Partial volume effects on vascular T2 measurements

被引:27
作者
Stainsby, JA [1 ]
Wright, GA [1 ]
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Dept Med Biophys, N York, ON M4N 3M5, Canada
关键词
T-2; partial volume effects; oximetry;
D O I
10.1002/mrm.1910400322
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Quantitative, vascular T-2 measurements are of interest for applications such as MR oximetry, In the situation of a vessel with long T-2 relaxation times embedded in tissue with relatively short T-2 values, contamination of the blood signal from the surrounding tissue can bias T-2 measurements. Limited data sampling and vessels running obliquely through the imaging slice can cause significant signal contamination, Using a model of these effects to predict the behavior of T-2 measurements under a range of conditions, a set of parameters that provides the best combination of measurement accuracy with a signal-to-noise ratio as high as possible is proposed.
引用
收藏
页码:494 / 499
页数:6
相关论文
共 11 条
[11]   ESTIMATING OXYGEN-SATURATION OF BLOOD INVIVO WITH MR IMAGING AT 1.5T [J].
WRIGHT, GA ;
HU, BS ;
MACOVSKI, A .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (03) :275-283