroplasty Technologies Does Not Reflect Real-World Loadings: Capturing Patient-Specific and Activity-Related Variation in Hip Contact Forces

被引:26
作者
Lunn, David E. [1 ,2 ]
De Pieri, Enrico [3 ]
Chapman, Graham J. [1 ,2 ,4 ]
Lund, Morten E. [5 ]
Redmond, Anthony C. [1 ,2 ]
Ferguson, Stephen J. [3 ]
机构
[1] Univ Leeds, Inst Rheumat & Musculoskeletal Med, Leeds, W Yorkshire, England
[2] NIHR Leeds Biomed Res Ctr, Leeds, W Yorkshire, England
[3] Swiss Fed Inst Technol, Inst Biomech, Zurich, Switzerland
[4] Univ Cent Lancashire, Sch Hlth Sci, Preston, Lancs, England
[5] AnyBody Technol AS, Aalborg, Denmark
基金
英国工程与自然科学研究理事会;
关键词
total hip arthroplasty; hip contact force; functional outcomes; activities of daily living; biomechanics; KNEE ARTHROPLASTY; IMPLANT REVISION; WEAR; REPLACEMENT; SURGERY; RISK; GAIT; PROJECTIONS; SIMULATOR; PATTERNS;
D O I
10.1016/j.arth.2019.10.006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Total hip arthroplasty (THA) implants are routinely tested for their tribological performance through regulatory preclinical wear testing (eg, ISO-14242). The standardized loading conditions defined in these tests consist of simplified waveforms, which do not specifically represent in vivo loads in different groups of patients. The aim of this study is to investigate, through musculoskeletal modeling, patient-specific and activity-related variation in hip contact forces (HCFs) in a large cohort of THA patients during common activities of daily living (ADLs). Methods: A total of 132 THA patients participated in a motion-capture analysis while performing different ADLs, including walk, fast walk, stair ascent, and descent (locomotor); sit to stand, stand to sit, squat, and lunge (nonlocomotor). HCFs were then calculated using the AnyBody Modeling System and qualitatively compared across all activities. The influence of gender on HCFs was analyzed through statistical parametric mapping analysis. Results: Systematic differences were found in HCF magnitudes and individual components in both locomotor and nonlocomotor ADLs. The qualitative analysis of the ADLs revealed a large range and a large variability in forces experienced at the hip during different activities. Significant differences in the 3-dimensional loading patterns were observed between males and females across most activities. Conclusion: THA patients present a large variability in the forces experienced at the hip joint during their daily life. The interpatient variation might partially explain the heterogeneity observed in implant survival rates. A more extensive preclinical implant testing standard under clinically relevant loading conditions has been advocated to better predict and avoid clinical wear problems. (c) 2019 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:877 / 885
页数:9
相关论文
共 46 条
[1]   Evaluation of the accuracy of three popular regression equations for hip joint centre estimation using computerised tomography measurements for metal-on-metal hip resurfacing arthroplasty patients [J].
Andersen, Michael S. ;
Mellon, Stephen ;
Grammatopoulos, George ;
Gill, Harinderjit S. .
GAIT & POSTURE, 2013, 38 (04) :1044-1047
[2]   Prediction of Polyethylene Wear Rates from Gait Biomechanics and Implant Positioning in Total Hip Replacement [J].
Ardestani, Marzieh M. ;
Amenabar Edwards, Pedro P. ;
Wimmer, Markus A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2017, 475 (08) :2027-2042
[3]   The effect of patient age at intervention on risk of implant revision after total replacement of the hip or knee: a population-based cohort study [J].
Bayliss, Lee E. ;
Culliford, David ;
Monk, A. Paul ;
Glyn-Jones, Sion ;
Prieto-Alhambra, Daniel ;
Judge, Andrew ;
Cooper, Cyrus ;
Carr, Andrew J. ;
Arden, Nigel K. ;
Beard, David J. ;
Price, Andrew J. .
LANCET, 2017, 389 (10077) :1424-1430
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]   Standardized Loads Acting in Hip Implants [J].
Bergmann, Georg ;
Bender, Alwina ;
Dymke, Joern ;
Duda, Georg ;
Damm, Philipp .
PLOS ONE, 2016, 11 (05)
[6]   Metal-on-metal hip simulator study of increased wear particle surface area due to 'severe' patient activity [J].
Bowsher, JG ;
Hussain, A ;
Williams, PA ;
Shelton, JC .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :279-287
[7]   TLEM 2.0-A comprehensive musculoskeletal geometry dataset for subject-specific modeling of lower extremity [J].
Carbone, V. ;
Fluit, R. ;
Pellikaan, P. ;
van der Krogt, M. M. ;
Janssen, D. ;
Damsgaard, M. ;
Vigneron, L. ;
Feilkas, T. ;
Koopman, H. F. J. M. ;
Verdonschot, N. .
JOURNAL OF BIOMECHANICS, 2015, 48 (05) :734-741
[8]   Future projections of total hip and knee arthroplasty in the UK: results from the UK Clinical Practice Research Datalink [J].
Culliford, D. ;
Maskell, J. ;
Judge, A. ;
Cooper, C. ;
Prieto-Alhambra, D. ;
Arden, N. K. .
OSTEOARTHRITIS AND CARTILAGE, 2015, 23 (04) :594-600
[9]   Patient characteristics affect hip contact forces during gait [J].
De Pieri, E. ;
Lunn, D. E. ;
Chapman, G. J. ;
Rasmussen, K. P. ;
Ferguson, S. J. ;
Redmond, A. C. .
OSTEOARTHRITIS AND CARTILAGE, 2019, 27 (06) :895-905
[10]   Refining muscle geometry and wrapping in the TLEM 2 model for improved hip contact force prediction [J].
De Pieri, Enrico ;
Lund, Morten E. ;
Gopalakrishnan, Anantharaman ;
Rasmussen, Kasper P. ;
Lunn, David E. ;
Ferguson, Stephen J. .
PLOS ONE, 2018, 13 (09)