Corneal Ablation Depth Readout of the MEL 80 Excimer Laser Compared to Artemis Three-dimensional Very High-frequency Digital Ultrasound Stromal Measurements

被引:12
作者
Reinstein, Dan Z. [1 ,2 ,3 ,4 ]
Archer, Timothy J. [1 ]
Gobbe, Marine [1 ]
机构
[1] London Vis Clin, London W1G 7LA, England
[2] Kings Coll London, Dept Ophthalmol, St Thomas Hosp, London WC2R 2LS, England
[3] Cornell Univ, Dept Ophthalmol, Weill Med Coll, New York, NY 10021 USA
[4] Ctr Hosp Natl Ophtalmol, Paris, France
关键词
IN-SITU KERATOMILEUSIS; OPTICAL COHERENCE TOMOGRAPHY; EPITHELIAL THICKNESS PROFILE; PHOTOREFRACTIVE KERATECTOMY; FLAP THICKNESS; ORBSCAN PACHYMETRY; DIURNAL-VARIATIONS; LASIK ENHANCEMENT; PROBABILITY MODEL; TOPOGRAPHY;
D O I
10.3928/1081597X-20100114-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the accuracy of the ablation depth readout for the MEL 80 excimer laser (Carl Zeiss Meditec). METHODS: Artemis 1 very high-frequency digital ultrasound measurements were obtained before and at least 3 months after LASIK in 121 eyes (65 patients). The Artemis-measured ablation depth was calculated as the maximum difference in stromal thickness before and after treatment. Laser in situ keratomileusis was performed using the MEL 80 excimer laser and the Hansatome microkeratome (Bausch & Lomb). The Aberration Smart Ablation profile was used in 56 eyes and the Tissue Saving Ablation profile was used in 65 eyes. All ablations were centered on the corneal vertex. Comparative statistics and linear regression analysis were performed between the laser readout ablation depth and Artemis-measured ablation depth. RESULTS: The mean maximum myopic meridian was -6.66 +/- 2.40 diopters (D) (range: -1.50 to -10.00 D) for Aberration Smart Ablation-treated eyes and -6.50 +/- 2.56 D (range: -1.34 to -11.50 D) for Tissue Saving Ablation-treated eyes. The MEL 80 readout was found to overestimate the Artemis-measured ablation depth by 20 +/- 12 mu m for Aberration Smart Ablation and by 21 +/- 12 mu m for Tissue Saving Ablation profiles. CONCLUSIONS: The accuracy of ablation depth measurement was improved by using the Artemis stromal thickness profile measurements before and after surgery to exclude epithelial changes. The MEL 80 readout was found to overestimate the achieved ablation depth. The linear regression equations could be used by MEL 80 users to adjust the ablation depth for predicted residual stromal thickness calculations without increasing the risk of ectasia due to excessive keratectomy depth as long as a suitable flap thickness bias is included. [J Refract Surg. 2010; 26(12):949-959.] doi:10.3928/1081597X-20100114-02
引用
收藏
页码:949 / 959
页数:11
相关论文
共 60 条
[1]   Geometrical analysis of the loss of ablation efficiency at non-normal incidence [J].
Arba-Mosquera, Samuel ;
de Ortueta, Diego .
OPTICS EXPRESS, 2008, 16 (06) :3877-3895
[2]   Central Ablation Depth and Postoperative Refraction in Excimer Laser Myopic Correction Measured With Ultrasound, Scheimpflug, and Optical Coherence Pachymetry [J].
Arbelaez, Maria Clara ;
Vidal, Camila ;
Mosquera, Samuel Arba .
JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (08) :699-708
[3]  
BARRAQUER JI, 1980, QUERATOMILEUSIS QUER
[4]   Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes [J].
Chakrabarti, HS ;
Craig, JP ;
Brahma, A ;
Malik, TY ;
McGhee, CNJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) :1823-1828
[5]   Errors of residual stromal thickness estimation in LASIK [J].
Cheng, Huey-Chuan ;
Chen, Yen-Tung ;
Yeh, Shu-I ;
Yau, Chi-Wang .
OPHTHALMIC SURGERY LASERS & IMAGING, 2008, 39 (02) :107-113
[6]  
COLEMAN DJ, 1992, RADIOL CLIN N AM, V30, P1105
[7]   EXCIMER-LASER ABLATION RATE AND CORNEAL HYDRATION [J].
DOUGHERTY, PJ ;
WELLISH, KL ;
MALONEY, RK .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1994, 118 (02) :169-176
[8]   The predictability of corneal flap thickness and tissue laser ablation in laser in situ keratomileusis [J].
Durairaj, VD ;
Balentine, J ;
Kouyoumdjian, G ;
Tooze, JA ;
Young, D ;
Spivack, L ;
Taravella, MJ .
OPHTHALMOLOGY, 2000, 107 (12) :2140-2143
[9]   Confocal microscopy evaluation of stromal ablation depth after myopic laser in situ keratomileusis and photorefractive keratectomy [J].
Erie, JC ;
Hodge, DO ;
Bourne, WM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (02) :321-325
[10]   The effect of water content on the 193 nm excimer laser ablation [J].
Feltham, MH ;
Stapleton, F .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2002, 30 (02) :99-103