Obstruction of mechanical valve prostheses: Clinical diagnosis and surgical or nonsurgical treatment

被引:26
作者
Aoyagi, S
Fukunaga, S
Suzuki, S
Nishi, Y
Oryoji, A
Kosuga, K
机构
[1] Second Department of Surgery, Kurume University, School of Medicine, Kurume 830
来源
SURGERY TODAY-THE JAPANESE JOURNAL OF SURGERY | 1996年 / 26卷 / 06期
关键词
prosthetic valve obstruction; valve thrombosis; cineradiography; thrombolysis; prosthetic valve replacement;
D O I
10.1007/BF00311926
中图分类号
R61 [外科手术学];
学科分类号
摘要
Twenty patients underwent nonsurgical and/or surgical treatment for obstruction of mechanical prosthetic valves, The obstructed prosthetic valve was in the aortic position in 11 patients, in the mitral position in 5, and in the tricuspid position in 4, Twelve patients had a bileaflet valve (3 aortic, 5 mitral, 4 tricuspid), and 8 had a tilting disk valve (all aortic), The diagnosis of prosthetic valve obstuction was made by cineradiography and echocardiography, Thrombolytic therapy was instituted in a series of our 10 most recent patients (11 cases), except for one patient with acute renal failure, regardless of the position of the obstructed prosthetic valve. Successful thrombolysis was achieved in 6 cases (54.5%), Six patients required surgical treatment subsequent to either failed or incomplete thrombolysis, and one patient died of congestive heart failure 1 month after surgery, Nonfatal neurologic events occurred in 2 cases (18.2%). A total of 16 patients underwent surgical treatment, Two (12.6%) of the 16 patients died of causes unrelated to the operative procedures before discharge from the hospital, These results suggest that thrombolytic therapy appears to be an attractive nonsurgical alternative for valve thrombosis when the patient's clinical condition is not critical, and thus surgical treatment should only be performed in an emergency on seriously ill patients.
引用
收藏
页码:400 / 406
页数:7
相关论文
共 26 条
[1]  
Aoyagi S, 1993, J Heart Valve Dis, V2, P279
[2]   OBSTRUCTION OF THE ST-JUDE MEDICAL VALVE - DIAGNOSTIC AND THERAPEUTIC VALUES OF CINERADIOGRAPHY [J].
AOYAGI, S ;
HIGA, Y ;
MATSUZOE, S ;
NISHI, Y ;
TANAKA, K ;
KAWARA, T ;
ORYOJI, A ;
KOSUGA, K ;
OISHI, K .
THORACIC AND CARDIOVASCULAR SURGEON, 1993, 41 (06) :357-363
[3]  
Aoyagi S., 1995, Japanese Journal of Artificial Organs, V24, P405
[4]  
AOYAGI S, 1989, JINKO ZOKI, V18, P753
[5]  
AYUSO LA, 1982, J THORAC CARDIOV SUR, V84, P906
[6]   THROMBOSIS ON BJORK-SHILEY AORTIC-VALVE PROSTHESIS - CLINICAL, ARTERIOGRAPHIC, ECHOCARDIOGRAPHIC AND THERAPEUTIC OBSERVATIONS IN 7 CASES [J].
BENZVI, J ;
HILDNER, FJ ;
CHANDRARATNA, PAN ;
SAMET, P .
AMERICAN JOURNAL OF CARDIOLOGY, 1974, 34 (05) :538-544
[7]  
BJORK VO, 1979, J THORAC CARDIOV SUR, V78, P331
[8]  
DEVERI E, 1991, J AM COLL CARDIOL, V17, P646
[9]   THE RISKS AND BENEFITS OF THROMBOLYTIC THERAPY IN ACUTE AORTIC AND MITRAL PROSTHETIC VALVE DYSFUNCTION - REPORT OF A CASE AND REVIEW OF THE LITERATURE [J].
GRAVER, LM ;
GELBER, PM ;
TYRAS, DH .
ANNALS OF THORACIC SURGERY, 1988, 46 (01) :85-88
[10]  
HORSTKOTTE D, 1983, CIRCULATION, V68, P175