A mainstream monitoring system for respiratory CO2 concentration and gasflow

被引:7
作者
Yang, Jiachen [1 ,2 ]
Chen, Bobo [1 ]
Burk, Kyle [3 ,4 ]
Wang, Haitao [1 ]
Zhou, Jianxiong [1 ]
机构
[1] Tianjin Univ, Sch Elect Informat Engn, Tianjin, Peoples R China
[2] Tianjin Yaan Polytron Technol Inc, Tianjin, Peoples R China
[3] Univ Utah, Dept Anesthesiol, Salt Lake City, UT USA
[4] Univ Utah, Dept Bioengn, Salt Lake City, UT 84112 USA
基金
中国国家自然科学基金;
关键词
Respiratory gas monitoring; CO2; Gasflow; Capnography; NDIR; Differential pressure; FEM; SENSOR;
D O I
10.1007/s10877-015-9739-y
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Continuous respiratory gas monitoring is an important tool for clinical monitoring. In particular, measurement of respiratory concentration and gasflow can reflect the status of a patient by providing parameters such as volume of carbon dioxide, end-tidal respiratory rate and alveolar deadspace. However, in the majority of previous work, concentration and gasflow have been studied separately. This study focuses on a mainstream system which simultaneously measures respiratory concentration and gasflow at the same location, allowing for volumetric capnography to be implemented. A non-dispersive infrared monitor is used to measure concentration and a differential pressure sensor is used to measure gasflow. In developing this new device, we designed a custom airway adapter which can be placed in line with the breathing circuit and accurately monitor relevant respiratory parameters. Because the airway adapter is used both for capnography and gasflow, our system reduces mechanical deadspace. The finite element method was used to design the airway adapter which can provide a strong differential pressure while reducing airway resistance. Statistical analysis using the coefficient of variation was performed to find the optimal driving voltage of the pressure transducer. Calibration between variations and flows was used to avoid pressure signal drift. We carried out targeted experiments using the proposed device and confirmed that the device can produce stable signals.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 16 条
  • [1] Tidal breath analysis for infant pulmonary function testing
    Bates, JHT
    Schmalisch, G
    Filbrun, D
    Stocks, J
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2000, 16 (06) : 1180 - 1192
  • [2] IMPROVED RESPONSE TIME WITH A NEW MINIATURISED MAIN-STREAM MULTIGAS MONITOR
    Berggren, Mattias
    Hosseini, Nasser
    Nilsson, Krister
    Stenqvist, Ola
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2009, 23 (06) : 355 - 361
  • [3] A Wireless Portable System With Microsensors for Monitoring Respiratory Diseases
    Cao, Zhe
    Zhu, Rong
    Que, Rui-Yi
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (11) : 3110 - 3116
  • [4] Can mainstream end-tidal carbon dioxide measurement accurately predict the arterial carbon dioxide level of patients with acute dyspnea in ED
    Cinar, Orhan
    Acar, Yahya Ayhan
    Arziman, Ibrahim
    Kilic, Erden
    Eyi, Yusuf Emrah
    Ocal, Ramazan
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2012, 30 (02) : 358 - 361
  • [5] Continuous Monitoring of Respiratory Flow and CO2 Challenges of On-Airway Measurements
    Jaffe, Michael B.
    Orr, Joseph A.
    [J]. IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2010, 29 (02): : 44 - 52
  • [6] Versatile Gas Detection System Based on Combined NDIR Transmission and Photoacoustic Absorption Measurements
    Kuehn, Karsten
    Pignanelli, Eliseo
    Schuetze, Andreas
    [J]. IEEE SENSORS JOURNAL, 2013, 13 (03) : 934 - 940
  • [7] Pekdemir M., 2013, RESP CARE
  • [8] Low Voltage CO2-Gas Sensor Based on III-V Mid-IR Immersion Lens Diode Optopairs: Where we Are and How Far we Can Go?
    Sotnikova, Galina Yu.
    Gavrilov, Gennadyy A.
    Aleksandrov, Sergey E.
    Kapralov, Alexander A.
    Karandashev, Sergey A.
    Matveev, Boris A.
    Remennyy, Maxim A.
    [J]. IEEE SENSORS JOURNAL, 2010, 10 (02) : 225 - 234
  • [9] St John Robert E, 2003, Crit Care Nurse, V23, P83
  • [10] Fourier domain local narrow-band signal extraction algorithm and its application to real-time infrared gas detection
    Xie, Qiong
    Li, Jianping
    Gao, Xiaoguang
    Jia, Jian
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2010, 146 (01) : 35 - 39